It is now well-known that chatter (Figure 1) is caused by vibration between the microtome arm and the diamond knife. It is usually observed as a cyclical variation in “optical” density of an electron micrograph due to sample thickness variations perpendicular to the cutting direction. This vibration might be induced by using too large a block face, too large a clearance angle, excessive cutting speed, non-uniform embedding medium or microtome vibration. Another prominent cause is environmental vibration caused by inadequate building construction. Microtomes should be installed on firm, solid floors. The best floors are thick, ground-level concrete pads poured over a sand bed and isolated from the building walls. Even when these precautions are followed, we recommend an additional isolation pad placed on the top of a sturdy table.