Determination of “free” glycerol in human serum reference materials by isotope-dilution gas chromatography—mass spectrometry

Author(s):  
John T. Bernert ◽  
Carol J. Bell ◽  
James E. McGuffey ◽  
Parvin P. Waymack
1989 ◽  
Vol 72 (2) ◽  
pp. 349-354 ◽  
Author(s):  
Lee Q Huang

Abstract A multiresidue method was developed for the simultaneous determination of low parts per billion (ppb) concentrations of the herbicides alachlor, metolachlor, atrazine, and simazine in water and soil using isotope dilution gas chromatography/mass spectrometry (GC/MS). Known amounts of 15N,13C-alachlor and 2H5-atrazine were added to each sample as internal standards. The samples were then prepared by a solid phase extraction with no further cleanup. A high resolution GC/low resolution MS system with data acquisition in selected ion monitoring mode was used to quantitate herbicides in the extract. The limit of detection was 0.05 ppb for water and 0.5 ppb for soil. Accuracy greater than 80% and precision better than 4% was demonstrated with spiked samples.


1991 ◽  
Vol 37 (4) ◽  
pp. 540-546 ◽  
Author(s):  
Linda Thienpont ◽  
Lothar Siekmann ◽  
Alexander Lawson ◽  
Elisabeth Colinet ◽  
Andrȳ De Leenheer

Abstract The Community Bureau of Reference of the European Communities has produced four batches of lyophilized serum Certified Reference Materials, two for cortisol (CRM 192 and 193) and two for progesterone (CRM 347 and 348). For cortisol, one of the pools consisted of serum from healthy blood donors, whereas the second batch was supplemented with pure cortisol. The progesterone Reference Materials contained only endogenous hormone concentrations. Assessment of vial-to-vial variability in the cortisol and progesterone concentrations showed no between-sample inhomogeneity, and the materials were stable. The quality of the materials was therefore considered sufficient for certification of the values for the cortisol and progesterone concentrations by a collaborative study involving several laboratories from the European Communities, using isotope dilution gas chromatography-mass spectrometry. Inaccuracy in reconstitution of the lyophilized materials was less than 0.3%; imprecision of sampling was less than 0.2%. For determinations of cortisol and progesterone concentrations, the mean within-laboratory coefficients of variation (CVs) were 1.76% (CRM 192), 1.19% (CRM 193), 1.64% (CRM 347), and 1.75% (CRM 348). The between-laboratory CVs were greater: CRM 192, 1.79%; CRM 193, 1.48%; CRM 347, 2.08%; and CRM 348, 2.16%. The concentrations in the reconstituted Reference Materials were certified to be 273 nmol/L in CRM 192 and 763 nmol/L in CRM 193 for cortisol and 10.13 nmol/L in CRM 347 and 40.3 nmol/L in CRM 348 for progesterone. Uncertainties at the 0.95 confidence level--6 (CRM 192), 14 (CRM 193), 0.21 (CRM 347), and 1.0 nmol/L (CRM 348)--were considered compatible with the intended use of the materials.


Sign in / Sign up

Export Citation Format

Share Document