scholarly journals Correction to: The evaluation of Rolimeter, KLT, KiRA and KT-1000 arthrometer in healthy individuals shows acceptable intra-rater but poor inter-rater reliability in the measurement of anterior tibial knee translation

Author(s):  
Armin Runer ◽  
Tommaso Roberti di Sarsina ◽  
Vasco Starke ◽  
Alessandra Iltchev ◽  
Gernot Felmet ◽  
...  
Author(s):  
Armin Runer ◽  
Tommaso Roberti di Sarsina ◽  
Vasco Starke ◽  
Alessandra Iltchev ◽  
Gernot Felmet ◽  
...  

Abstract Purpose To assess measurement equivalence, inter- and intra-rater reliability, standard error of measurements (SEM) and false positive measurements (FPM) of four different knee arthrometers (KLT,Karl Storz; KiRA, I + ; KT-1000 MEDmetric Corp; Rolimeter, Aircast) in healthy patients. Methods Four different investigators (two advanced (AR) and two beginners (BR)) examined 12 participants with healthy knees at two time points with regards to anterior tibial translation (ATT) and side-to-side difference (SSD). Test equivalence was assessed using the TOST (two-one-sided t test) procedure with ± 1 mm equivalence boundaries. Intraclass correlation coefficients (ICCs) were calculated using two-way mixed effects models. Furthermore, false positive-(SSD > 3 mm) and SEMs were assessed. Results A total of 2304 Lachman Tests were performed. Between-rater SSDs were equivalent between AR and BR raters for the Rolimeter only. Inter-rater ICC values (SSD, ATT) were graded as “poor” to “moderate” for all devices. Equivalent test–retest results were observed for all raters using the Rolimeter, KLT and KT-1000, whereas measurement consistency with KiRA was given in the advanced examiners group only. Intra-rater ICC values (Range: SSD, ATT) were graded as “poor” to “moderate” for SSD values and “moderate” to “good” for ATT. SEMs were lowest for the Rolimeter and highest for KiRA. FPM were never obtained with the Rolimeter (0%), twice (2.1%) with the KT-1000, three times (3.1%) with the KLT and 33 times (34.4%) using KiRA. Conclusion There is acceptable intra-rater but poor inter-rater reliability with all tested arthrometers. Measures of knee laxity are comparable between Rolimeter, KLT and KT-1000 but higher for KiRA. Clinically, the present study shows that repeated arthrometry measurements should always be performed by the same investigators.


2020 ◽  
Vol 48 (3) ◽  
pp. 573-580 ◽  
Author(s):  
Guan-yang Song ◽  
Hui Zhang ◽  
Jin Zhang ◽  
Zhi-jun Zhang ◽  
Tong Zheng ◽  
...  

Background: Anterior tibial subluxation (ATS) in extension after anterior cruciate ligament (ACL) injury highlights an increased anterior position of the tibia relative to the femur. Recent studies demonstrated that subluxation is sometimes irreducible and the normal tibiofemoral relationship is not restored by ACL reconstruction (ACLR), which raises concerns regarding clinical outcomes after ACLR. Hypothesis: Excessive preoperative ATS in extension is associated with inferior knee stability after anatomic ACLR. Study Design: Cohort study; Level of evidence, 3. Methods: From March 2016 to January 2017, a total of 487 consecutive patients with clinically diagnosed noncontact ACL injuries who underwent primary anatomic ACLR were retrospectively analyzed. Of these patients, 430 met the criteria for inclusion in this study. Anterior subluxation of the lateral and medial compartments (ASLC and ASMC) in extension relative to the femoral condyles was measured on preoperative magnetic resonance imaging. Twenty patients (study group) who demonstrated excessive (>10 mm) ASLC and ASMC in extension were matched 1:2 to 40 participants (control group) who showed minimal or no (<3 mm) ASLC and ASMC in extension. The amount of ASLC and ASMC in extension relative to the femoral condyles at 2 years postoperatively was the primary outcome. Moreover, the Lysholm score, IKDC grade (International Knee Documentation Committee), and stability assessments (pivot-shift test and KT-1000 arthrometer side-to-side difference) were evaluated preoperatively and at the last follow-up visit. Results: The preoperative mean ASLC and ASMC in extension of the study group were both significantly larger than those of the control group (study group vs control group: ASLC, 13.5 mm vs 1.2 mm; ASMC, 12.4 mm vs 1.0 mm; P < .05). Moreover, patients in the study group showed significantly larger posterior tibial slope than the patients in the control group (17.8°± 2.5° vs 9.5°± 1.5°; P < .05). At the final follow-up visit, the mean ASLC and ASMC of the study group were 8.1 mm and 7.3 mm, which were significantly larger than those of the control group (ASLC, 0.9 mm; ASMC, 0.7 mm; P < .05). In addition, the study group showed inferior knee stability when compared with the control group in terms of both the pivot-shift test (study group vs control group: 2 grade 2, 10 grade 1, and 8 grade 0 vs 1 grade 1 and 39 grade 0; P < .05) and the KT-1000 arthrometer side-to-side difference (study group vs control group: 4.4 ± 1.2 mm vs 1.5 ± 0.6 mm; P < .05). Furthermore, the study group showed significantly lower mean Lysholm score (study group vs control group: 80.3 ± 6.3 vs 93.3 ± 4.3, P < .05) and IKDC grading results (study group vs control group: 3 grade C, 16 grade B, and 1 grade A vs 3 grade B and 37 grade A; P < .05) as compared with the control group. Conclusion: In this short-term study, the excessive (>10 mm) preoperative ATS in extension after ACL injury was associated with inferior knee stability after anatomic ACLR.


