scholarly journals Reproducibility of Central Corneal Thickness Measurements in Normal Eyes Using the Zeiss Cirrus 5000 HD-OCT and Pentacam HR

2018 ◽  
Vol 12 (1) ◽  
pp. 72-83 ◽  
Author(s):  
Elmira Baghdasaryan ◽  
Xiwen Huang ◽  
Kenneth M. Marion ◽  
Tudor C. Tepelus ◽  
Homayoun Bagherinia ◽  
...  

Objectives:To determine the repeatability and reproducibility of Central Corneal Thickness (CCT) measurements using two different anterior segment imaging modalities, including those obtained with the new anterior segment lens attachments for the Cirrus 5000 HD-OCT.Methods:A total of 32 eyes from 16 normal volunteers (8 male, 8 female) were enrolled in this prospective study. CCT was measured by the same examiner using the Cirrus 5000 HD-OCT and Pentacam HR. The results of CCT obtained by each method were averaged and compared using t-test analysis. The agreement between the measurement methods was evaluated. Coefficient of Repeatability (CoR) and Intra-Class Correlation Coefficient (ICC) were computed.Results:The mean measurements taken with the Cirrus OCT anterior chamber lens (CCTAC), HD cornea lens (CCTHDC) and pachymetry scans (CCTPach) were 545.35 ± 31.02, 537.87 ± 26.82, and 532.04 ± 29.82 µm, respectively. The mean CCT obtained with the Pentacam (CCTPent) was 545.51 ± 30.71 µm. CCTPentwere significantly higher than CCTHDCand CCTPach(p< 0.0001). In contrast, the CCTPentand CCTACwere similar (p=0.87). CCT, as evaluated by the two different instruments, showed excellent correlation (r> 0.98,p< 0.0001) with an ICC > 0.99 (95% CI, 0.97 – 0.99). CoR was the highest for CCTPach(3.7 ± 1.4, 95% CI (3.0- 4.6)).Conclusion:CCT measurements from the Cirrus OCT using the new anterior segment lens attachments and the Pentacam HR are highly correlated. This should allow the use of a standardized correction factor if necessary to inter-relate the measurements between the two devices.

2021 ◽  
Vol 13 ◽  
pp. 251584142199563
Author(s):  
Mehmet Barış Üçer ◽  
Erdinç Bozkurt

Purpose: The purpose of this study was to compare and evaluate the agreement of central corneal thickness (CCT) values obtained with three different devices working according to optical principle in healthy eyes. Methods: 60 eyes of 60 individuals (30 men and 30 women) were enrolled in this study. CCT measurements performed with Scheimpflug–Placido topographer (Sirius), spectral-domain optical coherence tomography (RTVue) with an anterior segment module, and optical biometer (AL-Scan) were compared. Bland–Altman analysis was used to demonstrate agreement between methods. Results: The mean age was 30.07 ± 7.313 years (range, 18–47 years). The mean CCT values obtained by RTVue, Sirius, and AL-Scan were 518.25 ± 36.38 µm, 526.08 ± 36.33 µm, and 513.50 ± 39.09 µm, respectively. The mean differences in CCT were 7.83 ± 14.15 µm between Sirius and RTVue, 12.58 ± 11.87 µm between Sirius and AL-Scan, and 4.75 ± 4.50 µm between RTVue and AL-Scan. The mean CCT was statistically different among the three groups ( p < 0.05). All three modalities of CCT measurements correlated closely with each other, with Pearson’s correlation coefficients ranging from 0.924 to 0.961. The 95% limits of agreement were −19.90 to 35.56 µm between Sirius and RTVue, −10.69 to 35.85 µm between Sirius and AL-Scan, and −4.07 to 13.58 µm between RTVue and AL-Scan. Conclusion: Different results could be obtained through different noncontact devices in CCT measurements. Although the measurement values obtained by these devices show a high level of correlation, it would be a more correct approach to not use them directly interchangeably in clinical practice. Evaluation and follow-up of CCT should be performed using the same device.


