Lambda monoclonal free light chain abnormalities detected by a serum immunofixation electrophoresis assay are underrepresented by quantitative serum free light chain results

2021 ◽  
Vol 156 (Supplement_1) ◽  
pp. S13-S14
Author(s):  
Rebecca Treger ◽  
Kathleen Hutchinson ◽  
Andrew Bryan ◽  
Chihiro Morishima

Abstract Protein and immunofixation (IFIX) electrophoresis are used to diagnose and monitor monoclonal gammopathies. While IFIX detects clonal production of intact immunoglobulins and free light chains (FLC), the latter can also be quantified using a serum free light chain (SFLC) assay, in which polyclonal antisera detects epitopes specific for free kappa (KFLC) or lambda light chains (LFLC). An abnormal KFLC: LFLC ratio (KLR) serves as a surrogate for clonality. While the SFLC assay is highly sensitive, normal LFLC (<2.63mg/dL) and KLR results (>0.26 & <1.65) were found in samples with distinct lambda monoclonal free light chains visualized by IFIX (X-LMFLC). To investigate this discordance, contemporaneous SFLC or KLR values were evaluated for their ability to accurately classify monoclonal FLCs identified by IFIX. We performed a retrospective analysis of serum and urine IFIX (Sebia Hydrasys) and SFLC (Freelite®, Binding Site) results from our institution between July 2010 through December 2020, using R 4.0.2 and Tidyverse packages. From among 9,594 encounters in which a single monoclonal component was initially identified by IFIX, 157 X-LMFLC and 131 X-KMFLC samples were analyzed. Elevated LFLC with normal KFLC was identified in 105/157 X-LMFLC samples (67%), while both LFLC and KFLC were elevated in 42/157 samples (27%). Concordance between X-KMFLC and KFLC was markedly higher, where 122/131 samples (93%) displayed elevated kappa FLC (>1.94mg/dL) with normal LFLC, and only 7/131 X-KMFLC samples (5%) possessed both elevated KFLC and LFLC. The use of KLR to identify pathogenic monoclonal free light chains improved lambda concordance to 85%; however, 19/157 (12%) of X-LMFLC samples still exhibited normal KLR. High concordance of 98% was again observed for X-KMFLC with abnormal KLR. When samples were segregated according to normal or impaired renal function (eGFR > or ≤60mL/min/1.73m², respectively), this disparate identification of X-LMFLC and X-KMFLC by the SFLC assay persisted, suggesting that renal dysfunction (as measured by eGFR) does not underlie this phenomenon. Lastly, we corroborated the above findings in a larger sample population by examining patients with urine Bence Jones FLC identified by IFIX who had free or intact monoclonal components in serum (N=724), grouped by lambda or kappa light chain involvement. The cause(s) of the discrepant performance by the Freelite® SFLC assay, relative to the Sebia Hydrasys IFIX assay, for identifying lambda FLC components is currently unclear. Possible contributory factors include assay reference range cutoffs, other patient disease parameters, and differences in assay-specific polyclonal antisera. Future analyses of these factors will help to further characterize SFLC assay performance and elucidate how interpretation of composite serum FLC test results can be improved to better guide patient management.

Author(s):  
Lauren Campbell ◽  
Dawn Simpson ◽  
Adrian Shields ◽  
Berne Ferry ◽  
Karthik Ramasamy ◽  
...  

Background The measurement of monoclonal free light chains is being increasingly utilized since the introduction of serum-based assays. It is important for laboratories to determine their own reference ranges in order to reflect the local population. The aim of this study was to determine if age-adjusted reference ranges for serum free light chains would have implications for demand management of further laboratory investigations including immunofixation. Methods After certain exclusions, 4293 samples from individuals seen in primary care across Oxfordshire between 2014 and 2016 were identified for analysis of patient characteristics, serum free light chain results and estimated glomerular filtration rate. Results We found age to be an independent variable when considering serum free light chain concentrations, ratio and estimated glomerular filtration rate. The reference ranges derived from our data differ markedly from the original Binding Site ranges. When the age-specific ranges are retrospectively applied to our population, there is a 38% decrease in follow-up testing with no loss of specificity. Conclusion We feel confident implementing new age-specific serum free light chain reference ranges in our laboratory. We have developed a simple algorithm for evaluating serum free light chains based on age and estimated glomerular filtration rate. We encourage laboratories to establish their own local reference ranges using large cohorts and their chosen serum free light chain assay platform.


