Application of the high-speed self-reversal background corrector to the determination of cadmium by chemical vapor generation atomic absorption spectrometry

2012 ◽  
Vol 90 (10) ◽  
pp. 874-879 ◽  
Author(s):  
Christophe Waterlot ◽  
Aurélie Pelfrêne ◽  
Francis Douay

Concentrations of cadmium (Cd) in extracting solutions (neutral salts) from contaminated soils are often too low to be determined by conventional flame atomic absorption spectrometry. For this reason, determination of Cd requires sensitive analytical methods free from interference problems generated by samples. In this context, vapor generation atomic absorption spectrometry (HGAAS) was combined with a high-speed self-reversal background corrector. This new approach was successfully applied after optimization of the analytical parameters to obtain a maximal absorbance signal of the volatile Cd species. The optimum condition was achieved with a 3% (m/v) NaBH4 in 1.5% (m/v) NaOH reducing solution and a solution containing 0.3 mol/L HNO3. The detection limit was 1 ng mL–1 under the previous conditions and the relative standard deviation was up to 5% for 10 replicate analyses of Cd at 0.2 and 1 ng mL–1, reflecting a very highly sensitive and reproducible method. Moreover, the results showed that the proposed combination was an efficient method to overcome the interference problems caused by different coexisting cations, As, Al, Ca, Cu, Fe, Mg, Mn, Ni, Pb, Se and Zn, up to 10 µg mL–1. The method was validated with analyses of two standard reference materials and was used for Cd determination in 0.01 mol/L CaCl2 extracts from contaminated kitchen garden soils. The data were compared with those obtained by two other more conventional methods, electrothermal atomic absorption spectrometry (ETAAS) and flame atomic absorption spectrometry (FAAS). The analytical results obtained by the ETAAS and HGAAS were in a good agreement, suggesting the suitability of the method for Cd determination in 0.01 mol/L CaCl2 extracting solution.

2015 ◽  
Vol 98 (5) ◽  
pp. 1417-1422 ◽  
Author(s):  
Ali Behbahani ◽  
Mehdi Ardjmand

Abstract A new method was applied to produce modified nanoporous silica as a novel sorbent for Pb(II) ion SPE from industrial wastewater samples. In this modified method, the produced nanoporous silica has a higher functional group loading, which leads to a higher preconcentration factor as well as a lower LOD. This modified nanoporous silica was used for preconcentration prior to subsequent determination of Pb(II) ions by flame atomic absorption spectrometry. Various parameters such as the eluent, pH of the sample solution, and flow rate were optimized during this work. Also, the effect of a variety of ions on preconcentration and recovery of Pb(II) ions was investigated. The LOD, defined as five times the SD of the blank, was determined to be lower than 0.1 mg/L with an RSD of <2%. The accuracy of the method was established by analyzing standard reference materials with certified Pb concentrations. Finally, the established method was successfully applied for determination of the Pb(II) ion concentration in industrial wastewater samples.


Sign in / Sign up

Export Citation Format

Share Document