HIGH RESOLUTION X-RAY SPECTROSCOPY WITH THE DOUBLE CRYSTAL SPECTROMETER

1987 ◽  
Vol 48 (C9) ◽  
pp. C9-83-C9-86
Author(s):  
W. C. SAUDER
1992 ◽  
Vol 63 (10) ◽  
pp. 5022-5022
Author(s):  
R. Barnsley ◽  
A. Abbey ◽  
J. Dunn ◽  
S. Lea ◽  
N. J. Peacock

2001 ◽  
Vol 72 (6) ◽  
pp. 2566-2574 ◽  
Author(s):  
J. Weinheimer ◽  
I. Ahmad ◽  
O. Herzog ◽  
H.-J. Kunze ◽  
G. Bertschinger ◽  
...  

2018 ◽  
Vol 25 (5) ◽  
pp. 1541-1547 ◽  
Author(s):  
Jagannath ◽  
U. K. Goutam ◽  
R. K. Sharma ◽  
J. Singh ◽  
K. Dutta ◽  
...  

The Hard X-ray Photo-Electron Spectroscopy (HAXPES) beamline (PES-BL14), installed at the 1.5 T bending-magnet port at the Indian synchrotron (Indus-2), is now available to users. The beamline can be used for X-ray photo-emission electron spectroscopy measurements on solid samples. The PES beamline has an excitation energy range from 3 keV to 15 keV for increased bulk sensitivity. An in-house-developed double-crystal monochromator [Si (111)] and a platinum-coated X-ray mirror are used for the beam monochromatization and manipulation, respectively. This beamline is equipped with a high-energy (up to 15 keV) high-resolution (meV) hemispherical analyzer with a microchannel plate and CCD detector system with SpecsLab Prodigy and CasaXPS software. Additional user facilities include a thin-film laboratory for sample preparation and a workstation for on-site data processing. In this article, the design details of the beamline, other facilities and some recent scientific results are described.


Sign in / Sign up

Export Citation Format

Share Document