Author(s):  
Sagili Chandrasekhara Reddy ◽  
Mohd Mansor Shariff ◽  
Aina Malindri Dasrilsyah

Purpose: To determine the central corneal thickness in myopic adult patients scheduled for laser corneal refractive surgery and to explore its correlation with degree of refractive error. Materials and Methods: The case records of 130 myopic patients who underwent laser corneal refractive surgery in a military hospital over a period of two years were reviewed to determine the central corneal thickness. All patients had 6/6 vision with best correction, and did not have any other anterior segment or fundus diseases in both eyes. The central corneal thickness was measured with Visante Carl Zeiss anterior segment optical coherence tomography instrument.                                                                                                                        Results: Out of 130 patients, males were more (73, 56.2%); mean age of patients was 33.8 years (range 18-60 years) and majority were Malays (110, 84.6%).  The spherical power of myopia ranged from – 0.5 to – 10.00 D, and the cylindrical power ranged from – 0.25 to – 3.25 D. The mean central corneal thickness of both eyes was 528.2 µm (range 331- 615 µm); in the mild degree of myopia (- 0.50 to - 2.00 D) 527.9 µm, moderate degree (- 2.25 to -5.00 D) 529.4 µm, and high degree (-5.25 to -11.00 D) 523.9 µm. Conclusion: The anterior segment optical coherence tomography provides noncontact, rapid, pachymetry mapping of the corneal thickness. In Malaysian patients, the mean central corneal thickness of both eyes in myopia was 528.1 µm (range 331- 615 µm). There was no correlation between the mean central corneal thickness and degree of myopia, different genders, age groups, ethnic groups and two eyes.


2017 ◽  
Vol 2017 ◽  
pp. 1-6 ◽  
Author(s):  
Katarzyna Krysik ◽  
Dariusz Dobrowolski ◽  
Katarzyna Polanowska ◽  
Anita Lyssek-Boron ◽  
Edward A. Wylegala

Purpose. Comparative analysis of central and peripheral corneal thickness in PEX patients using three different imaging systems: Pentacam-Scheimpflug device, time-domain optical coherence tomography (OCT) Visante, and swept-source OCT Casia. Materials and Methods. 128 eyes of 80 patients with diagnosed PEX were examined and compared with 112 normal, non-PEX eyes of 72 cataract patients. The study parameters included 5 measured zones: central and 4 peripheral (superior, inferior, nasal, and temporal). Results. The mean CCT in eyes with PEX syndrome measured with all three instruments was thicker than that in normal eyes. Corneal thickness measurements in the PEX group were statistically significantly different between Pentacam and OCT Casia: central corneal thickness (p=0.04), inferior corneal zone (p=0.01), and nasal and temporal corneal zones (p<0.01). Between Pentacam and OCT Visante inferior, nasal and temporal corneal zones were statistically significantly different (p<0.01). Between OCT Casia and OCT Visante, there were no statistically significant differences in measured parameters values. Conclusion. The central corneal thickness in eyes with PEX syndrome measured with three different independent methods is higher than that in the non-PEX group, and despite variable peripheral corneal thickness, this one parameter is still crucial in intraocular pressure measurements.


2017 ◽  
Vol 26 (10) ◽  
pp. 860-865 ◽  
Author(s):  
Riccardo Scotto ◽  
Alessandro Bagnis ◽  
Marina Papadia ◽  
Carlo Alberto Cutolo ◽  
Domenico Risso ◽  
...  

2021 ◽  
Vol 37 (3) ◽  
Author(s):  
Muhammad Suhail Sarwar ◽  
Sehrish Shahid ◽  
Muhammad Arslan Ashraf ◽  
Shaista Kanwal

