Inappropriate shock due to quadruple counting in a patient with subcutaneous implantable cardioverter-defibrillator and a dual-chamber pacemaker

EP Europace ◽  
2020 ◽  
Vol 22 (5) ◽  
pp. 796-796
Author(s):  
Halim Marzak ◽  
Olivier Morel ◽  
Laurence Jesel
2021 ◽  
pp. 1-3
Author(s):  
Sezen Gulumser Sisko ◽  
Hasan Candas Kafali ◽  
Yakup Ergul

Abstract We report a patient with long QT syndrome who received an inappropriate implantable cardioverter-defibrillator shock due to electrical interference from a refrigerator. This electrical interference was mistakenly detected as an episode of ventricular fibrillation and ended with an inappropriate delivery of shock without any warning symptoms before.