Blood ◽  
2001 ◽  
Vol 97 (9) ◽  
pp. 2900-2902 ◽  
Author(s):  
Mark Drayson ◽  
Lian X. Tang ◽  
Roger Drew ◽  
Graham P. Mead ◽  
Hugh Carr-Smith ◽  
...  

Abstract Using sensitive, automated immunoassays, increased concentrations of either κ or λ free light chains (and abnormal κ/λ ratios) were detected in the sera of 19 of 28 patients with nonsecretory multiple myeloma. Four other patients had suppression of one or both light chains, and the remaining 5 sera had normal or raised free light-chain concentrations with substantially normal κ/λ ratios. Six of the patients with an elevated single free light chain, who were studied during follow-up, had changes in disease activity that were reflected by the changes in free light-chain concentrations. It is concluded that quantification of free light chains in serum should prove useful for the diagnosis and monitoring of many patients with nonsecretory myeloma.


Author(s):  
Richard B Fulton ◽  
Suran L Fernando

Background The potential for serum free light chain (sFLC) assay measurements to replace urine electrophoresis (uEPG) and to also diminish the need for serum immunofixation (sIFE) in the screening for monoclonal gammopathy was assessed. A testing algorithm for monoclonal protein was developed based on our data and cost analysis. Methods Data from 890 consecutive sFLC requests were retrospectively analysed. These included 549 samples for serum electrophoresis (sEPG), 447 for sIFE, and 318 for uEPG and urine immunofixation (uIFE). A total of 219 samples had sFLC, sEPG, sIFE and uEPG + uIFE performed. The ability of different test combinations to detect the presence of monoclonal proteins was compared. Results The sFLC κ/ λ ratio (FLC ratio) indicated monoclonal light chains in 12% more samples than uEPG + uIFE. The combination of sEPG and FLC ratio detected monoclonal proteins in 49% more samples than the combination of sEPG and sIFE. Furthermore, the sEPG + FLC ratio combination detected monoclonal protein in 6% more samples than were detected by the combined performance of sEPG, sIFE, uEPG and uIFE. However, non-linearity of the assay, the expense of repeat determinations due to the narrow measuring ranges, and frequent antigen excess checks were found to be limitations of the sFLC assay in this study. Conclusion The FLC ratio is a more sensitive method than uIFE in the detection of monoclonal light chains and may substantially reduce the need for onerous 24 h urine collections. Our proposed algorithm for the evaluation of monoclonal gammopathy incorporates the sFLC assay, resulting in a reduction in the performance of labour intensive sIFE and uEPG + uIFE while still increasing the detection of monoclonal proteins.


2013 ◽  
Vol 2013 ◽  
pp. 1-6 ◽  
Author(s):  
Aurélie Grados ◽  
Mikael Ebbo ◽  
José Boucraut ◽  
Frédéric Vély ◽  
Pierre Aucouturier ◽  
...  

Immunoglobulin free light chains are produced in excess during normal antibody synthesis. Their evaluation is commonly used in case of a monoclonal gammopathy. In polyclonal hypergammaglobulinemia related to the Sjögren syndrome or systemic lupus, erythematosus serum free light chain levels are increased and could correlate with disease activity. We show here that theκ() andλ() free light chains and theκ : λratio () are increased in sixteen patients with IgG4-related disease when compared to healthy controls. The increase ofκandλfree light chains probably reflects the marked polyclonal B cell activation of the disease. We could not assess in this small cohort of patients a significative correlation of serum free light chain levels and disease activity or extension.


