scholarly journals REVERSE-PHASE HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION AND FORCED DEGRADATION STUDIES OF EMTRICITABINE, RILPIVIRINE, AND TENOFOVIR ALAFENAMIDE IN SOLID DOSAGE FORM

Author(s):  
Juluri Krishna Dutta Tejaswi ◽  
Govinda Rajan R

Objective: A stability indicating reverse-phase high-performance liquid chromatography (RP-HPLC) method was developed and validated for the estimation of emtricitabine (EMT), rilpivirine (RIL), and tenofovir alafenamide (TAF) in combined dosage forms and its API.Methods: Chromatographic separation was achieved on Waters ACQUITY RP-HPLC with PDA detector having Zodiac C18 Column (250×4.6×5μ) using mobile phase mixture of phosphate buffer: acetonitrile in the ratio of 40:60 v/v at 262 nm.Results: The assay was performed with tablets, and the % assay was found to be 100.104 for EMT, 99.74 for RIL, and 102.41 for TAF which shows that the method is useful for routine analysis. The linearity was found to be linear with a correlation coefficient of 0.999, which shows that the method is capable of producing good sensitivity. The retention time (RT) of EMT, RIL, and TAF using optimum conditions was found to be 2.517, 3.273, and 6.697 min. Forced degradation studies (FDS) were performed on sample using acid, base, thermal, photolytic, and peroxide degradation.Conclusion: Due to its simplicity, rapidness, high precision, and low RT value, this method was successfully applied to the estimation of EMT, RIL, and TAF combined dosage form. The drugs were found to be stable at FDS, and the net degradation was found to be within the limits.

Author(s):  
Juluri Krishna Dutta Tejaswi ◽  
Govinda Rajan R

Objective: A stability indicating reverse-phase high-performance liquid chromatography (RP-HPLC) method was developed and validated for the estimation of emtricitabine (EMT), rilpivirine (RIL), and tenofovir alafenamide (TAF) in combined dosage forms and its API.Methods: Chromatographic separation was achieved on Waters ACQUITY RP-HPLC with PDA detector having Zodiac C18 Column (250×4.6×5μ) using mobile phase mixture of phosphate buffer: acetonitrile in the ratio of 40:60 v/v at 262 nm.Results: The assay was performed with tablets, and the % assay was found to be 100.104 for EMT, 99.74 for RIL, and 102.41 for TAF which shows that the method is useful for routine analysis. The linearity was found to be linear with a correlation coefficient of 0.999, which shows that the method is capable of producing good sensitivity. The retention time (RT) of EMT, RIL, and TAF using optimum conditions was found to be 2.517, 3.273, and 6.697 min. Forced degradation studies (FDS) were performed on sample using acid, base, thermal, photolytic, and peroxide degradation.Conclusion: Due to its simplicity, rapidness, high precision, and low RT value, this method was successfully applied to the estimation of EMT, RIL, and TAF combined dosage form. The drugs were found to be stable at FDS, and the net degradation was found to be within the limits.


Author(s):  
SRUTHI A ◽  
UTTAM PRASAD PANIGRAHY

Objective: A rapid, sensitive and specific reverse phase High performance liquid Chromatography (RP-HPLC) method was developed for the estimation of Fimasartan in bulk and pharmaceutical dosage form. Method: The RP-HPLC analysis was performed isocratically on a Primacel C18 column (150 mm × 4.6 mm internal diameter, 5 μm particle size) using mobile phase of composition Acetonitrile and 0.1% orthophosphoric Acid in 80:20, v/v proportions with a flow rate of 0.8 ml/min. Results: The analyte was monitered with UV-detector at 265 nm. In the developed method Fimasartan elutes at a typical retention time of 2.4 min. The proposed method is having linearity in the concentration ranging from 5-30 μg/ml of Fimasartan. Conclusion: : The method was statistically validated and had been applied to analysis of the drug in bulk and pharmaceutical dosage form.


Author(s):  
Geetha Susmita A ◽  
Rajitha G ◽  
Ramya Yadav Y ◽  
Uma P

Objective: The objective of this study was to develop and validate a stability-indicating reverse-phase high-performance liquid chromatography (RP-HPLC) method for the simultaneous estimation of the metformin and empagliflozin in tablet dosage forms.Methods: The chromatographic conditions were optimized and it was run through Std. BDS (250 mm × 4.6 mm, 5 m) column with mobile phase consisting of 0.1% orthophosphoric acid buffer: acetonitrile in the ratio of 50:50. The flow rate was 1 ml/min and optimized wavelength was 210 nm. Temperature was maintained at 30°C.Results: The retention times of metformin and empagliflozin were found to be 2.588 min and 3.679 min and percentage relative standard deviation (RSD) of the metformin and empagliflozin was found to be 0.59 and 1.2, respectively. Percentage recovery was in the range of 100.01–100.65% for metformin and empagliflozin, respectively.Conclusion: A sensitive, rapid, and specific method has been developed for the simultaneous estimation of metformin and empagliflozin using RP-HPLC in tablet dosage form.


