Flow Injection On-Line Preconcentration and Flame AAS Determination of Copper, Cadmium and Lead in Marine Sediment Samples

1997 ◽  
Vol 30 (5) ◽  
pp. 1037-1050 ◽  
Author(s):  
S. Kartikeyan ◽  
B. Vijayalekshmy ◽  
S. Chandramouleeswaran ◽  
T. P. Rao ◽  
C. S. P. Iyer
1995 ◽  
Vol 17 (1) ◽  
pp. 25-29 ◽  
Author(s):  
J. L. Burguera ◽  
M. Burguera ◽  
P. Carrero ◽  
J. Marcano ◽  
C. Rivas ◽  
...  

The paper describes the development of an automatic on-line column pre-concentration technique using a time based-flow injection atomic absorption spectrometry system. A manifold incorporating a micro-column containing 25 mg of Dowex 50W-X8 was used with a time-based injector for the pre-concentration and determination of copper in natural and drinking waters. The system features depend on the alternate positions of a solenoid valve. The 3σ detection limits, enrichment factors, sampling frequency, relative standard deviations and linear calibration graphs were, respectively, in the range 0.6-1.5 μg/l, 25-60, 15-30 measurements/h, 1.0-3.1% and 1-65 μg/ml for pre-concentration times of 1 min. The procedure was successfully applied to a range of water samples and the accuracy was assessed through recovery experiments, the analysis of certified reference water samples and by independent analysis by atomic absorption spectrometry with electrothermal atomization.