Author(s):  
Georg Mattiassich ◽  
Reinhold Ortmaier ◽  
Harald Kindermann ◽  
Jürgen Barthofer ◽  
Imre Vasvary ◽  
...  

Abstract Background Anterior cruciate ligament (ACL) injury can lead to reduced function, meniscal lesions, and early joint degeneration. Preservation of a torn ACL using the Internal Brace technique might re-establish normal knee kinematics, avoid donor-site morbidity due to tendon harvesting, and potentially maintain proprioception of the knee. Methods Fifty subjects were recruited for this study between December 2015 and October 2016. Two groups of individuals who sustained a unilateral ACL rupture were included: those who underwent surgery with preservation of the injured ACL (Internal Brace technique; IB) and those who underwent ACL reconstruction using a hamstring tendon graft (all-inside technique; AI). Subjective self-administered scores were used: the German version of the IKDC Subjective Knee Form (International Knee Documentation Committee), the German version of the WOMAC (Western Ontario and McMaster Universities Arthritis Index), SF-36 (short form), the German version of the KOOS (Knee Osteoarthritis Outcome Score), and the German version of themodified Lysholm Score by Lysholm and Gillquist. Anterior tibial translation was assessed using the KT-1000 Arthrometer (KT-1000 Knee Ligament Arthrometer, MEDmetric Corp., San Diego, CA, USA). Magnetic resonance evaluation was performed in all cases. Results Twenty-three subjects (46 %) were men, and the mean age was 34.7 years. The objective IKDC scores were “normal” in 15 and 14 patients, “nearly normal” in 11 and 7 patients, and “abnormal” in 1 and 2 patients, in the IB and AI groups, respectively. KT-1000 assessment showed a sideto-side difference of more than 3 mm on maximum manual testing in 11 (44 %) and 6 subjects (28.6 %) in the IB and AI groups, respectively. In the postoperative MRI, 20 (74 %) and 22 subjects (96 %) in the IB and AI groups had an intact ACL. Anterior tibial translation was significantly higher in the IB group compared with the AI group in the manual maximum test. Conclusions Preservation of the native ACL with the Internal Brace primary repair technique can achieve comparable results to ACL reconstruction using Hamstring autografts over a short term. Clinically relevant limitations such as a higher incidence of pathologic laxity, with patients more prone to pivot-shift phenomenon were observed during the study period.


2021 ◽  
pp. 036354652110218
Author(s):  
Anne Gro Heyn Faleide ◽  
Liv Heide Magnussen ◽  
Bård Erik Bogen ◽  
Torbjørn Strand ◽  
Ingunn Fleten Mo ◽  
...  

