step and shoot
Recently Published Documents


TOTAL DOCUMENTS

314
(FIVE YEARS 26)

H-INDEX

30
(FIVE YEARS 1)

2022 ◽  
Vol 17 (01) ◽  
pp. P01006
Author(s):  
M. Saed ◽  
A. Sadremomtaz ◽  
H. Mahani

Abstract Background: the need for simultaneous high-sensitivity and high-resolution breast SPECT imaging mandates to design and optimize dedicated scanners. Therefore, this work aims to design and optimize a novel breast-dedicated SPECT system with multi-lofthole collimator. Materials and Methods: in this research, a novel breast-dedicated scanner is designed and then optimized. The scanner is equipped with a single full-ring multi-lofthole collimation long with modular NaI(Tl) detectors. The step-and-shoot data acquisition was considered with two steps. Then, an analytic optimization was conducted to balance the existing sensitivity-resolution tradeoff. To do so, several scanner geometries were investigated. The optimal configuration maximized the system sensitivity at a given system resolution. Furthermore, the scanner was also modeled within the GATE simulator. Then, detector energy resolution, septal penetration and scattering, and system sensitivity were calculated. Analytic findings were also compared with the simulated ones. Results: the results showed that high sensitivity of about 2 cps/kBq can be obtained for a diameter of lofthole 3.05 mm with a 75° opening angle. Results of GATE simulations showed clinically acceptable performance of the system offering 9% energy resolution for a point source. The septal penetration and scattering were approximately 0.5% and 0.2%, respectively, for cylindrical water phantom and tungsten as collimator material. Conclusion: the designed SPECT scanner provides promising results in terms of sensitivity and spatial resolution and therefore outperforms the traditional multi-pinhole collimation by a much higher sensitivity at a given system resolution.


2021 ◽  
Author(s):  
Mengyu Jia ◽  
Yan Wu ◽  
Yong Yang ◽  
Lei Wang ◽  
Cynthia Chuang ◽  
...  

Author(s):  
Alison Gray ◽  
Omemh Bawazeer ◽  
Sankar Arumugam ◽  
Philip Vial ◽  
Joseph Descallar ◽  
...  
Keyword(s):  

2021 ◽  
Vol 8 (2) ◽  
Author(s):  
De Ornelas M ◽  
◽  
Mihaylov IB ◽  

Purpose: To determine the sensitivity of changes to IMRT delivery parameters for mass-based optimization schemes: Dose-Mass- (DM) and Energy-based (Energy), compared to Dose-Volume-based (DV) optimization. Methods: Twelve Head-and-Neck (HN) and twelve lung cases were retrospectively optimized using DM and Energy optimization. In both optimization approaches nine equidistant, split beams were used for step-and-shoot deliverable IMRT. Changes to two parameters were investigated: the number of IMRT segments (5 and 10 per beam) and the minimum allowed segment area (2 and 6 cm²). Plans were normalized such that 95% of the PTV received the same dose. Dose Indices (DIs) were used for evaluation. For the lung cases, DIs included: 1%_cord, 33%_heart, 20% and 30%_both-lungs, and 50%_ esophagus. In the HN cases: 1%_cord, 1%_brainstem, left/right parotids_50%, 50%_larynx, and 50%_esophagus. Results: The lung cases demonstrated that the Energy plans were more sensitive to segment area; changing the segment area resulted in a statistically significant dose increase for 1%_cord, 30%_both-lungs and 50%_esophagus. Changes to the number of segments yielded on average statistically significant differences in dose to 1%_cord in Energy plans, 50%_esophagus in DM plans, and 20%_both-lungs in DV plans. When the segment area was changed, the HN cases yielded statistically significant differences in doses to 1%_cord, 1%_ brainstem, 50%_left and right parotids, and 50%_larynx for the Energy plans and 50%_larynx for DM plans. Moreover, changing the number of segments resulted in significant dose decrease for 50%_parotids and 50%_esophagus for the Energy plans and 50%_larynx for DV plans. Conclusions: This study showed that both lung and HN Energy plans exhibit larger sensitivity than DV and DM plans to changing IMRT delivery parameters, especially when increasing the minimum segment area rather than with varying the number of segments.


2021 ◽  
Author(s):  
John Baines ◽  
Ariadne Shoobridge

Abstract During the adaptive workflow associated with MRgRT, a secondary dose calculation is required and MU2net (DOSIsoft, France) is one commercial option. The suitability of MU2net to be used in conjunction with the online Monaco treatment planning system of the Elekta Unity (Elekta AB, Stockholm, Sweden), is evaluated in this work. Monaco and MU2net point doses are compared for various fields on and off axis and at different SSDs. To investigate the comparative effects of attenuation due to the cryostat, couch and posterior coil, measured, MU2net and Monaco dose outputs at the isocentre, as a function of gantry angle, were compared. Point doses for the beams of nine step and shoot IMRT (SSIMRT) test plans (courtesy Elekta) were calculated with Monaco v5.4 and compared to corresponding doses computed with MU2net. In addition, Monaco v5.4 and MU2net point doses were compared for 1552 beams treated on the Unity at our facility. For the on-axis fields investigated the agreement between MU2net and measured data is acceptable. MU2net and Monaco point doses for the Elekta SSIMRT test plans were within ± 5.0 % and ± 6.4 % for beams delivered from gantry zero and at planned beam angles, respectively. For the 1552 beams delivered approximately 80.0 % of MU2net and Monaco point doses agree within ± 5.0 %, therefore it is recommended to correlate MU2net Dose Reference Points (DRPs )with pre and post treatment dosimetry verification. Computational accuracy of MU2net could be enhanced with improved modelling of attenuation due to the couch, cryostat and posterior MR imaging coil.