2013 ◽  
Vol 66 (11) ◽  
pp. 992-995
Author(s):  
Annalise E Zemlin ◽  
Megan A Rensburg ◽  
Hayley Ipp ◽  
Jurie J Germishuys ◽  
Rajiv T Erasmus

Monoclonal serum free light chain measurements are used to follow up and manage patients with monoclonal gammopathies, and abnormal serum free light chain ratios are associated with risk of progression in certain diseases. We aimed to validate the reference intervals in our population. Reference intervals for κ and λ free light chains were established on 120 healthy adults. Creatinine levels were measured to exclude renal dysfunction and serum protein electrophoresis was performed. All creatinine values were within normal limits. After exclusion of subjects with abnormal serum protein electrophoreses, 113 subjects were available for analysis. The 95% reference interval was 6.3–20.6 mg/L for κ free light chains, 8.7–25.9 mg/L for λ free light chains and 0.46–1.23 for free light chain ratio. Most of the values fell within the manufacturer's recommended limits and therefore could be used for our population.


Author(s):  
Henk te Velthuis ◽  
Mark Drayson ◽  
John P. Campbell

AbstractRecently, serum free light chain (FLC) assays incorporating anti-kappa (κ) and anti-lambda (λ) FLC monoclonal antibodies have become available: N Latex FLC assay (Siemens) and Seralite


2020 ◽  
Vol 5 (6) ◽  
pp. 1358-1371
Author(s):  
Gurmukh Singh

Abstract Background Laboratory methods for diagnosis and monitoring of monoclonal gammopathies have evolved to include serum and urine protein electrophoresis, immunofixation electrophoresis, capillary zone electrophoresis, and immunosubtraction, serum-free light chain assay, mass spectrometry, and newly described QUIET. Content This review presents a critical appraisal of the test methods and reporting practices for the findings generated by the tests for monoclonal gammopathies. Recommendations for desirable practices to optimize test selection and provide value-added reports are presented. The shortcomings of the serum-free light chain assay are highlighted, and new assays for measuring monoclonal serum free light chains are addressed. Summary The various assays for screening, diagnosis, and monitoring of monoclonal gammopathies should be used in an algorithmic approach to avoid unnecessary testing. Reporting of the test results should be tailored to the clinical context of each individual patient to add value. Caution is urged in the interpretation of results of serum-free light chain assay, kappa/lambda ratio, and myeloma defining conditions. The distortions in serum-free light chain assay and development of oligoclonal bands in patients‘ status post hematopoietic stem cell transplants is emphasized and the need to note the location of original monoclonal Ig is stressed. The need for developing criteria that consider the differences in the biology of kappa and lambda light chain associated lesions is stressed. A new method of measuring monoclonal serum-free light chains is introduced. Reference is also made to a newly defined entity of light chain predominant intact immunoglobulin monoclonal gammopathy. The utility of urine testing in the diagnosis and monitoring of light chain only lesions is emphasized.


Blood ◽  
2016 ◽  
Vol 128 (22) ◽  
pp. 5698-5698
Author(s):  
Abhishek Chilkulwar ◽  
Prerna Mewawalla ◽  
Anna Miller ◽  
Gina Berteotti ◽  
Entezam Sahovic ◽  
...  