Background: Deciding when patients are ready to return to sport (RTS) after an anterior cruciate ligament (ACL) reconstruction (ACLR) is challenging. The understanding of which factors affect readiness and how they may be related is limited. Therefore, despite widespread use of RTS testing, there is a lack of knowledge about which tests are informative on the ability to resume sports. Purpose: To examine whether there is an association between knee laxity and psychological readiness to RTS after ACLR and to evaluate the predictive value of these measures on sports resumption. Study Design: Cohort study; Level of evidence, 2. Methods: Patients aged ≥16 years engaged in physical activity/sports before injury were recruited at routine clinical assessment 9-12 months after ACLR. Exclusion criteria were concomitant ligament surgery at ACLR and/or previous ACL injury in the contralateral knee. At baseline, a project-specific activity questionnaire and the ACL–Return to Sport After Injury (ACL-RSI) scale were completed. Knee laxity was assessed by use of the Lachman test, KT-1000 arthrometer, and pivot-shift test. Two years after surgery, knee reinjuries and RTS status (the project-specific questionnaire) were registered. Associations between psychological readiness and knee laxity were evaluated with the Spearman rho test, and predictive ability of the ACL-RSI and knee laxity tests were examined using regression analyses. Results: Of 171 patients screened for eligibility, 132 were included in the study. There were small but significant associations between the ACL-RSI score and the Lachman test (rho = −0.18; P = .046) and KT-1000 arthrometer measurement (rho = −0.18; P = .040) but no association between the ACL-RSI and the pivot-shift test at the time of recruitment. Of the total patients, 36% returned to preinjury sport level by 2 years after surgery. Higher age, better psychological readiness, and less anterior tibial displacement (KT-1000 arthrometer measurement) were significant predictors of 2-year RTS (explained variance, 33%). Conclusion: Small but significant associations were found between measurements of psychological readiness and anterior tibial displacement, indicating that patients with less knee laxity after ACLR feel more ready to RTS. ACL-RSI and KT-1000 arthrometer measurements were independent predictors of 2-year RTS and should be considered in RTS assessments after ACLR.


2020 ◽  
Author(s):  
Georg Mattiassich ◽  
Reinhold Ortmaier ◽  
Harald Kindermann ◽  
Jürgen Barthofer ◽  
Imre Vasvary ◽  
...  

Abstract Background Anterior cruciate ligament (ACL) injury can lead to reduced function, meniscal lesions, and early joint degeneration. Preservation of a torn ACL using the Internal Brace® technique might re-establish normal knee kinematics, avoid donor-site morbidity due to tendon harvesting, and potentially maintain proprioception of the knee. Methods Fifty subjects were recruited for this study between December 2015 and October 2016. Two groups of individuals who sustained unilateral ACL rupture were included: those who underwent surgery with preservation of the injured ACL (Internal Brace® technique; IB) and those who underwent ACL reconstruction using a hamstring tendon graft (all-inside technique; AI). Subjective self-administered scores were used: the German Version of the IKDC Subjective Knee Form (International Knee Documentation Committee), the German Version of the WOMAC (Western Ontario and McMaster Universities Arthritis Index), SF-36 (short form), the German Version of the KOOS (Knee Osteoarthritis Outcome Score), and the German Version of the modified Lysholm-score by Lysholm and Gillquist. Anterior tibial translation was assessed using the KT-1000 arthrometer® (KT-1000 Knee Ligament Arthrometer, MEDmetric Corp., San Diego, CA, USA). Magnetic resonance evaluation was performed in all cases. Results Twenty-three subjects (46%) were men, and the mean age was 34.7 years. The objective IKDC scores were “normal” in 15 and 14 patients, “nearly normal” in 11 and 7 patients, and “abnormal” in 1 and 2 patients, in the IB and AI groups, respectively. KT-1000 assessment showed a side-to-side difference of more than 3 mm on maximum manual testing in 11 (44%) and 6 subjects (28.6%) in the IB and AI groups, respectively. In the postoperative MRI, 20 (74%) and 22 subjects (96%) in the IB and AI groups showed an intact ACL. Anterior tibial translation was significantly higher in the IB group compared to the AI group in the manual maximum test. Conclusions Preservation of the native ACL with the Internal Brace ® primary repair technique can achieve comparable results to ACL reconstruction using Hamstring autografts over a short term. Clinically relevant limitations such as a higher incidence of pathologic laxity, with patients more prone to pivot shift phenomenon were observed during the study period.


2020 ◽  
Vol 48 (14) ◽  
pp. 3486-3494
Author(s):  
Guan-yang Song ◽  
Qian-kun Ni ◽  
Tong Zheng ◽  
Zhi-jun Zhang ◽  
Hua Feng ◽  
...  

