scholarly journals Durable Polymer Versus Biodegradable Polymer Drug-Eluting Stents After Percutaneous Coronary Intervention in Patients with Acute Coronary Syndrome: The HOST-REDUCE-POLYTECH-ACS Trial

Author(s):  
Hyo-Soo Kim ◽  
Jeehoon Kang ◽  
Doyeon Hwang ◽  
Jung-Kyu Han ◽  
Han-Mo Yang ◽  
...  

Background: Large scale randomized comparison of drug-eluting stents (DES) based on durable polymer (DP) versus biodegradable polymer (BP) technology is currently insufficient in patients with acute coronary syndrome (ACS). The present study sought to prove the non-inferiority of the DP-DES compared with the BP-DES in such patients. Methods: The HOST-REDUCE-POLYTECH-ACS trial is an investigator-initiated, randomized, open-label, adjudicator-blinded, multicenter, non-inferiority trial which compared the efficacy and safety of DP-DES and BP-DES in patients with ACS. The primary endpoint was patient oriented composite outcome (POCO, a composite of all-cause death, non-fatal myocardial infarction (MI), and any repeat revascularization) at 12 months. The key secondary endpoint was device oriented composite outcome (DOCO; a composite of cardiac death, target-vessel MI, or target lesion revascularization) at 12 months. Results: A total of 3413 patients were randomized to receive the DP-DES (1713 patients) and BP-DES (1700 patients). At 12 months, POCO occurred in 5.2% in the DP-DES group and 6.4% in the BP-DES group (Absolute risk difference: -1.2%, P non-inferiority <0.001). The key secondary endpoint, DOCO, occurred less frequently in the DP-DES group (DP-DES vs. BP-DES: 2.6% vs. 3.9%, HR 0.67, 95% CI 0.46-0.98, p=0.038), mostly due to a reduction in target lesion revascularization. The rate of spontaneous non-fatal MI and stent thrombosis were extremely low, with no significant difference between the 2 groups (0.6% vs. 0.8%; p=0.513 and 0.1% vs 0.4%; p=0.174, respectively). Conclusions: In ACS patients receiving percutaneous coronary intervention (PCI), DP-DES was non-inferior to BP-DES with regard to POCO at 12 months after index PCI. Clinical Trial Registration: URL: clinicaltrials.gov Unique identifier: NCT02193971

2021 ◽  
Vol 10 (6) ◽  
pp. 1278
Author(s):  
Endrin Koni ◽  
Wojciech Wanha ◽  
Jakub Ratajczak ◽  
Zhongheng Zhang ◽  
Przemysław Podhajski ◽  
...  

Among drug-eluting stents (DESs), the durable polymer everolimus-eluting stent (EES) and resolute zotarolimus-eluting stent (R-ZES) are widely used in clinical practice and have contributed to improve the outcomes of patients undergoing percutaneous coronary intervention (PCI). Few studies addressed their long-term comparative performance in patients with acute coronary syndrome (ACS). We aimed to investigate the 5 year comparative efficacy of EES and R-ZES in ACS. We queried ACTION-ACS, a large-scale database of ACS patients undergoing PCI. The treatment groups were analyzed using propensity score matching. The primary endpoint was a composite of mortality, myocardial infarction (MI), stroke, repeat PCI, and definite or probable stent thrombosis, which was addressed at the five-year follow-up. A total of 3497 matched patients were analyzed. Compared with R-ZES, a significant reduction in the primary endpoint at 5 years was observed in patients treated with EES (hazard ratio (HR) [95%CI] = 0.62 [0.54–0.71], p < 0.001). By landmark analysis, differences between the two devices emerged after the first year and were maintained thereafter. The individual endpoints of mortality (HR [95%CI] = 0.70 [0.58–0.84], p < 0.01), MI (HR [95%CI] = 0.55 [0.42–0.74], p < 0.001), and repeat PCI (HR [95%CI] = 0.65 [0.53–0.73], p < 0.001) were all significantly lower in the EES-treated patients. Stroke risk did not differ between EES and R-ZES. In ACS, a greater long-term clinical efficacy with EES vs. R-ZES was observed. This difference became significant after the first year of the ACS episode and persisted thereafter.


Sign in / Sign up

Export Citation Format

Share Document