Ionization efficiency of a gas-flow ion chamber used for measuring radioactive gases by Monte Carlo simulation

Author(s):  
Tatsuo Torii
2019 ◽  
Vol 25 (2) ◽  
pp. 101-110
Author(s):  
Itumeleng Setilo ◽  
Oluwaseyi Michael Oderinde ◽  
Freek Cp du Plessis

Abstract Introduction: Small fields photon dosimetry is associated with many problems. Using the right detector for measurement plays a fundamental role. This study investigated the measurement of relative output for small photon fields with different detectors. It was investigated for three-photon beam energies at SSDs of 90, 95, 100 and 110 cm. As a benchmark, the Monte Carlo simulation was done to calculate the relative output of these small photon beams for the dose in water. Materials and Methods: 6, 10 and 15 MV beams were delivered from a Synergy LINAC equipped with an Agility 160 multileaf collimator (MLC). A CC01 ion chamber, EFD-3G diode, PTW60019 microdiamond, EBT2 radiochromic film, and EDR2 radiographic film were used to measure the relative output of the linac. Measurements were taken in water for the CC01 ion chamber, EFD-3G diode, and the PTW60019. Films were measured in water equivalent RW3 phantom slabs. Measurements were made for 1 × 1, 2 × 2, 3 × 3, 4 × 4, 5 × 5 and a reference field of 10 × 10 cm2. Field sizes were defined at 100cm SSD. Relative output factors were also compared with Monte Carlo (MC) simulation of the LINAC and a water phantom model. The influence of voxel size was also investigated for relative output measurement. Results and Discussion: The relative output factor (ROF) increased with energy for all fields large enough to have lateral electronic equilibrium (LEE). This relation broke down as the field sizes decreased due to the onset of lateral electronic disequilibrium (LED). The high-density detector, PTW60019 gave the highest ROF for the different energies, with the less dense CC01 giving the lowest ROFs. Conclusion: These are results compared to MC simulation, higher density detectors give higher ROF values. Relative to water, the ROF measured with the air-chamber remained virtually unchanged. The ROFs, as measured in this study showed little variation due to increased SSDs. The effect of voxel size for the Monte Carlo calculations in water does not lead to significant ROF variation over the small fields studied.


2011 ◽  
Vol 26 (3) ◽  
pp. 193-200 ◽  
Author(s):  
Abid Hussain ◽  
Sikander Mirza ◽  
Nasir Mirza ◽  
Muhammad Siddique

GEANT4 based Monte Carlo simulations have been carried out for the determination of efficiency and conversion factors of a gas-flow ionization chamber for beta particles emitted by 86 different radioisotopes covering the average-b energy range of 5.69 keV-2.061 MeV. Good agreements were found between the GEANT4 predicted values and corresponding experimental data, as well as with EGS4 based calculations. For the reported set of b-emitters, the values of the conversion factor have been established in the range of 0.5?1013-2.5?1013 Bqcm-3/A. The computed xenon-to-air conversion factor ratios have attained the minimum value of 0.2 in the range of 0.1-1 MeV. As the radius and/or volume of the ion chamber increases, conversion factors approach a flat energy response. These simulations show a small, but significant dependence of ionization efficiency on the type of wall material.


Sign in / Sign up

Export Citation Format

Share Document