Impact of implant depth on hydrodynamic function with the ACURATE neo transcatheter heart valve following valve-in-valve transcatheter aortic valve replacement in Mitroflow bioprosthetic valves: an ex vivo bench study

2019 ◽  
Vol 15 (1) ◽  
pp. 78-87 ◽  
Author(s):  
Janarthanan Sathananthan ◽  
Stephanie L. Sellers ◽  
Rob Fraser ◽  
Danny Dvir ◽  
Mark Hensey ◽  
...  
2021 ◽  
Vol 10 (6) ◽  
Author(s):  
Sascha Macherey ◽  
Max Meertens ◽  
Victor Mauri ◽  
Christian Frerker ◽  
Matti Adam ◽  
...  

Background During the past decade, the use of transcatheter aortic valve replacement (TAVR) was extended beyond treatment‐naïve patients and implemented for treatment of degenerated surgical bioprosthetic valves. Selection criteria for either valve‐in‐valve (viv) TAVR or redo surgical aortic valve replacement are not well established, and decision making on the operative approach still remains challenging for the interdisciplinary heart team. Methods and Results This review was intended to analyze all studies on viv‐TAVR focusing on short‐ and mid‐term stroke and mortality rates compared with redo surgical aortic valve replacement or native TAVR procedures. A structured literature search and review process led to 1667 potentially relevant studies on July 1, 2020. Finally, 23 studies fulfilled the inclusion criteria for qualitative analysis. All references were case series either with or without propensity score matching and registry analyses. Quantitative synthesis of data from 8509 patients revealed that viv‐TAVR is associated with mean 30‐day stroke and mortality rates of 2.2% and 4.2%, respectively. Pooled data analysis showed no significant differences in 30‐day stroke rate, 30‐day mortality, and 1‐year mortality between viv‐TAVR and comparator treatment (native TAVR [n=11 804 patients] or redo surgical aortic valve replacement [n=498 patients]). Conclusions This review is the first one comparing the risk for stroke and mortality rates in viv‐TAVR procedures with native TAVR approach and contributes substantial data for the clinical routine. Moreover, this systematic review is the most comprehensive analysis on ischemic cerebrovascular events and early mortality in patients undergoing viv‐TAVR. In this era with increasing numbers of bioprosthetic valves used in younger patients, viv‐TAVR is a suitable option for the treatment of degenerated bioprostheses.


2021 ◽  
Author(s):  
Frane Runjic ◽  
◽  
Andrija Matetic ◽  
Matjaz Bunc ◽  
Nikola Crncevic ◽  
...  

Background: Patient-prothesis mismatch (PPM) is a serious potential complication following surgical aortic valve replacement (SAVR). If it develops, valve-in-valve transcatheter aortic valve replacement (TAVR) is a reasonable therapeutic option. However, there is low evidence on the management of small degenerated surgical bioprosthetic valves, not prone to balloon-valve fracture (BVF). Case Presentation: This case report presents a successful valve-in-valve TAVR in acute heart failure due to degenerative surgical bioprosthetic valve Trifecta (21 mm) that is not susceptible to BVF. Standard preparation for transfemoral TAVR with a self-expandable valve was conducted, including the over-the-wire pacing. Thereafter, a successful valve-in-valve primary implantation of the self-expanding, supra-annular valve Evolut R 26 (Medtronic™) has been achieved. Follow-up at 3 months showed mild paravalvular leak in the region with clinical and heart function improvements of the patient. Follow-up echocardiographic parameters showed the reduction of anterograde flow impairment and improved effective orifice area (~0.85 cm2/m2). Conclusion: In conclusion, supra-annular valve-in-valve TAVR is a potential therapeutic option for PPM of small degenerated surgical bioprosthetic valves which are not prone to BVF.


2021 ◽  
Vol 14 (2) ◽  
pp. 211-220 ◽  
Author(s):  
Michel Pompeu B.O. Sá ◽  
Jef Van den Eynde ◽  
Matheus Simonato ◽  
Luiz Rafael P. Cavalcanti ◽  
Ilias P. Doulamis ◽  
...  

2021 ◽  
Vol 5 (5) ◽  
Author(s):  
Klaus-Dieter Hönemann ◽  
Steffen Hofmann ◽  
Frank Ritter ◽  
Gerold Mönnig

Abstract Background A rare, but serious, complication following transcatheter aortic valve replacement (TAVR) is the occurrence of an iatrogenic ventricular septal defect (VSD). Case summary We describe a case of an 80-year-old female who was referred with severe aortic stenosis for TAVR. Following thorough evaluation, the heart team consensus was to proceed with implantation via a transapical approach of an ACURATE neo M 25 mm valve (Boston Scientific, Natick, MA, USA). The valve was deployed harnessing transoesophageal echocardiographic (TOE) guidance under rapid pacing with post-dilation. Directly afterwards a very high VSD close to the aortic annulus was detected. As the patient was haemodynamically stable, the procedure was ended. The next day another TOE revealed a shunt volume (left-to-right ventricle) between 50% and 60%. Because the defect was partly located between the stent struts of the ACURATE valve decision was made to fix this leakage with implantation of a further valve and we chose an EVOLUT Pro 29 mm (Medtronic Inc., Minneapolis, MN, USA). The valve-in-valve was implanted 2–3 mm below the lower edge of the first valve, more towards the left ventricular outflow tract (LVOT) with excellent result: VSD was reduced to a very small residual shunt without any hemodynamic relevance. Discussion We suggest that an iatrogenic VSD located near the annulus may be treated percutaneously in a bail-out situation with implantation of a second valve that should be implanted slightly more into the LVOT to cover the VSD.