Joint Meeting of the German Society of Clinical Chemistry and of the German Society of Laboratory Medicine, Rostock, 1619 September 2001

Author(s):  
Rainer Haeckel ◽  
Werner Wosniok ◽  
Thomas Streichert

AbstractThe organizers of the first EFLM Strategic Conference “Defining analytical performance goals” identified three models for defining analytical performance goals in laboratory medicine. Whereas the highest level of model 1 (outcome studies) is difficult to implement, the other levels are more or less based on subjective opinions of experts, with models 2 (based on biological variation) and 3 (defined by the state-of-the-art) being more objective. A working group of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL) proposes a combination of models 2 and 3 to overcome some disadvantages inherent to both models. In the new model, the permissible imprecision is not defined as a constant proportion of biological variation but by a non-linear relationship between permissible analytical and biological variation. Furthermore, the permissible imprecision is referred to the target quantity value. The biological variation is derived from the reference interval, if appropriate, after logarithmic transformation of the reference limits.


Sign in / Sign up

Export Citation Format

Share Document