2021 ◽  
pp. 20210084
Author(s):  
Yoshihiro Ozaki ◽  
Hiroshi Watanabe ◽  
Tohru Kurabayashi

Objective: The objective of this study was to perform effective dose estimation in cone-beam CT for dental use (CBCT) using a Monte-Carlo simulation employing a step-and-shoot method as well as to determine the optimal number of steps. Methods: We simulated 3DX Accuitomo FPD8 as a CBCT model and estimated the effective doses of a large and a small field of view (FOV) examination against the virtual Rando phantom using a particle and heavy ion transport code system. We confirmed the results compared to those from a thermo-luminescence dosemeter (TLD) system in a real phantom and investigated how the reduced angle calculations could be accepted. Results: The effective doses of both FOVs estimated with each one degree were almost the same as those estimated from the TLD measurements. Considering the effective doses and the itemized organ doses, simulation with 5° and 10° is acceptable for the large and small FOV, respectively. We tried to compare an effective dose with a large FOV as well as with multiple small FOVs covering the corresponding area and found that the effective dose from six small FOVs was approximately 1.2 times higher than that of the large FOVs. Conclusion: We successfully performed a Monte-Carlo simulation using a step-and-shoot method and estimated the effective dose in CBCT. Our findings indicate that simulation with 5° or 10° is acceptable based on the FOV size, while a small multiple FOV scan is recommended from a radiation protection viewpoint.


PLoS ONE ◽  
2021 ◽  
Vol 16 (4) ◽  
pp. e0250243
Author(s):  
Chun-Yen Yu ◽  
Shih-Wen Wan ◽  
Yih-Chyang Weng ◽  
Ching-Han Hsu

We measured the overshoot effect in a linac and reduced it using block correction, reverse-sequence correction, and index correction. A StarTrack detector was used on a Varian iX. Five segments, 1 × 10 cm2 in area, were designed; the centers were at −4, −2, 0, 2, and 4 cm on the x axis for measuring the overshoot effect on a 10 × 10 cm2 collimator setting. Block correction was applied to two segments. The first was on the new first segment at −6 cm, and the other was on the new last segment at 6 cm. Both two new segments were obtained from the 10 × 10 cm2 collimator setting. The order of segments was reversed in reverse-sequence correction. Reverse-sequence correction averages the dose at every segment after two irradiations. When we used MLC Shaper, index correction reduced the first segment’s index (cumulative radiation occupation) by 60% and increased the last segment’s radiation occupation by 60% in a new MLC.log file. As for relative dose, the first segment had an overdose of 52.4% and the last segment had an underdose of 48.6%, when irradiated at 1 MU at 600 MU/min. The relative doses at the first segment, irradiated at 1 MU, after block correction, reverse-sequence correction, and index correction were applied decreased from 152.5% to 95.1%, 104.8%, and 100.1%, respectively. The relative doses at the last segment, irradiated at 1 MU, after block correction, reverse-sequence correction, and index correction were applied increased from 48.6% to 97.3%, 91.1%, and 95.9%, respectively. The overshoot effect depended on the speed of irradiation. High irradiation speeds resulted in notable overdosing and underdosing at the first and last segments, respectively. The three corrections mitigated the overshoot effect on dose. To save time and effort, the MLC.log file should be edited with a program in the future.


Author(s):  
M. Athiyaman ◽  
A. Hemalatha ◽  
Arun Chougule ◽  
Mary Joan ◽  
HS Kumar

Abstract Introduction: The control points (CP) play a significant role in the delivery of segmented based Intensity-Modulated Radiation Therapy (IMRT) delivery, particularly in dynamic mode. The number of segments is determined by control points and these segments will transfer from one to the other either during beam ON called dynamic delivery or during beam OFF called static delivery or step and shoot. This study was aimed at indirect estimation of the total monitor units (MU) to be delivered per field by exploiting the control points and also to find the MUs at any nth segment. Materials and methods: This study was performed in the Eclipse treatment planning software version 13.8.0. The details of control points, metre set weight per segment, leaf positions for each segment, field size, etc. were taken into consideration. Results: TPS calculated MU value and analytically estimated MU value were compared and the percentage of difference was estimated. The overall mean percentage of deviation was 1·03% between the TPS calculated method and the analytical method. The paired sample t-test was performed and, p-value <0·05, no significant difference was found. The analytical relationship determined to estimate the total number of MU delivered for any nth control point was also evaluated. Conclusion: The control points are a potential parameter in the conventional IMRT delivery. Through this study, we have addressed the indirect method to estimate the monitor units delivered per segment.


Sign in / Sign up

Export Citation Format

Share Document