Combination of knotted reactor with portable tungsten coil electrothermal atomic absorption spectrometer for on-line determination of trace cadmium

2016 ◽  
Vol 124 ◽  
pp. 60-64 ◽  
Author(s):  
Xiaodong Wen ◽  
Shengchun Yang ◽  
Haizhu Zhang ◽  
Qingwen Deng
1998 ◽  
Vol 52 (8) ◽  
pp. 1086-1091 ◽  
Author(s):  
James D. Batchelor ◽  
Susan E. Thomas ◽  
Bradley T. Jones

A portable atomic absorption spectrometer is described that is powered by a 12 V car battery. The atomization device is a tungsten coil extracted from a common projector lamp bulb. The coil is housed in a glass cell, and the atomization environment is purged with 10% H2 in Ar (available as a standard commercial welding gas) to prevent oxidation. The detection system is a miniature charge-coupled device (CCD) spectrometer mounted on a PC board. The device is controlled by a laptop computer. Background correction measurements are performed by the near-line method. Aqueous standard solutions provide a Cd detection limit of 3 μg/L (60 pg) for a 20 μL sample volume. Accuracies for the determination of Cd in urine, soil, and stream water are typically better than 90%, with relative standard deviations in the 10% range.