Background:
We developed a new device for quantifying capillary refill time (CRT) by applying the pulse oximeter principle, and reported the excellent correlation between quantitative CRT (Q-CRT) and tissue hypoperfusion status as represented by arterial blood lactate levels in critically ill patients in the pilot study.
Methods:
Diagnostic accuracy study was undergone in ICU of a tertiary emergency medical center. While the pulse oxygen saturation sensor was placed on the finger of the patients, transmitted light intensity was measured with a pulse oximeter (OLV-3100, Nihon Kohden Corp., Japan) before start and during compression of the finger. Correlation between delta Ab, defined as the absorbance of light intensity of infra-red (A-ir) minus that of red (A-r), and arterial blood lactate level was analyzed.
Results:
The delta Ab was analyzed a total of 69 waveforms in 23 patients. The sensitivity, specificity, positive predictive value, and negative predictive value of the delta Ab less than 0.06 for predicting a lactate level more than 2.0 mmol/l were 100%, 93%, 89%, 100%, respectively.
Conclusions:
The delta Ab was well correlated to high level of blood lactate level in critically ill patients. Further study will confirm to place the delta Ab as one of non-invasive predictors of tissue perfusion status.