MO-D-AUD-06: Dependence of Total Scatter Factors of Small Beams On the Radial Distribution of the Electron Beam Incident On the Target: A Multi-Detector and Monte Carlo Study

2007 ◽  
Vol 34 (6Part15) ◽  
pp. 2521-2522
Author(s):  
P Francescon ◽  
S Cora ◽  
C Cavedon ◽  
V Santangelo ◽  
P Scalchi
2015 ◽  
Vol 27 (1) ◽  
pp. 14001
Author(s):  
孙剑锋 Sun Jianfeng ◽  
胡杨 Hu Yang ◽  
孙江 Sun Jiang ◽  
杨海亮 Yang Hailiang ◽  
张鹏飞 Zhang Pengfei ◽  
...  

2016 ◽  
Vol 22 (1) ◽  
pp. 5-9
Author(s):  
Samuel Okon Inyang ◽  
Alan Chamberlain

Abstract The use of a dual electron multileaf collimator (eMLC) to collimate therapeutic electron beam without the use of cutouts has been previously shown to be feasible. Further Monte Carlo simulations were performed in this study to verify the nature and appearance of the isodose distribution in water phantom of irregular electron beams delivered by the eMLC. Electron fields used in this study were selected to reflect those used in electron beam therapy. Results of this study show that the isodose distribution in a water phantom obtained from the simulation of irregular electron beams through the eMLC conforms to the pattern of the eMLC used in the delivery of the beam. It is therefore concluded that the dual eMLC could deliver isodose distributions reflecting the pattern of the eMLC field that was used in the delivery of the beam.