Agreement between anterior segment Optical Coherence Tomography, Lenstar and Ultrasound Biomicroscopy measurements of anterior segment parameters
Abstract Purpose To compare anterior chamber depth (ACD) and central corneal thickness (CCT) measurements by anterior segment optical coherence tomography (AS-OCT), Lenstar and ultrasound biomicroscopy (UBM). Methods A retrospective cross section study. A total of 83 eyes from 54 patients were included. Central corneal thickness (CCT) and anterior chamber depth (ACD) values obtained from three devices (AS-OCT, Lenstar and UBM) were recorded. All patients were collected from Oct. 2015 to Aug. 2016 in the Fourth Affiliated Hospital of China Medical University. Differences of these three devices (ACD and CCT) were statistically analyzed by randomized blocks analysis. Pearson correlative analysis and Bland-Altman analysis were used to compare the correlation and agreement. Results There’s no significant difference in measuring ACD by Lenstar and UBM (P>0.05), but there were significant differences in other two groups (AS-OCT and UBM, AS-OCT and Lenstar) (P<0.05). There were significant differences in the mean CCT measured by the three devices (P<0.05). There were high correlation in measuring ACD (all P<0.001, r1=0.928,r2=0.982,r3=0.932, respectively), and CCT(all P<0.001,r1=0.957,r2=0.949,r3=0.928, respectively). Conclusion AS-OCT, Lenstar and UBM yielded comparable results and are all can be used interchangeably for anterior segment measurements.