Full energy peak efficiency calibration for the assay of large volume radioactive waste drums in a segmented gamma scanner

2019 ◽  
Vol 144 ◽  
pp. 80-86
Author(s):  
Sabyasachi Patra ◽  
Chhavi Agarwal ◽  
Sanhita Chaudhury
2016 ◽  
Vol 19 (2) ◽  
pp. 71-80
Author(s):  
Chuong Dinh Huynh ◽  
Dan Tieu Luu ◽  
Nguyen Hoang Vo ◽  
Thanh Thien Tran ◽  
Tao Nhat Chau

In this paper, we present an analytical calculation model of full energy peak efficiency for cylindrical detectors without collimator based on efficiency transfer method. A calculation program by Mathematica language is developed to apply to this model. The validity of the calculation model was checked by comparison with MCNP5 simulated efficiency values for measurements of point source in the waste drum containing matrix of rubber or concrete. The discrepancy between MCNP5 simulated and calculated efficiencies is smaller 11 %. This shows that the calculation model is reliable and can be applied to calculate the full energy peak efficiency for assays of radioactive waste drums. Besides, the calculated time by the this program is much faster than the simulation using MCNP5 program.