SU-FF-T-326: A Systematic Methodology in Dosimetric Verification of Anisotropic Analytical Algorithm in Eclipse Treatment Planning System

2009 ◽  
Vol 36 (6Part14) ◽  
pp. 2596-2596
Author(s):  
R He ◽  
C Yang
2013 ◽  
Vol 4 (1) ◽  
pp. 43-49
Author(s):  
M Jahangir Alam ◽  
Syed Md Akram Hussain ◽  
Kamila Afroj ◽  
Shyam Kishore Shrivastava

A three dimensional treatment planning system has been installed in the Oncology Center, Bangladesh. This system is based on the Anisotropic Analytical Algorithm (AAA). The aim of this study is to verify the validity of photon dose distribution which is calculated by this treatment planning system by comparing it with measured photon beam data in real water phantom. To do this verification, a quality assurance program, consisting of six tests, was performed. In this program, both the calculated output factors and dose at different conditions were compared with the measurement. As a result of that comparison, we found that the calculated output factor was in excellent agreement with the measured factors. Doses at depths beyond the depth of maximum dose calculated on-axis or off-axis in both the fields or penumbra region were found in good agreement with the measured dose under all conditions of energy, SSD and field size, for open and wedged fields. In the build up region, calculated and measured doses only agree (with a difference 2.0%) for field sizes > 5 × 5 cm2 up to 25 × 25 cm2. For smaller fields, the difference was higher than 2.0% because of the difficulty in dosimetry in that region. Dose calculation using treatment planning system based on the Anisotropic Analytical Algorithm (AAA) is accurate enough for clinical use except when calculating dose at depths above maximum dose for small field size.DOI: http://dx.doi.org/10.3329/bjmp.v4i1.14686 Bangladesh Journal of Medical Physics Vol.4 No.1 2011 43-49


2013 ◽  
Vol 55 ◽  
pp. 60-63 ◽  
Author(s):  
Jeong-Eun Rah ◽  
Dongho Shin ◽  
Jeong-hoon Park ◽  
Ui-Jung Hwang ◽  
Do Hoon Oh ◽  
...  

2020 ◽  
Vol 26 (3) ◽  
pp. 127-134
Author(s):  
Md Akhtaruzzaman ◽  
Paweł Kukołowicz

AbstractPurpose: The purpose of the study was to investigate the dependence of tissue inhomogeneity correction factors (ICFs) on the photon beam quality index (QI).Materials and Methods: Heterogeneous phantoms, comprising semi-infinite slabs of the lung (0.10, 0.20, 0.26 and 0.30 g/cm3), adipose tissue (0.92 g/cm3) and bone (1.85 g/cm3) in water, were constructed in the Eclipse treatment planning system. Several calculation models of 6 MV and 15 MV photon beams for quality index (TPR20,10) = 0.670±k*0.01 and TPR20,10 = 0.760±k*0.01, k = -3, -2, -1, 0, 1, 2, 3 respectively were built in the Eclipse. The ICFs were calculated with the anisotropic analytical algorithm (AAA) for several beam sizes and points lying at several depths inside of and below inhomogeneities of different thicknesses.Results: The ICFs increased for lung and adipose tissues with increasing beam quality (TPR20,10), while decreased for bone. Calculations with AAA predict that the maximum difference in ICFs of 1.0% and 2.5% for adipose and bone tissues, respectively. For lung tissue, changes of ICFs of a maximum of 9.2% (6 MV) and 13.8% (15 MV). For points where charged particle equilibrium exists, a linear dependence of ICFs on TPR20,10 was observed. If CPE doesn’t exist, the dependence became more complex. For points inside of the low-density inhomogeneity, the dependence of the ICFs on energy was not linear but the changes of ICFs were smaller than 3.0%. Measurements results carried out with the CIRS phantom were consistent with the calculation results.Conclusions: A negligible dependence of the ICFs on energy was found for adipose and bone tissue. For lung tissue, in the CPE region, the dependence of ICFs on different beam quality indexes with the same nominal energy may not be neglected, however, this dependence was linear. Where there is no CPE, the dependence of the ICFs on energy was more complicated.


1999 ◽  
Vol 44 (2) ◽  
pp. 463-478 ◽  
Author(s):  
Lei Xing ◽  
Bruce Curran ◽  
Robert Hill ◽  
Tim Holmes ◽  
Lijun Ma ◽  
...  