Use of a closed-system drug transfer device reduces contamination with doxorubicin during bolus injection
Background: Administration of doxorubicin via bolus injection may result in environmental contamination and a risk of nurses becoming exposed. Small spills are frequently observed by nurses when syringes are connected to, and disconnected from, infusion lines. Aims: The effect of a closed-system drug transfer device (CSTD) on the release of doxorubicin was studied during administration via bolus injections. Methods: 10 administrations with the currently used technique and 10 administrations using the CSTD were compared by analysis of doxorubicin contamination on gauze pads, tissues and gloves. Findings: Using the current technique, contamination was found during nine administrations, which was mainly on the gauze pads and, to a lesser extent, on the tissues and gloves, indicating release of doxorubicin during administration. With use of the CSTD, contamination was found only on one pair of gloves. Conclusion: Use of a CSTD significantly decreased the number of spills and level of contamination compared with the currently used technique and, consequently, the use of such devices offers a safer working environment for nurses.