Abstract Background: There have been significant improvements in the outcomes for patients with multiple myeloma over the past 10 years. The course of the disease remains highly variable. While some patients experience 10-15 year survivals, others succumb to highly refractory disease within a few months. Many studies have focused on the description of prognostic factors capable of predicting this heterogeneity in survival. Baseline Serum free light chain concentration is a major prognostic indicator for plasma cell neoplasms and normalization of free light chain ratio with treatment has been reported as an indicator for favorable prognosis. High Baseline free light chain concentration at presentation indicates aggressive disease. Furthermore High serum free light chain ratio correlates with elevated serum creatinine, elevated LDH and extensive marrow infiltration. Objectives: To determine if serum free light chain concentration > 1000mg/dl at the time of diagnosis and at disease progression was an independent prognostic marker for multiple myeloma. Methods: The results of all the serum free light chain analyses of patients evaluated at western Pennsylvania hospital between 2007 -2015 were reviewed and patients with serum free light chain concentration >1000mg/dl at the time of initial diagnosis and disease progression were identified Retrospective chart review was done to study the survival in these patients. Results: Total of 15 patients in whom serum free light chain concentration was greater than 1000 mg/dl at diagnosis were identified. Median age at diagnosis was 60.81 years (45.2-77.3). The median survival in this population is 1.85 years (range 0.06-8.85 years), with 5 deaths. Of these 15, 9 patients received an autologous PBSCT as a part of their initial therapy. Six of them are alive with a median overall survival of 2.44 years and mean survival of 3.89 years (range 0.87-8.89). In the remaining 6 patients that did not undergo an auto PBSCT 3 of them are alive with median overall survival of 1.21 years and a mean survival 1.20 years (range 06-2.76 years). There is an overall trend toward very poor prognosis in this specific group of patients irrespective of the age, sex, stage at time of diagnosis, the immunoglobulin subtype and the free light chain subtype (kappa vs lambda). We also conducted a retrospective review of patients in whom serum free light chains concentration was greater than 1000mg/dl at disease progression. Twenty patients were identified all of whom are deceased with a median overall survival from the time of progression with serum free light chains greater than 1000 mg /dl was dismal at 0.27 years and mean survival was 0.53 years (range 0.04-2.52 years) Conclusion: Patients with multiple myeloma who presented with a serum free light chain concentration greater than 1000 mg/dl at diagnosis have very poor prognosis. These specific subgroups of patients should be identified and treated aggressively at the time of diagnosis to prevent complications from multiple myeloma and improve their overall survival. In our small cohort of patients those who underwent auto PBSCT as a part of their initial therapy tend to have better outcomes in terms of overall survival compared to patients who received therapy without auto PBSCT consolidation. Patients with serum free light chains greater than 1000 mg/dl at disease progression will need to be treated aggressively with perhaps a salvage auto PBSCT to improve their overall outcome. A larger study to evaluate elevated serum free light chains (greater than 1000 mg/dl) as a prognostic indicator at diagnosis and progression should be done to further validate these findings Table 1 Table 1. Disclosures No relevant conflicts of interest to declare.


Blood ◽  
2004 ◽  
Vol 104 (11) ◽  
pp. 4880-4880 ◽  
Author(s):  
Bradley M. Augustson ◽  
Steven D. Reid ◽  
Graham P. Mead ◽  
Mark T. Drayson ◽  
J. Anthony Child ◽  
...  

Abstract Introduction: Patients with asymptomatic myeloma fulfil two of the diagnostic criteria for myeloma having more than 10% bone marrow plasma cells and an M protein of greater than 30g/l, but they are asymptomatic with no evidence of end organ or tissue damage. The median time to disease progression is 12–32 months. These patients do not require treatment but do require monitoring for progression to symptomatic myeloma. Predicting progression of asymptomatic myeloma would be of clinical benefit to optimise monitoring and initiate treatment prior to substantial end organ damage. However monoclonal spike, plasma cell labelling index, bone marrow plasmacytosis, immunoparesis and the presence of Bence Jones protein have limited value in predicting progression. Abnormal levels of serum free light chains are present in 95% of all multiple myeloma patients and have clinical benefit in diagnosis and monitoring of disease. In monoclonal gammopathy of undetermined significance (MGUS) 60% of patients have abnormal serum free light chain ratios and are an independent risk factor for progression to myeloma. The aim of this study was to examine the serum of asymptomatic patients for serum free light chains at diagnosis and to determine if they are predictive of disease progression. Methods: Archived presentation sera were studied from forty three asymptomatic myeloma patients who had been registered into United Kingdom Medical Research Council trials (1980 – 2002). Archived presentation sera were assayed for serum free light chains using the serum free light chain assay on an Olympus AU400 analyzer. Times to progression for those with abnormal versus normal serum free light chain ratios were compared. Times to progression were examined by Kaplan-Meier survival curves and log-rank sum statistical analysis. Results: Abnormal serum free light ratios were present in 36/43 (84%) of asymptomatic myeloma patients at the time of diagnosis and the remaining 7 patients had normal ratios. The median follow-up time for all 43 patients was 2807 days. Six patients with a normal kappa/lambda ratio had a median time to progression of 1323 days. In contrast, 26 patients with abnormal serum free kappa/lambda ratios had a median time to progression of 713 days. Ten patients who had an abnormal kappa/lambda ratio had not progressed at the time of follow-up. Although the median time to progression of patients with normal serum free light chain ratios was greater than those with abnormal ratios, this did not reach statistical significance (p<0.13). Conclusions: In summary, 84% of asymptomatic myeloma patients have an abnormal kappa/lambda ratio at diagnosis, in comparison with 95% of multiple myeloma and 60 % of (MGUS) patients. Furthermore, our data suggest that those with normal serum free light chain ratio may progress more slowly than those with abnormal ratios. Due to the small number of patients in this study, this did not reach statistical significance. In the spectrum of malignancy from MGUS to asymptomatic and symptomatic myeloma serum free light chain levels have an increasing frequency of abnormality and are associated with increased risk of progression to symptomatic myeloma.


