Comparison of two devices for measuring exercise transcutaneous oxygen pressures in patients with claudication

VASA ◽  
2015 ◽  
Vol 44 (5) ◽  
pp. 355-362 ◽  
Author(s):  
Marie Urban ◽  
Alban Fouasson-Chailloux ◽  
Isabelle Signolet ◽  
Christophe Colas Ribas ◽  
Mathieu Feuilloy ◽  
...  

Abstract. Summary: Background: We aimed at estimating the agreement between the Medicap® (photo-optical) and Radiometer® (electro-chemical) sensors during exercise transcutaneous oxygen pressure (tcpO2) tests. Our hypothesis was that although absolute starting values (tcpO2rest: mean over 2 minutes) might be different, tcpO2-changes over time and the minimal value of the decrease from rest of oxygen pressure (DROPmin) results at exercise shall be concordant between the two systems. Patients and methods: Forty seven patients with arterial claudication (65 + / - 7 years) performed a treadmill test with 5 probes each of the electro-chemical and photo-optical devices simultaneously, one of each system on the chest, on each buttock and on each calf. Results: Seventeen Medicap® probes disconnected during the tests. tcpO2rest and DROPmin values were higher with Medicap® than with Radiometer®, by 13.7 + / - 17.1 mm Hg and 3.4 + / - 11.7 mm Hg, respectively. Despite the differences in absolute starting values, changes over time were similar between the two systems. The concordance between the two systems was approximately 70 % for classification of test results from DROPmin. Conclusions: Photo-optical sensors are promising alternatives to electro-chemical sensors for exercise oximetry, provided that miniaturisation and weight reduction of the new sensors are possible.

2012 ◽  
Vol 49 (6) ◽  
pp. 913 ◽  
Author(s):  
Beatriz Crespo-Ruiz ◽  
Antonio J. del-Ama ◽  
Fernando J. Jimnez-Daz ◽  
Jennifer Morgan ◽  
Ana de la Pea-Gonzlez ◽  
...  

2018 ◽  
Vol 276 ◽  
pp. 117-123 ◽  
Author(s):  
Pierre Abraham ◽  
Yongquan Gu ◽  
Lianrui Guo ◽  
Knut Kroeger ◽  
Nafi Ouedraogo ◽  
...  

2017 ◽  
Vol 22 (6) ◽  
pp. 490-497 ◽  
Author(s):  
Alban Fouasson-Chailloux ◽  
Pierre Abraham ◽  
Christophe Colas-Ribas ◽  
Mathieu Feuilloy ◽  
Bruno Vielle ◽  
...  

Data on simultaneous hemodynamic changes and pain rating estimation in arterial claudication while walking are lacking. This study was conducted to determine if a difference in transcutaneous oxygen pressure (tc pO2) exists between proximal and distal localization at pain appearance (PAINapp), maximal pain (PAINmax) and pain relief (PAINrel) in proximal or distal claudication and if a relationship exists between tc pO2 changes and pain intensity. We analyzed the pain rating (Visual Analog Scale (VAS)) to lower limb ischemia, measured with the decrease from rest of oxygen pressure (DROP) tc pO2 index during constant-load treadmill tests in patients with calf ( n = 41) or buttock ( n = 19) claudication. Calves versus buttocks results were analyzed with ANOVA tests. The R2 correlation coefficient between individual VAS versus DROP was calculated. Ischemia intensity versus pain rating changes were correlated. Significant ischemia was required for pain appearance, but pain disappeared despite the persistence of ischemia. We observed no statistical difference for DROP at PAINapp, PAINmax or PAINrel between proximal or distal claudication. A significant correlation between pain rating versus DROP was found: from PAINapp to PAINmax, R2 = 0.750 (calves) and 0.829 (buttocks), and from PAINmax to PAINrel, R2 = 0.608 (calves) and 0.560 (buttocks); p<0.05. Pain appeared after a significant decrease of hemodynamic parameters but disappeared while parameters were not normalized. No difference in pain rating was found in proximal versus distal claudication.


Sign in / Sign up

Export Citation Format

Share Document