Internal standardization and least-squares background correction in high-resolution continuum source flame atomic absorption spectrometry to eliminate interferences on determination of Pb in phosphoric acid

2008 ◽  
Vol 63 (9) ◽  
pp. 992-995 ◽  
Author(s):  
Jorge L. Raposo ◽  
Silvana R. Oliveira ◽  
Joaquim A. Nóbrega ◽  
José A. Gomes Neto
2014 ◽  
Vol 1010-1012 ◽  
pp. 417-421
Author(s):  
Chao Li ◽  
Shang Long Chen

The aim of this work was to evaluate the microemulsification after incomplete digestion as sample preparation procedure for determination of Ca in leisure foods by high resolution continuum source flame atomic absorption spectrometry (HR-CS FAAS). Microemulsions were prepared with digestive liquor and n-butyl alcohol as an auxiliary emulsifier after incomplete digestion treatment. The appropriate fuel flow, 70 L/h, and the optimum burner height, 5 mm, were obtained by single factor experimental design. The results showed that the correlation coefficient was better than 0.999, the characteristic concentration was 0.086 mg/L, the precision (RSD) for 6 replicate measurements of Ca was 1.7% and the average of recoveries was 102.1%. F-test and t-test (95% confidence level) in between the proposed method and the comparative method, using microwave digestion-HR-CS FAAS, had no significant difference. Therefore, the proposed method was accurate and stable with a high practical value. It provided scientific basis for determination of metal elements in food.