Blood ◽  
2012 ◽  
Vol 120 (21) ◽  
pp. 4568-4568 ◽  
Author(s):  
Katerina Sarris ◽  
Vassiliki Bartzis ◽  
Dimitris Maltezas ◽  
Efstathios Koulieris ◽  
Tatiana Tzenou ◽  
...  

Abstract Abstract 4568 Background and Aim: Symptomatic CLL patients need treatment immediately. For these patients, molecular-genetic factors (mutated-unmutated, ZAP 70, ATM, p53) are important prognostic factors of response and survival. Nevertheless, 2/3 of newly diagnosed patients are asymptomatic and require only of follow up that can last for months or years. For these patients overall survival (OS) depends on the time to first treatment (TFT). The most frequent paraprotein produced in CLL is serum free light chain in 50% of the patients. It has recently been shown that serum free light chains (sFLC) and their sum above 60 (κ+λ above 60) are useful prognostic factors for TFT. We therefore studied the eventual prognostic implication of sFLC and the summated FLC-kappa plus FLC-lambda in a CLL patients' series. Patients and Methods: 143 CLL patients were studied of which 18 needed immediate treatment while 37 more needed treatment during their follow up. 64% and 72%, 28% and 18%, 7.5% and 10%, were in stage 0 and A, 1 and β, 2 and C according to Rai and Binet respectively. Median patients' follow up was 32 months (range 4–228). Light chain restriction was established by flow cytometry or bone marrow biopsy immunohistochemistry. Serum free light chain values were retrospectively determined by nephelometry (Freelite™, the Binding Site Birmingham, UK) in frozen sera drawn at diagnosis. Elevated sFLC values were defined using as cut-off values the 95th percentile range of healthy individuals. Statistical analysis was performed using SPSS v15.0. Hazard ratios and prognostic significance of abnormal sFLC, HLC and ratios were determined by univariate Cox regression analysis. Kaplan Meier method was used for pictorial representation of survival and time to treatment. Results: Increased sFLC were found in 45% of the patients while the summated FLC-kappa plus FLC-lambda was higher than 60 mg/dl in 14%. Increased sFLC values as well as values of FLC κ+λ>60 were related to shorter TFT (p=0,0005 and p=0,000003 respectively). In addition, high levels of sFLC and FLC κ+λ >60 correlated with β2-microglobulin (r=0.2, p=0.009 και r=0.2, p=0.03 respectively), serum albumin (r=0.2, p=0.009 only for FLC κ+λ > 60), negatively with hemoglobin (r=-0.3, p=0.000003 και r=-0.2, p=0.0002 respectively), increased LDH (p=0.037 και 0.001 respectively), Rai stage (p=0.03 και 0.003 respectively) and Binet stage (p=0.02 only for FLC κ+λ > 60) and with the presence of beta-symptoms (p=0.004 only for FLC κ+λ > 60). Finally, increased sFLC and FLC κ+λ>60 values correlated with shorter OS (p=0.05 and p=0.003 respectively). Conclusion: The results of our study confirmed the significance of sFLC in CLL with regard to TFT and their relationship with adverse prognostic clinical and laboratory parameters but also demonstrated for the first time their impact on OS. Disclosures: No relevant conflicts of interest to declare.


Sign in / Sign up

Export Citation Format

Share Document