Heterogeneous-Membrane, Carbon-Support, and Coated-Wire Ion-Selective Electrodes

2018 ◽  
pp. 131-176 ◽  
Author(s):  
R. W. Cattrall
2003 ◽  
Vol 36 (11) ◽  
pp. 2417-2426 ◽  
Author(s):  
Manal R. Al-Saraj ◽  
Salman M. Saadeh ◽  
Monzir S. Abdel-Latif

1972 ◽  
Vol 44 (4) ◽  
pp. 856-857 ◽  
Author(s):  
Helen J. James ◽  
Gary. Carmack ◽  
Henry. Freiser

2003 ◽  
Vol 58 (7) ◽  
pp. 658-662 ◽  
Author(s):  
Manal R. Al-Saraj ◽  
Salman M. Saadeh ◽  
Monzir S. Abdel-Latif

Several versions of Cu(II) ion selective electrodes (ISE), based on cyclized N,N'-bis(salicylidene)-o-phenylenediamine (salophen) complexes with Cu(II), were fabricated for determination of Cu(II) in aqueous solutions. The response of the ISE was optimized by variation of membrane composition and evaluation of various experimental conditions. Near Nernstian slopes (~ 28-32 mV/decade) were obtained for some preparations. The linear range of the ISE ranged from 5 × 10-5 to 1 × 10-2M Cu(II). Coated-wire and coated disc ISE resulted practically in a similar response as screen printed electrodes (SPE). The potentiometric selectivity coefficients (Kij) for all electrodes were determined for Na+, NH4+, Ca2+, Mg2+, Ni2+, Pb2+, Zn2+, Cd2+, Co2+, Fe3+, Hg2+, CO32-, H2PO4-, HPO42-, SO42-, CH3COO-, Br-, I-, NO3-, and SCN-. The selectivity coefficients were in the range from 10-2 to 10-3 for all ions tested except Hg2+, I-, and to less extent Fe3+. Fabricated ISE using the Cu(II)-salophen complex are reliable and stable.