scholarly journals A paradigmatic case of haemolysis and pseudohyperkalemia in blood gas analysis

2018 ◽  
Vol 29 (1) ◽  
pp. 169-172
Author(s):  
Gian Luca Salvagno ◽  
Davide Demonte ◽  
Giuseppe Lippi

A 51-year old male patient was admitted to the hospital with acute dyspnea and history of chronic asthma. Venous blood was drawn into a 3.0 mL heparinized syringe and delivered to the laboratory for blood gas analysis (GEM Premier 4000, Instrumentation Laboratory), which revealed high potassium value (5.2 mmol/L; reference range on whole blood, 3.5-4.5 mmol/L). This result was unexpected, so that a second venous blood sample was immediately drawn by direct venipuncture into a 3.5 mL lithium-heparin blood tube, and delivered to the laboratory for repeating potassium testing on Cobas 8000 (Roche Diagnostics). The analysis revealed normal plasma potassium (4.6 mmol/L; reference range in plasma, 3.5-5.0 mmol/L) and haemolysis index (5; 0.05 g/L). Due to suspicion of spurious haemolysis, heparinized blood was transferred from syringe into a plastic tube and centrifuged. Potassium and haemolysis index were then measured in this heparinized plasma, confirming high haemolysis index (50; 0.5 g/L) and pseudohyperkalemia (5.5 mmol/L). Investigation of this case revealed that spurious haemolysis was attributable to syringe delivery in direct ice contact for ~15 min. This case emphasizes the importance of avoiding sample transportation in ice and the need of developing point of care analysers equipped with interference indices assessment.

Resuscitation ◽  
2017 ◽  
Vol 118 ◽  
pp. e45
Author(s):  
Susanne Betz ◽  
Tobias Grübl ◽  
Martin Sassen ◽  
Erich Wranze ◽  
Heiko Hartmann ◽  
...  

1994 ◽  
Vol 40 (1) ◽  
pp. 124-129 ◽  
Author(s):  
R J Wong ◽  
J J Mahoney ◽  
J A Harvey ◽  
A L Van Kessel

Abstract We evaluated a new portable instrument, the PPG StatPal II pH and Blood Gas Analysis System, designed for "point-of-care" measurements of blood gases and pH. Inaccuracy (% of target value) and imprecision (CV%) were assessed by blood tonometry and comparison with a Corning 178. Within-day results for PCO2 inaccuracy and imprecision ranged from 98.2% to 102.9% and 3.3% to 3.9%, respectively; for PO2, these were 95.5% to 102.3% and 2.3% to 3.0%, respectively. Between-day results for PCO2 inaccuracy and imprecision ranged from 99.2% to 99.3% and from 2.9% to 3.2%, respectively; for PO2, the ranges were 96.2% to 98.2% and 2.6% to 3.0%, respectively. Two PCO2 outliers (in 645 samples = 0.3%) were observed. In general, tonometry recovery, measurement stability, and pH bias results for the StatPal II and Corning 178 were comparable. We conclude that the StatPal II performs within acceptable ranges of inaccuracy and imprecision.


Author(s):  
Lukasz Krzych ◽  
Olga Wojnarowicz ◽  
Paweł Ignacy ◽  
Julia Dorniak

Introduction. Reliable results of an arterial blood gas (ABG) analysis are crucial for the implementation of appropriate diagnostics and therapy. We aimed to investigate the differences (Δ) between ABG parameters obtained from point-of-care testing (POCT) and central laboratory (CL) measurements, taking into account the turnaround time (TAT). Materials and methods. A number of 208 paired samples were collected from 54 intensive care unit (ICU) patients. Analyses were performed using Siemens RAPIDPoint 500 Blood Gas System on the samples just after blood retrieval at the ICU and after delivery to the CL. Results. The median TAT was 56 minutes (IQR 39-74). Differences were found for all ABG parameters. Median Δs for acid-base balance ere: ΔpH=0.006 (IQR –0.0070–0.0195), ΔBEef=–0.9 (IQR –2.0–0.4) and HCO3–act=–1.05 (IQR –2.25–0.35). For ventilatory parameters they were: ΔpO2=–8.3 mmHg (IQR –20.9–0.8) and ΔpCO2=–2.2 mmHg (IQR –4.2––0.4). For electrolytes balance the differences were: ΔNa+=1.55 mM/L (IQR 0.10–2.85), ΔK+=–0.120 mM/L (IQR –0.295–0.135) and ΔCl–=1.0 mM/L (IQR –1.0–3.0). Although the Δs might have caused misdiagnosis in 51 samples, Bland-Altman analysis revealed that only for pO2 the difference was of clinical significance (mean: –10.1 mmHg, ±1.96SD –58.5; +38.3). There was an important correlation between TAT and ΔpH (R=0.45, p<0.01) with the safest time delay for proper assessment being less than 39 minutes. Conclusions. Differences between POCT and CL results in ABG analysis may be clinically important and cause misdiagnosis, especially for pO2. POCT should be advised for ABG analysis due to the impact of TAT, which seems to be the most important for the analysis of pH.