Determination of cadmium in aqueous media by flow injection cold vapour atomic absorption spectrometry. Application to natural water samples

1998 ◽  
Vol 13 (4) ◽  
pp. 295-300 ◽  
Author(s):  
María Liva Garrido ◽  
Riansares Muñoz-olivas ◽  
Carmen Cámara
1995 ◽  
Vol 17 (1) ◽  
pp. 25-29 ◽  
Author(s):  
J. L. Burguera ◽  
M. Burguera ◽  
P. Carrero ◽  
J. Marcano ◽  
C. Rivas ◽  
...  

The paper describes the development of an automatic on-line column pre-concentration technique using a time based-flow injection atomic absorption spectrometry system. A manifold incorporating a micro-column containing 25 mg of Dowex 50W-X8 was used with a time-based injector for the pre-concentration and determination of copper in natural and drinking waters. The system features depend on the alternate positions of a solenoid valve. The 3σ detection limits, enrichment factors, sampling frequency, relative standard deviations and linear calibration graphs were, respectively, in the range 0.6-1.5 μg/l, 25-60, 15-30 measurements/h, 1.0-3.1% and 1-65 μg/ml for pre-concentration times of 1 min. The procedure was successfully applied to a range of water samples and the accuracy was assessed through recovery experiments, the analysis of certified reference water samples and by independent analysis by atomic absorption spectrometry with electrothermal atomization.


Sign in / Sign up

Export Citation Format

Share Document