volar plate osteosynthesis
Recently Published Documents


TOTAL DOCUMENTS

12
(FIVE YEARS 2)

H-INDEX

6
(FIVE YEARS 0)

2016 ◽  
Vol 17 (1) ◽  
Author(s):  
Alexaner Synek ◽  
Lars Borgmann ◽  
Hannes Traxler ◽  
Wolfgang Huf ◽  
Ekkehard Euler ◽  
...  

2013 ◽  
Vol 38 (7) ◽  
pp. 751-757 ◽  
Author(s):  
I. Mehling ◽  
P. Rittstieg ◽  
A. P. Mehling ◽  
R. Küchle ◽  
L. P. Müller ◽  
...  

The purpose of this study was to analyze the practicability and benefit of intraoperative C-arm computed tomography (CT) imaging in volar plate osteosynthesis of unstable distal radius fractures. During a 1 year period, intraoperative three dimensional (3D) imaging with the ARCADIS Orbic 3D was performed in addition to standard fluoroscopy in 51 cases. The volar angular stable plate oesteosyntheses were analyzed intraoperatively and, if necessary, improved immediately. The duration of the scan and radiation exposure dose were measured. On average, performance of the scan and analysis of the CT dataset took 6.7 minutes. In 31.3% of the surgeries a misplacement of screws was detected and correction was done immediately. C-arm CT imaging can easily be integrated in the normal course of surgery. As a complement to the standard 2D-fluoroscopy, the C-arm CT is a useful tool to evaluate the quality of osteosynthesis.


Sign in / Sign up

Export Citation Format

Share Document