Determination of Total Selenium Content in Cereals and Bakery Products by Flow Injection Hydride Generation Graphite Furnace Atomic Absorption Spectrometry Applying in-situ Trapping on Iridium-Treated Graphite Platforms

2005 ◽  
Vol 150 (1) ◽  
pp. 1-8 ◽  
Author(s):  
Zsolt Ajtony ◽  
Norbert Szoboszlai ◽  
Zsuzsanna Bella ◽  
Szilvia Bolla ◽  
Pál Szakál ◽  
...  
2014 ◽  
Vol 97 (6) ◽  
pp. 1696-1700 ◽  
Author(s):  
Tasneem G Kazi ◽  
Nida F Kolachi ◽  
Hassan I Afridi ◽  
Kapil Dev Brahman ◽  
Faheem Shah

Abstract The total selenium (Se) was determined in herbal and pharmaceutical supplements used for liver diseases. The total Se contents were determined in different pharmaceutical and herbal supplements by hydride generation atomic absorption spectrometry (HGAAS) and graphite furnace atomic absorption spectrometry (GFAAS) after microwave-assisted acid digestion. The accuracy of the techniques was evaluated by using certified reference material and the standard addition method. The recoveries of total Se were 99.4 and 99.0% for HGAAS and GFAAS, respectively. The precision of the techniques expressed as RSD were 2.34 and 4.54% for HGAAS and GFAAS measurements, respectively. The LOD values for HGAAS and GFAAS were 0.025 and 0.052 μg/g, respectively. The concentrations of Se in pharmaceutical and herbal supplements were found in the range of 19.2–53.8 and 25.0–42.5 μg/g, respectively, corresponding to 35–76% and 45–76% of the total recommended dose of Se for adults.


Sign in / Sign up

Export Citation Format

Share Document