scholarly journals Scatter rejection and low-contrast performance of a slot-scan digital chest radiography system with electronic aft-collimation: A chest phantom study

2008 ◽  
Vol 35 (6Part1) ◽  
pp. 2391-2402 ◽  
Author(s):  
Xinming Liu ◽  
Chris C. Shaw ◽  
Chao-Jen Lai ◽  
Mustafa C. Altunbas ◽  
Lingyun Chen ◽  
...  
1994 ◽  
Author(s):  
Carey E. Floyd, Jr. ◽  
Harrell G. Chotas ◽  
Carl E. Ravin

1994 ◽  
Vol 21 (4) ◽  
pp. 509-516 ◽  
Author(s):  
Ulrich Neitzel ◽  
Ingo Maack ◽  
Susanne Günther-Kohfahl

1989 ◽  
Vol 30 (4) ◽  
pp. 337-342 ◽  
Author(s):  
L. G. Månsson ◽  
S. Kheddache ◽  
D. Schlossman ◽  
J. Börjesson ◽  
E. Håkansson ◽  
...  

A digital system for chest radiography based on a large image intensifier was compared with a conventional film-screen system. The diagnostic performance was evaluated with special reference to the digital monitor images with a modified version of receiver operating characteristic (ROC) analysis—free response ROC (FROC) analysis—on a chest equivalent phantom. Measurements of spatial resolution and energy imparted were also performed. The detectability of low-contrast objects as well as spatial resolution was better for the full-size film-screen radiographs than for both the digital monitor images and the 100 mm photofluorograms. The image-intensifier system has a potential for considerable dose savings in relation to the conventional technique provided that fluoroscopy is excluded in the positioning of the patients.