Gap plasmon-based fluorescence correlation spectroscopy for anomalous molecular diffusion

Author(s):  
Hongki Lee ◽  
Donghyun Kim ◽  
Gwiyeong Moon ◽  
Seongmin Im
2017 ◽  
Author(s):  
Dominic Waithe ◽  
Falk Schneider ◽  
Jakub Chojnacki ◽  
Mathias Clausen ◽  
Dilip Shrestha ◽  
...  

AbstractScanning Fluorescence Correlation Spectroscopy (scanning FCS) is a variant of conventional point FCS that allows molecular diffusion at multiple locations to be measured simultaneously. It enables disclosure of potential spatial heterogeneity in molecular diffusion dynamics and also the acquisition of a large amount of FCS data at the same time, providing large statistical accuracy. Here, we optimize the processing and analysis of these large-scale acquired sets of FCS data. On one hand we present FoCuS-scan, scanning FCS software that provides an end-to-end solution for processing and analysing scanning data acquired on commercial turnkey confocal systems. On the other hand, we provide a thorough characterisation of large-scale scanning FCS data over its intended time-scales and applications and propose a unique solution for the bias and variance observed when studying slowly diffusing species. Our manuscript enables researchers to straightforwardly utilise scanning FCS as a powerful technique for measuring diffusion across a broad range of physiologically relevant length scales without specialised hardware or expensive software.