Self-poisoning as a cause of admission in a medical intensive care unit and a question of misuse of prescription medications

Heart & Lung ◽  
2022 ◽  
Vol 51 ◽  
pp. 17-21
Author(s):  
Ana Vujaklija Brajković ◽  
Marta Grgat ◽  
Luka Bielen ◽  
Jurica Brajković ◽  
Ozrenka Zlopaša ◽  
...  
2021 ◽  
pp. 193229682110275
Author(s):  
Wannita Tingsarat ◽  
Patinut Buranasupkajorn ◽  
Weerapan Khovidhunkit ◽  
Patchaya Boonchaya-anant ◽  
Nitchakarn Laichuthai

Objective: To assess the accuracy of continuous glucose monitoring (CGM) in medical intensive care unit (MICU) patients. Methods: A Medtronic Enlite® sensor accuracy was assessed versus capillary blood glucose (CBG) and plasma glucose (PG) using the mean absolute relative difference (MARD), surveillance error grid (SEG) analysis and modified Bland-Altman plots. Results: Using CBG as a reference, MARD was 6.6%. Overall, 99.7% of the CGM readings were within the “no risk” zone. No significant differences in accuracy were seen within vasopressor subgroups. Using PG as the reference, MARD was 8.8%. The surveillance error grid analysis showed 95.2% of glucose readings were within the “no risk” zone. There were no device-related adverse events. Conclusion: The CGM sensor showed acceptable accuracy in MICU patients, regardless of vasopressor use.


Sign in / Sign up

Export Citation Format

Share Document