scholarly journals Study on the removal of interferences for the determination of ⁸⁷Sr/⁸⁶Sr isotopic ratio in petroleum drill-hole water samples using isotope dilution – inductively coupled plasma mass spectrometry (ID – ICP-MS)

2021 ◽  
Vol 8 (4) ◽  
pp. 34-43
Author(s):  
Dung Nguyen Thi Kim ◽  
Thuy Thai Thi Thu

87Sr/86Sr isotopic ratio is one of the useful tools that can authenticate the original source of the natural products from the earth-created and/or geological processes. However, the effect of interferences in petroleum drill-holes water sample such as thickness of sample matrix or isotopic signal of 87Rb might cause the low precision of 87Sr/86Sr determination using quadrupole inductively coupled plasma spectrometry (ICP-MS). The elimination of these mentioned effects was thus studied by using the ion - exchange chromatography. Calcium in sample matrix was separated on anionite column (Bio-Rad AG1-X8 resin) in methanol medium with the high efficiency while rubidium wasremoved from strontium on cation exchange resin (Bio-Rad AG50-X8) with strontium recovery over 99%. The isotope dilution technique with 86Sr - enriched isotopic standard solution was used for the control of separation process. The 87Sr/86Sr isotopic ratio was thus determined using ICP-MS with the signal correction by a strontium isotopic ratio standard reference material (NIST SRM 987).

2021 ◽  
Vol 7 (3) ◽  
pp. 8-16
Author(s):  
Kim Dung Nguyen Thi ◽  
Thi Lien Nguyen

The determination of 10B/11B isotope ratio and boron concentration in various watersamples using isotope dilution technique with inductively coupled plasma mass spectrometry (ICPMS) was studied. The interferences on precision and accuracy in isotopic ratio determination by ICPMS such as memory effects, dead time, spectral overlap of 12C were investigated for the selection of optimum conditions. By the addition of certain amounts of enriched 10B into samples, the 10B/11B ratio was determined through ICP-MS signal of 10B and 11B. The detection limit for 10B and 11B was experimentally obtained as 0.26 µg/L and 0.92 µg/L, respectively. The ratios of 10B/11B in measured water samples varied in the ranged between 0.1905 and 0.2484 for different matrices. This method has been then applied for the determination of boron isotopic ratio in VVER-1000 reactor-type simulated primary coolant water and in some environmental water samples.


2021 ◽  
Author(s):  
Andrew J. Mason ◽  
Anton Vaks ◽  
Sebastian F. M. Breitenbach ◽  
John N. Hooker ◽  
Gideon M. Henderson

Abstract. We describe a new method for the measurement of U/Pb ratios by isotope dilution multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) for the dating of geologically young clean carbonates, particularly speleothems. The method is intended for materials containing little or no initial 232Th. We illustrate and validate the method with four examples ranging from 0.57 Ma to 20 Ma old. The new method is capable of applying the 235U-207 and 238U-234U-206Pb chronometers, common Pb and quantifiable residual 234U/238U disequilibrium permitting. These provide an alternative to the more widely used 238U-206Pb chronometer, which can be highly inaccurate for samples a few million years old, owing to uncertainties in the excess initial 234U (hence, excess radiogenic 206Pb) commonly observed in speleothems.