Accuracy and sensitivity of three-dimensional echocardiography to detect changes in right ventricular volumes: comparison study with cardiac magnetic resonance

Author(s):  
K. A. Myhr ◽  
C. B. Kristensen ◽  
F. H. G. Pedersen ◽  
C. Hassager ◽  
N. Vejlstrup ◽  
...  
2021 ◽  
Vol 8 ◽  
Author(s):  
Ashfaq Ahmad ◽  
He Li ◽  
Xiaojing Wan ◽  
Yi Zhong ◽  
Yanting Zhang ◽  
...  

Background: A novel, fully automated right ventricular (RV) software for three-dimensional quantification of RV volumes and function was developed. The direct comparison of the software performance with cardiac magnetic resonance (CMR) was limited. Therefore, the aim of this study was to test the feasibility, accuracy, and reproducibility of a fully automated RV quantification software against CMR imaging as a reference.Methods: A total of 170 patients who underwent both CMR and three-dimensional echocardiography were enrolled. RV end-diastolic volume (RVEDV), RV end-systolic volume (RVESV), and RV ejection fraction (RVEF) were obtained using fully automated three-dimensional RV quantification software and compared with a CMR reference. For inter-technical agreement, Spearman correlation and Bland–Altman analysis were used.Results: The fully automated RV quantification software was feasible in 149 patients. RVEDV and RVESV were underestimated, and RVEF was overestimated compared with CMR values. RV measurements obtained from the manual editing method correlated better with CMR values than that without manual editing (RVEDV, 0.924 vs. 0.794: RVESV, 0.955 vs. 0.854; RVEF, 0.941 vs. 0.781 respectively, all p < 0.0001) with less bias and narrower limit of agreement (LOA). The bias and LOA for RV volumes and EF using the automated software without and with manual editing were greater in patients with severely impaired RV function or low frame rate than those with normal and mild impaired RV function, or high frame rate. The fully automated RV three-dimensional measurements were highly reproducible.Conclusion: The novel fully automated RV software shows good feasibility and reproducibility, and the measurements had a high correlation with CMR values. These findings support the routine application of the novel 3D automated RV software in clinical practice.


2021 ◽  
Vol 17 (1) ◽  
pp. 71-83
Author(s):  
Halszka Kamińska ◽  
Łukasz A. Małek ◽  
Marzena Barczuk-Falęcka ◽  
Bożena Werner

IntroductionThe role of three-dimensional echocardiography (3D-ECHO) chamber quantification in children is still underestimated.Material and methodsIn 43 children 3D-ECHO measurements of end-diastolic (EDV) and end-systolic ventricular volumes (ESV) were compared to cardiac magnetic resonance (CMR) using Bland-Altman analysis and linear regression. The values of left and right ventricular volumes calculated in 3D-ECHO were compared with each other and verified by CMR.ResultsThe values of LV-EDV and LV-ESV measured in 3D-ECHO showed highly significant correlations with CMR (for LV-EDV r = 0.892, p < 0.00001; for LV-ESV r = 0.896, p < 0.00001). In the case of the right ventricle the correlation of 3D-ECHO results with CMR was still high (RV-EDV r = 0.848, p < 0.00001, RV-ESV r = 0.914, p < 0.00001), although mean RV-EDV and RV-ESV in 3D-ECHO were underestimated compared to CMR (by 38% for RV-EDV and 45% for RV-ESV). Correction of 3D-ECHO results using the coefficient of 1.38 and 1.45 for RV-EDV and RV-ESV, respectively, significantly improved the consistency of the results with CMR. 3D-ECHO offered lower mean values of right ventricular volumes compared to the left ventricle. The discrepancy was again reduced by the calculated coefficients.Conclusions3D-ECHO is a valuable tool for assessment of left ventricular volume, which strongly correlates and agrees with CMR. The right ventricular volumes calculated in 3D-ECHO tend to be significantly underestimated in comparison to CMR and corresponding left ventricular volumes obtained from 3D-ECHO. The use of coefficients developed by the study improves the consistency of right ventricular volumes measured by 3D-ECHO with results obtained by CMR and reduces the volumetric discrepancy between ventricles in 3D-ECHO.


Sign in / Sign up

Export Citation Format

Share Document