A simple and fast procedure for the determination of Al, Cu, Fe and Mn in biodiesel using high-resolution continuum source electrothermal atomic absorption spectrometry

2011 ◽  
Vol 66 (5) ◽  
pp. 373-377 ◽  
Author(s):  
Daiane P.C. Quadros ◽  
Martinho Rau ◽  
Muhammad Idrees ◽  
Eduardo S. Chaves ◽  
Adilson J. Curtius ◽  
...  
2017 ◽  
Vol 37 (1) ◽  
pp. 45
Author(s):  
Tiago Varão Silva ◽  
Jardes Figuerêdo Rêgo ◽  
Mercedes De Moraes ◽  
Alberto José Cavalheiro ◽  
José Anchieta Gomes Neto ◽  
...  

A simple and rugged method to determine Ca, K and Mg in a single aliquot of medicinal plants by high-resolution continuum source flame atomic absorption spectrometry is proposed. Secondary lines for Ca (239.856 nm) and K (404.414 nm), and the alternate line measured at wing of the secondary line for Mg at 202.588 nm allowed calibration within the 20 – 500 mg L-1 Ca and K, and 1.0 – 80 mg L-1 Mg. Twenty samples and three plant certified materials were analyzed. Results were in agreement at a 95% confidence level with reference values. Limits of detection were 2.4 mg L-1 Ca, 1.9 mg L-1 K and 0.3 mg L-1 Mg. The RSD (n=12) were ≤ 5.1% and recoveries were between 83 and 108% for all analytes.