scholarly journals Evaluating the Effects of Dose Rate on Dynamic Intensity-Modulated Radiation Therapy Quality Assurance

2021 ◽  
Vol 32 (4) ◽  
pp. 116-121
Author(s):  
Kwon Hee Kim ◽  
Tae Seong Back ◽  
Eun Ji Chung ◽  
Tae Suk Suh ◽  
Wonmo Sung
2017 ◽  
Vol 17 (1) ◽  
pp. 85-95 ◽  
Author(s):  
Khalid Iqbal ◽  
Muhammad Mazhar Iqbal ◽  
Muhammad Akram ◽  
Saima Altaf ◽  
Saeed Ahmad Buzdar

AbstractPurposeThis study aimed to examine the dosimetric properties of Gafchromic® EBT3 film and intensity-modulated radiation therapy quality assurance (IMRT QA).Materials and methodsBeams characteristics dosimetric properties and 20 IMRT plans were created and irradiated on Varian dual-energy DHX-S Linac for 6 and 15 MV energies. EBT3 films were analysed using ‘film Pro QA 2014’ software.ResultsThe dosimetric comparison of EBT3 film (for red channel dosimetry) and ionisation ion chamber measurement showed that average deviations of symmetry, flatness, central axis, penumbra (left) and penumbra (right) of dose profile were 0·18, 1·34, 0·49%, 3·68 and 3·61 mm for 6 MV and 0·10, 1·3, 0·45, 2·65 and 2·71 mm for 15 MV, respectively. The blue and green channels dosimetry showed greater dose deviation as compared with red channel. IMRT QA verification plan complied about 95% at all different criteria. Reproducibility, stability and face orientation of film were within 1·4% for red channel.ConclusionsThe results advocate that the film can be used not only for dosimetric assessment but also as a reliable IMRT QA tool.