scholarly journals Determination of Chlorophenol Derivatives Using the Homogeneous Liquid-liquid Extraction in Ternary Component System-GC/MS Method.

2000 ◽  
pp. 291-293 ◽  
Author(s):  
Yoshitaka TAKAGAI ◽  
Chiaki MAEKOYA ◽  
Shukuro IGARASHI
2009 ◽  
Vol 6 (4) ◽  
pp. 1077-1084 ◽  
Author(s):  
Mohammad Reza Jamali ◽  
Yaghoub Assadi ◽  
Reyhaneh Rahnama Kozani ◽  
Farzaneh Shemirani

A simple and effective homogeneous liquid-liquid extraction method for selective separation, preconcentration and spectrophotometric determination of palladium(II) ion was developed by using a ternary component system (water / tetrabutylammonium ion (TBA+) / chloroform). The phase separation phenomenon occurred by an ion–pair formation of TBA+and perchlorate ion. Thio-Michler’s ketone (TMK), 4, 4ˊ-bis (dimethylamino) thiobenzophenone, was used as a complexing agent. After optimization of complexation and extraction conditions ([TMK]=5.0x10-2mol L-1, [TBA+] = 2.0×10-2mol L-1, [CHCl3] = 60.0 µL, [ClO4-] = 2.5×10-2mol L-1and pH= 3.0), a preconcentration factor 10 was obtained for 10 mL of sample. The analytical curve was linear in the range of 2-100 ng mL-1and the limit of detection was 0.4 ng mL-1. The relative standard deviation was 3.2% (n=10). Accuracy and application of the method was estimated by using test samples of natural and synthetic water spiked with different amounts of palladium(II) ion. The method is very simple and inexpensive.