Background: Steep posterior tibial slope (PTS; >13°), excessive anterior tibial subluxation (ATS) in extension (>10 mm), and meniscus posterior horn tears (MPHTs) have been identified to be associated with primary anterior cruciate ligament (ACL) reconstruction (ACLR) failure. Recent studies have reported that steep PTS is directly correlated with excessive ATS in extension and concomitant MPHTs, especially for those patients with chronic (>6 months) ACL deficiency. There is increasing biomechanical evidence that slope-reducing tibial osteotomy decreases ATS in extension and protects the ACL graft. Hypothesis: Slope-reducing tibial osteotomy combined with primary ACLR is effective for producing improved knee stability in patients with steep PTS (>13°), excessive ATS in extension (>10 mm), and concomitant chronic MPHTs (>6 months). Study Design: Case series; Level of evidence, 4. Methods: Between June 2016 and January 2018, 18 patients with ACL injuries who had steep PTS (>13°), excessive ATS in extension (>10 mm), and concomitant chronic MPHTs (>6 months) underwent slope-reducing tibial osteotomy combined with primary ACLR. The PTS and anterior subluxation of the lateral and medial compartment (ASLC and ASMC) in extension before and after the index procedures were regarded as primary clinical outcomes. Moreover, Lysholm score, Tegner activity score, International Knee Documentation Committee (IKDC) objective grade, pivot-shift test, and KT-1000 side-to-side difference were evaluated preoperatively and at the minimum 2-year follow-up visit. Results: The mean PTS was 18.5° (range, 17°-20°) preoperatively and 8.1° (range, 7°-9°) postoperatively ( P < .01). The mean ASLC and ASMC in extension were 12.1 mm and 11.9 mm preoperatively, which reduced to 1.0 mm and 1.5 mm at the last follow-up visit ( P < .05). In addition, all of the following showed significant improvements (pre- vs postoperatively): mean Lysholm score (46.5 vs 89.5; P < .05), mean Tegner activity score (5.7 vs 7.3; P < .05), IKDC objective grading results (18 grade D vs 14 grade A and 4 grade B; P < .05), pivot-shift tests (15 grade 2+ and 3 grade 3+ vs 18 grade 0; P < .01), and KT-1000 side-to-side difference (13.0 mm vs 1.6 mm; P < .01). Moreover, no graft reruptures were found at the final follow-up visit. Conclusion: In this study, slope-reducing tibial osteotomy combined with primary ACLR effectively improved knee stability in patients with steep PTS (>13°), excessive ATS in extension (>10 mm), and concomitant chronic MPHTs (>6 months).


2021 ◽  
Vol 13 (1) ◽  
Author(s):  
Yosuke Takahashi ◽  
Yuji Fujino ◽  
Kohei Miura ◽  
Ayumi Toida ◽  
Tadamitsu Matsuda ◽  
...  

Abstract Background Ultrasonography (US) is a feasible and accessible method for the measurement of skeletal muscle mass. This technique presents acceptable intra-rater reliability; however, there are a few reports on its inter-rater reliability. Additionally, relative reliability should equally be inspected to determine the presence of systematic errors. Therefore, this study aimed to investigate the intra- and inter-rater reliabilities and absolute reliability of rectus femoris muscle thickness as measured using US. Methods The participants included in our study comprised 12 healthy young men (26.5 ± 3.9 years. Rectus femoris muscle thickness was measured from the right side of the thigh using US by two trained physical therapists. Inter- and intra-rater reliabilities were determined using the intraclass correlation coefficient (ICC) (1, 1) and ICC (2, 1) methods, respectively. Absolute reliability was evaluated using Bland − Altman analysis. Additionally, we calculated the minimal detectable change at the 95% level of confidence (MDC95). Result According to the results of the Bland − Altman analysis, no fixed or proportional errors were present. The ICC (1, 1) was 0.95, and the ICC (2, 1) was 0.70. The MDC95 values of rectus femoris thickness for the intra- and inter-rater reliabilities were 2.0 mm and 4.3 mm, respectively. Conclusions In our study, intra- and inter-rater reliabilities were measured at “excellent” and “moderate” levels in the healthy individuals based on a previously defined scale. Moreover, we determined the measurement error for quantifying rectus femoris thickness. Therefore, the measurement of rectus femoris thickness using US could be considered applicable in clinical research.


2009 ◽  
Vol 17 (1) ◽  
pp. 77-79 ◽  
Author(s):  
S Arneja ◽  
J Leith

The KT-1000 knee arthrometer (KT-1000) is an objective instrument to measure anterior tibial motion relative to the femur for anterior cruciate ligament (ACL) reconstruction. Four studies between 1950 and 2007 regarding validity of the KT-1000 were identified using a Medline search. One had interpretable information on sensitivities, specificities, and predictive values to validate the instrument as a diagnostic tool in patients with acute or chronic ACL injuries. Three had limitations in methodology. We suggest that the KT-1000 should be used with caution as an objective instrument. Rather, using a KT-1000 score derived by subtracting the anterior tibial motion relative to the femur of the injured knee to that of the uninjured knee may be more appropriate as a dichotomous diagnostic test with a threshold of 2 or 3 mm.


Sign in / Sign up

Export Citation Format

Share Document