Purpose:  To check inter observer reproducibility of axial ocular measurements i.e. central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), anterior segment lens (ASL), vitreous length (VL) and axial length (AL) with non-contact HAAG-STRAIT biometer. Study Design:  Comparative Reproducibility Analysis. Place and Duration of Study:  College of ophthalmology and allied vision sciences (COAVS), Mayo Hospital, Lahore. Methods:  This study included 66 healthy students (132 eyes). Data was collected through self-designed proforma by 2 operators independently. SPSS 21 was used for data analysis. Interclass correlation was applied for agreement between the two readings. Interclass coefficient (ICC) value greater than 0.7 was considered as excellent correlation. Results:  The mean CCT, AD, ACD, LT, ASL, VL, and AL were 526.47 ± 35.72 µm and 526.47 ± 36.06 µm (ICC = 0.92); 2.93 ± 0.29 mm and 2.93 ± 0.29 mm (ICC = 0.81); 3.45 ± 0.30 mm and 3.46 ± 0.30 mm (ICC = 0.79); 3.58 ± 0.28 mm and 3.56 ± 0.22 mm (ICC = 0.76); 7.03 ± 0.30 mm and 7.02 ± 0.27 mm (ICC = 0.80); 16.56 ± 0.85 mm and 16.62 ± 0.81 mm (ICC = 0.72); and 23.59 ± 0.85 mm and 23.64 ± 0.87 mm (ICC: 0.76) of observer 1 and 2, respectively. Conclusion:  Non-contact Biometer (HAAG-STRAIT) has high inter-observer reproducibility with strong interclass coefficient of greater than 0.72. Key Words:  Biometry, Axial Length, Central Corneal Thickness, Anterior Chamber Depth.


2016 ◽  
Vol 2016 ◽  
pp. 1-5
Author(s):  
Jinhai Huang ◽  
Weicong Lu ◽  
Giacomo Savini ◽  
Hao Chen ◽  
Chengfang Wang ◽  
...  

Purpose.To compare between a new optical biometer (AL-Scan, Nidek Co., Aichi, Japan) and an anterior segment optical coherence tomographer (Visante AS-OCT, Carl Zeiss Meditec, Dublin, USA) for measuring central corneal thickness (CCT), anterior chamber depth (ACD), and aqueous depth (AD).Methods.Sixty-three eyes of 63 normal subjects were examined with AL-Scan and Visante AS-OCT in this prospective study. One eye per subject was measured three times with both devices to record their CCT, ACD, and AD. All procedures were performed by the same operator. Agreement between the two devices was assessed using pairedt-tests, Bland-Altman plots, and 95% limits of agreement (LoA).Results.The mean CCT, ACD, and AD measured by AL-Scan were538.59±27.37 μm,3.70±0.30 mm, and3.16±0.30 mm, respectively. The mean values obtained by the Visante OCT were536.14±26.61 μm for CCT,3.71±0.29 mm for ACD, and3.17±0.29 mm for AD. The mean CCT by the AL-Scan was higher than that obtained by the Visante AS-OCT (difference =2.45±6.07 μm,P<0.05). The differences in ACD and AD measurements were not statistically significant. The 95% LoA of CCT, ACD, and AD were between −9.44 and 14.35 μm, −0.15 and 0.12 mm, and −0.15 and 0.12 mm, respectively.Conclusions.Since these two devices were comparable for measuring CCT, ACD, and AD, their results can be interchangeably used in the clinic.


2015 ◽  
Vol 2015 ◽  
pp. 1-6 ◽  
Author(s):  
Jinhai Huang ◽  
Giacomo Savini ◽  
Chengfang Wang ◽  
Weicong Lu ◽  
Rongrong Gao ◽  
...  

Purpose.To assess the reliability and comparability of measuring central corneal thickness (CCT) and thinnest corneal thickness (TCT) using a new Scheimpflug-Placido analyzer (TMS-5, Japan) and ultrasound (US) pachymetry.Methods.Seventy-six healthy subjects were prospectively measured 3 times by 1 operator using the TMS-5, 3 additional consecutive scans were performed by a second operator, and ultrasound (US) pachymetry measurements were taken. The test-retest repeatability (TRT), coefficient of variation (CoV), and intraclass correlation coefficient (ICC) were calculated to evaluate intraoperator repeatability and interoperator reproducibility. Agreement among the devices was assessed using Bland-Altman plots and 95% limits of agreement (LoA).Results.The intraoperators TRT and CoV were <19 μm and 2.0%, respectively. The interoperators TRT and CoV were <12 μm and 1.0%, respectively, and ICC was >0.90. The mean CCT and TCT measurements using the TMS-5 were 15.97 μm (95% LoA from −26.42 to −5.52 μm) and 20.32 μm (95% LoA from −30.67 to −9.97 μm) smaller, respectively, than those using US pachymetry.Conclusions.The TMS-5 shows good repeatability and reproducibility for measuring CCT and TCT in normal subjects but only moderate agreement with US pachymetry results. Caution is warranted before using these techniques interchangeably.