scholarly journals Comparing imaging capabilities of spectral domain and swept source optical coherence tomography angiography in healthy subjects and central serous retinopathy

2018 ◽  
Vol 5 (1) ◽  
Author(s):  
Fupeng Wang ◽  
Qinqin Zhang ◽  
Anthony J. Deegan ◽  
Jun Chang ◽  
Ruikang K. Wang
Author(s):  
Anna Lentzsch ◽  
Laura Schöllhorn ◽  
Christel Schnorr ◽  
Robert Siggel ◽  
Sandra Liakopoulos

Abstract Purpose To compare swept-source (SS) versus spectral-domain (SD) optical coherence tomography angiography (OCTA) for the detection of macular neovascularization (MNV). Methods In this prospective cohort study, 72 eyes of 54 patients with subretinal hyperreflective material (SHRM) and/or pigment epithelial detachment (PED) on OCT possibly corresponding to MNV in at least one eye were included. OCTA scans were acquired using two devices, the PLEX Elite 9000 SS-OCTA and the Spectralis SD-OCTA. Fluorescein angiography (FA) was used as reference. Two graders independently evaluated en face OCTA images using a preset slab as well as a manually modified slab, followed by a combination of en face and cross-sectional OCTA. Results Sensitivity (specificity) for the automated slabs was 51.7% (93.0%) for SS-OCTA versus 58.6% (95.3%) for SD-OCTA. Manual modification of segmentation increased sensitivity to 79.3% for SS-OCTA but not for SD-OCTA (58.6%). The combination of en face OCTA with cross-sectional OCTA reached highest sensitivity values (SS-OCTA: 82.8%, SD-OCTA: 86.2%), and lowest number of cases with discrepancies between SS-OCTA and SD-OCTA (4.2%). Fleiss kappa as measure of concordance between FA, SS-OCTA, and SD-OCTA was 0.56 for the automated slabs, 0.60 for the manual slabs, and 0.73 (good agreement) for the combination of en face OCTA with cross-sectional OCTA. Concordance to FA was moderate for the automated slabs and good for manual slabs and combination with cross-sectional OCTA of both devices. Conclusion Both devices reached comparable results regarding the detection of MNV on OCTA. Sensitivity for MNV detection and agreement between devices was best when evaluating a combination of en face and cross-sectional OCTA.


PLoS ONE ◽  
2020 ◽  
Vol 15 (12) ◽  
pp. e0243830
Author(s):  
Yining Dai ◽  
Hao Zhou ◽  
Qinqin Zhang ◽  
Zhongdi Chu ◽  
Lisa C. Olmos de Koo ◽  
...  

Purpose To quantitatively assess choriocapillaris (CC) flow deficits in eyes with diabetic retinopathy (DR) using swept-source optical coherence tomography angiography (SS-OCTA). Methods Diabetic subjects with different stages of DR and age-matched healthy subjects were recruited and imaged with SS-OCTA. The en face CC blood flow images were generated using previously published and validated algorithms. The percentage of CC flow deficits (FD%) and the mean CC flow deficit size were calculated in a 5-mm-diameter circle centered on the fovea from the 6×6-mm scans. Results Forty-five diabetic subjects and 27 control subjects were included in the study. The CC FD% in diabetic eyes was on average 1.4-fold greater than in control eyes (12.34±4.14% vs 8.82±2.61%, P < 0.001). The mean CC FD size in diabetic eyes was on average 1.4-fold larger than in control eyes (2151.3± 650.8μm2 vs 1574.4±255.0 μm2, P < 0.001). No significant difference in CC FD% or mean CC FD size was observed between eyes with nonproliferative DR and eyes with proliferative DR (P = 1.000 and P = 1.000, respectively). Conclusions CC perfusion in DR can be objectively and quantitatively assessed with FD% and FD size. In the macular region, both CC FD% and CC FD size are increased in eyes with DR. SS-OCTA provides new insights for the investigations of CC perfusion status in diabetes in vivo.


Author(s):  
Pasha Anvari ◽  
Amin Najafi ◽  
Reza Mirshahi ◽  
Mahsa Sardarinia ◽  
Maryam Ashrafkhorasani ◽  
...  

Purpose: To compare the area of the foveal avascular zone (FAZ) in the superficial and deep retinal layers using two different spectral-domain optical coherence tomography angiography (OCTA) devices. Methods: A cross-sectional comparative study was conducted to obtain macular OCTA images from healthy subjects using Optovue RTVue XR Avanti (Optovue, Inc, Fremont, CA) and Spectralis HRA+OCTA (Heidelberg Engineering, Heidelberg, Germany). Two independent trained graders measured the FAZ area using automated slab segmentation. The FAZ area in the superficial and deep retinal layers were compared. Results: Twenty-three eyes of 23 subjects were included. The graders agreement was excellent (>0.86) for all measurements. The mean FAZ area was significantly larger at the superficial retinal layer as compared to the deep retinal layer on both devices (0.31 ± 0.08 mm2 vs 0.26 ± 0.08 mm2 in Optovue and 0.55 ± 0.16 mm2 vs 0.36 ± 0.13 mm2 in Spectralis, both P < 0.001). The mean FAZ area was significantly greater in the superficial and deep retinal layers using Spectralis as compared to Optovue measurements (P < 0.001 for both comparisons). Conclusion: In contrast to previous reports, the FAZ area was larger in the superficial retina as compared to deep retinal layers using updated software versions. Measurements from different devices cannot be used interchangeably.


