Kinetic Spectrophotometric Determination of Trace Aluminum by Catalytic Activity of Al 3+ on the Oxidation of Methylene Blue by Hydrogen Peroxide

2003 ◽  
Vol 143 (1) ◽  
pp. 53-57 ◽  
Author(s):  
Ren-Min Gong ◽  
Qiu-Yi Chen ◽  
Zhi-Li Liu
2009 ◽  
Vol 74 (8-9) ◽  
pp. 977-984
Author(s):  
Sofija Rancic ◽  
Snezana Nikolic-Mandic

A new reaction was suggested and a new kinetic method was elaborated for determination of Bi(III) in solution, based on its catalytic effect on the oxidation of phenyl-fluorone (PF) by hydrogen peroxide in ammonia buffer. By application of spectrophotometric technique, a limit of quantification (LQ) of 128 ng cm-3 was reached, and the limit of detection (LD) of 37 ng cm-3 was obtained, where LQ was defined as the ratio signal: noise = 10:1 and LD was defined as signal 3:1 against the blank. The RSD value was found to be in the range 2.8-4.8 % for the investigated concentration range of Bi(III). The influence of some ions upon the reaction rate was tested. The method was confirmed by determining Bi(III) in a stomach ulcer drug ('Bicit HP', Hemofarm A.D.). The obtained results were compared to those obtained by AAS and good agreement of results was obtained.


2008 ◽  
Vol 55 (1) ◽  
pp. 137-142 ◽  
Author(s):  
Zora M. Grahovac ◽  
Snežana S. Mitić ◽  
Todor G. Pecev ◽  
Emilija T. Pecev ◽  
Aleksandra N. Pavlović

Sign in / Sign up

Export Citation Format

Share Document