A new integrating sphere design for spectral radiant flux determination of light-emitting diodes

2009 ◽  
Vol 20 (9) ◽  
pp. 095111 ◽  
Author(s):  
P Hanselaer ◽  
A Keppens ◽  
S Forment ◽  
W R Ryckaert ◽  
G Deconinck
1989 ◽  
Vol 20 (5) ◽  
pp. 205-217
Author(s):  
J Deforges ◽  
P Garcia ◽  
J Bastie ◽  
F Marandet ◽  
J Bernard ◽  
...  

2021 ◽  
Author(s):  
C. Yuqin Zong ◽  
Cameron Miller

We have developed a new calibration capability for 200 nm to 400 nm ultraviolet light-emitting diodes (UV LEDs) using a Type D gonio-spectroradiometer. The recently-introduced mean differential continuous pulse (M-DCP) method is used to overcome the measurement difficulty associated with the initial forward voltage, VF, anomaly of a UV LED, which makes it impossible to use VF to infer junction temperature, TJ, during pulsed operation. The new measurement facility was validated indirectly by comparing the measured total luminous flux of a white LED with that measured using the NIST’s 2.5 m absolute integrating sphere. The expanded calibration uncertainty for the total radiant flux is approximately 2 % to 3 % (k = 2) depending the wavelength of the UV LED.