2018 ◽  
Vol 103 (9) ◽  
pp. 1320-1326 ◽  
Author(s):  
Riccardo Sacconi ◽  
Enrico Borrelli ◽  
Eleonora Corbelli ◽  
Luigi Capone ◽  
Alessandro Rabiolo ◽  
...  

AimTo analyse quantitative optical coherence tomography angiography (OCTA) parameters of choriocapillaris (CC) at the foveal, perifoveal and parafoveal regions in healthy subjects of different age.MethodsIn this observational cross-sectional study, consecutive healthy subjects underwent swept source OCTA (PLEX Elite 9000, Carl Zeiss Meditec, USA). 3×3 mm and 6×6 mm scans centred on the fovea were acquired analysed and postprocessed with thresholding and binarisation processes. The main outcome measures included CC quantitative OCTA features (perfusion density (PD), vessel length density (VLD) and vessel diameter index (VDI)) at the foveal, perifoveal and parafoveal regions, and their relationship with age.ResultsSeventy-two eyes (72 patients, mean age 47.4±19.2 years, mean axial length 24.0±0.7 mm) were included. The mean PD was statistically lower in the foveal region compared with the parafoveal region in 3×3 mm scans (p=0.009). In subjects aged 70–80, PD of the foveal region was significantly reduced compared with the perifoveal region in 6×6 mm scans (p=0.008). A strong negative correlation was found between PD and ageing in the foveal, parafoveal and perifoveal regions (p<0.001 in all analyses). Furthermore, PD in the nasal-perifoveal region was significantly lower than the temporal-perifoveal region (p=0.005). No significant correlation was found between VLD and age in all regions, whereas VDI was strictly correlated with age (p<0.001).ConclusionsWe reported in vivo a strong negative correlation between PD and ageing in the CC of healthy subjects. This reduction seems related to a reduced diameter, and not to a reduced number, of capillaries. The age-related changes were higher in the central area compared with the perifoveal area.


2017 ◽  
Vol Volume 11 ◽  
pp. 889-895 ◽  
Author(s):  
Grace L Su ◽  
Douglas M Baughman ◽  
Qinqin Zhang ◽  
Kasra Rezaei ◽  
Aaron Y Lee ◽  
...  

Retina ◽  
2020 ◽  
Vol 40 (3) ◽  
pp. 499-506 ◽  
Author(s):  
Maria Vittoria Cicinelli ◽  
Michele Cavalleri ◽  
Andrea Celestino Consorte ◽  
Alessandro Rabiolo ◽  
Riccardo Sacconi ◽  
...  

2019 ◽  
Vol 236 (04) ◽  
pp. 551-554 ◽  
Author(s):  
Stefan Trachsler ◽  
Arthur Baston ◽  
Marcel Menke

Abstract Purpose To compare 4 optical coherence tomography-angiography (OCT-A) devices for foveal avascular zone (FAZ) measurements in healthy subjects. Methods The central retinas of 24 eyes of 12 healthy subjects were scanned with 4 different OCT-A devices (Optovue RTVue-XR, Zeiss Cirrus 5000-HD-OCT, a prototype Spectralis OCT2, Heidelberg Engineering, and Topcon DRI-OCT Triton Swept-source OCT). For the Topcon, Zeiss, and Optovue devices, 3-mm and 6-mm scans were performed. The Heidelberg device only provided 4-mm scans. En-face OCT-A images of the superficial and deep capillary plexus of the macular area were generated. The FAZ areas were measured and compared. Results Twenty-four healthy eyes were included. OCT-A devices showed significant differences in FAZ measurements. The Zeiss OCT-A device measured the smallest values for foveal avascular area (mean 218.7 mm2), followed by the Optovue device (229.6 mm2), the Topcon device (239.3 mm2), and the Heidelberg device (250.4 mm2). Differences were statistically significant for following devices: Heidelberg versus Optovue (p < 0.001), Heidelberg versus Zeiss (p < 0.001), Topcon versus Zeiss (p < 0.001), and Optovue versus Zeiss (p = 0.046). For the Optovue device, FAZ measurements were significantly different between 3 mm (mean 220 mm2) and 6 mm (mean 239.3 mm2, p = 0.007) scans. All other devices showed no significant difference within scan modes. Conclusion Current OCT-A devices provide images that allow such measurements, but values showed significant differences between devices and, for the Optovue instrument, even within scan modes. The data for OCTA measurements cannot be transferred interchangeably between the devices. Therefore, a patient should always be measured with the same device.