Author(s):  
Geetha Susmita A ◽  
Rajitha G ◽  
Ramya Yadav Y ◽  
Uma P

Objective: The objective of this study was to develop and validate a stability-indicating reverse-phase high-performance liquid chromatography (RP-HPLC) method for the simultaneous estimation of the metformin and empagliflozin in tablet dosage forms.Methods: The chromatographic conditions were optimized and it was run through Std. BDS (250 mm × 4.6 mm, 5 m) column with mobile phase consisting of 0.1% orthophosphoric acid buffer: acetonitrile in the ratio of 50:50. The flow rate was 1 ml/min and optimized wavelength was 210 nm. Temperature was maintained at 30°C.Results: The retention times of metformin and empagliflozin were found to be 2.588 min and 3.679 min and percentage relative standard deviation (RSD) of the metformin and empagliflozin was found to be 0.59 and 1.2, respectively. Percentage recovery was in the range of 100.01–100.65% for metformin and empagliflozin, respectively.Conclusion: A sensitive, rapid, and specific method has been developed for the simultaneous estimation of metformin and empagliflozin using RP-HPLC in tablet dosage form.


Author(s):  
Swapnali Suresh Mankar ◽  
Muhammad Younas ◽  
Vaishavi Kalode ◽  
Pranita Jirvankar ◽  
Prasanna Phutane ◽  
...  

Aim: The recent analysis is required to do novel and simple, sensitive, precise, efficient, instant and reproducible reverse phase high performance liquid chromatography (RP-HPLC) method for estimation of antidiabetic drug in the unit dosage form. Validation of this method is also planned to make it suitable for the actual use. Study Design: Department of Pharmaceutical Chemistry, Datta Meghe Institute of Medical Sciences deemed to be university, Wardha in collaboration with Balkh university, Mazar-e-sharif, Afghanistan between August 2021 and December 2021. Methodology: In that article develop the method and validate it by estimation of antidiabetic drugs in solid dosage form by RP-HPLC, by using System suitability test, Repeatability, Precision studies (Intra-day and Interday/Intermediate), Linearity/Calibration studies, Robustness, Force degradation, Specificity, Drug recovery/accuracy studies. Results: as per ICH guidelines, the performance if system suitability in remogliflozin were achieved all guidelines; in that, tailing factor (T),separation factors(α),theoretical plates(N),capacity  factor (k’), resolution (R) and RSD (%). The validated stress degradation studies under thermal, oxidative, alkali and acid, in few degradation products for remogliflozin (REM). Conclusion: From the results we conclude that, this novel technique which validated for exploration by reverse phase high performance liquid chromatography (RP-HPLC) should be used for routine quality control of remogliflozin (REM) prediction from developed formulation.


Author(s):  
BHOOMI D PATEL ◽  
MEHTA BHAVYA ◽  
ANKIT B CHAUDHARY

Objective: The objective of the study was to develop and validate reverse-phase high-performance liquid chromatography (RP-HPLC) method and apply method to tablet dosage form. Methods: A simple, rapid, economical, precise, and accurate RP-HPLC method for simultaneous estimation of lamivudine and zidovudine in their combined dosage form has been developed. Results: A RP-HPLC method was developed for the simultaneous estimation of lamivudine and zidovudine. In their combined dosage form has been developed. The separation was achieved by LC-C18 column (150 mm ×4.6 mm, 5 μm) and water: methanol (65:35v/v) as mobile phase, at a flow rate of 0.8 ml/min. Detection was carried out at 272 nm. Retention time of lamivudine and zidovudine was found to be 3.007 min and 4.647, respectively. The method has been validated for linearity, accuracy, and precision. The assay method was found to be linear from 50% to 150% for lamivudine and zidovudine. Conclusion: Developed method was found to be accurate, precise, and rapid for simultaneous estimation of lamivudine and zidovudine in their combined dosage form.


2018 ◽  
Vol 10 (6) ◽  
pp. 242
Author(s):  
Juluri Krishna Dutta Tejaswi ◽  
R. Govinda Rajan

Objective: A stability indicating reverse phase high-performance liquid chromatography (RP-HPLC) method was developed and validated for the estimation of the combined tablet formulation of lamivudine (LAM) and raltegravir (RAL) in dosage forms and its API.Methods: Chromatographic separation was achieved on inertsil ODS C18 5 µm (4.6 X 150 mm) using a mobile phase (MP) consisting of a mixture of mixed orthophosphoric acid (OPA): acetonitrile (ACN) in the ratio 50:50 v/v which was determined at 242 nm respectively. Results: The assay of LAM and RAL was performed with tablets, and the % assay was found to be 100.12 and 99.89 which shows that the method is useful for routine analysis. The linearity of LAM and RAL was found to be linear with a correlation coefficient of 0.998 and 0.999, which shows that the method is capable of producing good sensitivity. The retention time of LAM and RAL was 1.99 min and 4.34 min respectively; linearity range was found to lie from 15 µg/ml to 75 µg/ml for LAM, 30 µg/ml to 150 µg/ml for RAL with a correlation coefficient of 0.999 respectively. Forced degradation studies were conducted in acidic, basic, thermal, photolytic and peroxide where all the degradation peaks were monitored.Conclusion: The proposed HPLC method was found to be simple, specific, precise, accurate, rapid and economical for simultaneous estimation of LAM and RAL in bulk and tablet dosage form. Thus the validated economical method was applied for forced degradation study of LAM and RAL tablet.