scholarly journals Preclinical Prototype Development of a Microwave Tomography System for Breast Cancer Detection

ETRI Journal ◽  
2010 ◽  
Vol 32 (6) ◽  
pp. 901-910 ◽  
Author(s):  
Seong-Ho Son
Author(s):  
Alvaro Diaz-Bolado ◽  
Tommy Henriksson ◽  
Paul-Andre Barriere ◽  
Hamidreza Memarzadeh-Tehran ◽  
Nadine Joachimowicz ◽  
...  

2013 ◽  
Vol 2013 ◽  
pp. 1-12 ◽  
Author(s):  
Alvaro Diaz-Bolado ◽  
Paul-Andre Barriere ◽  
Jean-Jacques Laurin

Microwave tomography (MT) measurement setups for different configurations based on breast compression are compared to classical circular measurement setups. Configurations based on compression allow measuring the evanescent component of the scattered field and lead to a compact measurement setup that allows direct image comparison with a standard mammography system. The different configurations are compared based on the singular value decomposition (SVD) of the radiation operator for a 2D TM case. This analysis allows determining under which conditions the image quality obtained from the reconstructions can be enhanced. These findings are confirmed by a series of reconstructions of breast phantoms based on synthetic data obtained at a single frequency of operation.