Use of distal protection filter device for rescue mechanical thrombectomy for symptomatic free-floating thrombus of the cervical internal carotid artery
Background Intraluminal free floating thrombus (FFT) of the cervical internal carotid artery (ICA) is a potentially high-risk lesion for recurrent ischemic strokes, with no single optimal treatment established. Objective To describe a novel interventional technique for mechanical thrombectomy of intraluminal free floating thrombus using a distal protection filter device and aspiration. Methods We identified two cases where anticoagulation was contraindicated in patients with intraluminal FFT and minimal underlying stenosis. In both cases, a distal protection device was used to extract the thrombus under proximal flow arrest and aspiration. Results Mechanical thrombectomy was successful in both cases, one with “en bloc” removal of the thrombus and one with piecemeal removal. Neither case showed any residual angiographic stenosis or atherosclerotic disease. Conclusion Mechanical thrombectomy using a distal protection device is a novel and technically feasible approach for select cases of cervical intraluminal thrombus. This may represent a therapeutic option in select cases with high risk of anticoagulation and in the unusual subset of younger patients with minimal atherosclerotidc disease.