scholarly journals METHOD DEVELOPMENT AND VALIDATION OF SELECTIVE AND HIGHLY SENSITIVE METHOD FOR DETERMINATION OF APIXABAN IN HUMAN PLASMA USING LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY

Author(s):  
CHAITANYA KRISHNA ATMAKURI ◽  
CHEEPURUPALLI PRASAD ◽  
RAJARAM S. PATIL ◽  
KARUNAKRANTH DHARANI

Objective: The present research work aims to develop and validate a selective and highly sensitive method for the determination of apixaban in human plasma using liquid chromatography-tandem mass spectrometry (LC-MS/MS). Methods: 200 µl of sodium heparin plasma samples were acidified and clean-up was performed by using solid-phase extraction (SPE). Apixaban 13C D3 was used as an internal standard (deuterated) to lower the relative matrix effects and a single step SPE was employed for sample clean up. 10 µl of SPE eluent was loaded onto Hypersil Beta Basic C18, 100×4.6 mm, 5 µ column for highly selective chromatographic separation using an isocratic mobile phase. 2 mmol ammonium acetate in water and acetonitrile were delivered by using a quaternary low-pressure gradient pump without premixing at a minimum flow rate of 0.50 ml/min. Results: LC-MS/MS method was successfully developed and validated to demonstrate the lowest detection limit of 0.05 ng/ml and a linear dynamic range from 1-250 ng/ml with r2>0.99. Method development and validation results proved that the method is selective and highly sensitive for the determination of apixaban in human plasma using LC-MS/MS. Conclusion: Current method can be applied for both therapeutic drug monitoring (TDM) and pharmacokinetic (PK) study analysis.

Author(s):  
L. Narasimha Rao K ◽  
N. Devanna ◽  
K.v.n. Suresh Reddy

<p><strong>Objective: </strong>The objective of this work is method development and validation study for quantitative determination of 1-[2-(2,4-difluorophenyl)-2,3-epoxypropyl]-1H-1,2,4-triazole, a genotoxic impurity and its precursor in a fluconazole drug sample by liquid chromatography–tandem mass spectrometry.</p><p><strong>Methods: </strong>LC-MS/MS analysis of these impurities was performed on Hypersil BDS C18 (100 mm x 4.0 mm, 3 µm) column. 5 mmol ammonium acetate and acetonitrile in the ratio of 65:35 (v/v) was used as the mobile phase with a flow rate of 0.4 ml/min. The developed method was accomplished with a short run time of 10 min. Triple quadrupole mass detector coupled with positive electrospray ionization was used for the quantification of genotoxic impurities in multiple reaction monitoring (MRM).</p><p><strong>Results: </strong>The method was validated as per International Conference on Harmonization (ICH) guidelines. The method was linear in the range of 0.30 µg/g to 11.37 µg/g for impurity A and 0.30 µg/g to 11.34 µg/g for impurity B with a correlation coefficient of 0.999. The accuracy of the method was in the range of 98.25 % to 100.53 % for both impurities.</p><p><strong>Conclusion: </strong>A specific, selective, highly sensitive and more accurate analytical method using LC-MS/MS coupled with positive electrospray ionization has been developed for the quantification of genotoxic impurity (1-[2-(2,4-difluorophenyl)-2,3-epoxypropyl]-1H-1,2,4-triazole) and its precursor (1-(2,4-difluorophenyl)-2-[1,2,4]triazol-1-yl-ethanone) at 0.3 µg/g with respect to the 5.0 mg/ml of fluconazole.</p>


Sign in / Sign up

Export Citation Format

Share Document