Planar processed polysilicon sealed cavities for pressure transducer arrays

Author(s):  
H. Guckel ◽  
D.W. Burns
1990 ◽  
Vol 23 (1-3) ◽  
pp. 1070-1079 ◽  
Author(s):  
S. Terry ◽  
J.S. Eckerle ◽  
R.D. Kornbluh ◽  
T. Low ◽  
C.M. Ablow

2021 ◽  
Vol 11 (9) ◽  
pp. 4025
Author(s):  
Dario Messenio ◽  
Marco Ferroni ◽  
Federica Boschetti

Glaucoma is the second cause of irreversible blindness in the world. Intraocular pressure (IOP) is a recognized major risk factor for the development and progression of glaucomatous damage. Goldmann applanation tonometry (GAT) is internationally accepted as the gold standard for the measurement of IOP. The purpose of this study was to search for correlations between Goldmann tonometry and corneal mechanical properties and thickness by means of in vitro tests. IOP was measured by the Goldmann applanation tonometer (GIOP), and by a pressure transducer inserted in the anterior chamber of the eye (TIOP), at increasing pressure levels by addition of saline solution in the anterior chamber of enucleated pig eyes (n = 49). Mechanical properties were also determined by inflation tests. The GAT underestimated the real measurements made by the pressure transducer, with most common differences in the range 15–28 mmHg. The difference between the two instruments, highlighted by the Bland–Altman test, was confirmed by ANOVA, normality tests, and Mann–Whitney’s tests, both on the data arranged for infusions and for the data organized by pressure ranges. Pearson correlation tests revealed a negative correlation between (TIOP-GIOP) and both corneal stiffness and corneal thickness. In conclusion, data obtained showed a discrepancy between GIOP and TIOP more evident for softer and thinner corneas, that is very important for glaucoma detection.


Author(s):  
Martin Angerer ◽  
Michael Zapf ◽  
Benjamin Leyrer ◽  
Nicole V. Ruiter

1967 ◽  
Vol 38 (8) ◽  
pp. 1145-1147 ◽  
Author(s):  
H. L. Dunegan ◽  
C. A. Tatro

Sign in / Sign up

Export Citation Format

Share Document