scholarly journals A New Optical Coherence Tomography Method for Subfoveal Choroidal Thickness Measurement: A Single-center Observational Study

2020 ◽  
Author(s):  
Kyung Tae Kang ◽  
Ji Hye Jang ◽  
Yu Cheol Kim

Abstract Background: Several macular pathologies are associated with choroidal vascular supply, and thus require choroidal thickness evaluation. However, there is no standard subfoveal choroidal thickness (SFCT) measuring method. This study investigated a new method to measure the SFCT, using optical coherence tomography.Methods: This is a single-center observational study. Eighty-two senior medical students participated as observers and independently measured the SFCT on five standard optical coherence tomography images of healthy adults in a tertiary care setting. They used three different methods: (1) the observer’s own method, (2) the conventional method, perpendicular to the retinal pigment epithelium, and (3) the new method, along a virtual line connecting the umbo with the most elevated point of the ellipsoid. Additionally, the SFCT angle—the angle between the measurement line and the vertical line of the image—was measured and compared between methods. The intraclass correlation coefficient was used to determine interpersonal variability. Results: The intraclass correlation coefficients for SFCT measured by methods 1, 2, and 3 were 0.853, 0.880, and 0.896, respectively (P < 0.001 for all). It was the highest with method 3; the intraclass correlation coefficients SFCT angles were 0.647, 0.842, and 0.307, respectively (P < 0.001 for all).Conclusions: The new method showed the lowest interpersonal variability, and could therefore be a reliable standard for SFCT measurement, even in foveae with a steep slope on optical coherence tomography.Trial registration: Not applicable

Author(s):  
Daniel Krause ◽  
Niklas Mohr ◽  
Mehdi Shajari ◽  
Wolfgang J. Mayer ◽  
Siegfried Priglinger ◽  
...  

Abstract Purpose To evaluate the reliability of spectral-domain optical coherence tomography (SD-OCT; RTVue XR; Optovue, Inc., Fremont, CA, USA) for thickness mapping of the entire cornea (CT), corneal epithelium (ET). and corneal stroma (ST) over a 9-mm zone in healthy eyes. We sought to develop reference values for different age groups and elucidate potential sex- and age-dependent characteristics of corneal sublayer pachymetry maps. Methods Three consecutive SD-OCT scans were obtained in 166 healthy right eyes (mean age = 50 ± 20 years). The thickness maps contain 25 sectors over a 9-mm diameter zone. To test measurement reliability, intraclass correlation coefficients (ICC), coefficients of variation (CoV), and within-subject standard deviations (WSSD) were calculated. Results CT, ET, and ST ICCs ranged from 0.961 to 0.998, 0.896 to 0.945, and 0.955 to 0.998, respectively. CoV values for CT, ET, and ST ranged between 0.3 and 1.5%, 1.6 and 4.2%, and 0.4 and 1.7%, respectively. WSSD ranged from 6 to 41, 4 to 8, and 7 to 46 µm, respectively. A negative correlation was found between age and ET (p < 0.05) but not between age and ST or CT. No gender-related differences in CT, ET, or ST were detected. CoV of CT, ET, and ST measurements showed a positive correlation with age in 28, 64, and 28% of the sectors, respectively. Conclusion SD-OCT is a rapid and noninvasive technique that provides excellent reliability for corneal sublayer thickness measurements over a 9-mm zone. The reliability of the ET measurement seems to be negatively affected by age. Peripheral CT and global ET thin with age.


2019 ◽  
Vol 8 (10) ◽  
pp. 1732 ◽  
Author(s):  
Matthew S. Pihlblad ◽  
Andrew Troia ◽  
Sapna Tibrewal ◽  
Parth R. Shah

OCT (optical coherence tomography) is widely used in ophthalmology and pediatric ophthalmology, but limited research has been done on the use of OCT in strabismus. This study investigates the use of different OCT machines to image rectus muscle insertions pre-, intra-, and post-operatively in pediatric strabismus patients. The OCT machines used in the study were a Bioptigen (Leica Microsystems Inc., Buffalo Grove, IL, USA), Spectralis HRA+OCT with Anterior Segment Module (Heidelberg Engineering, Heidelberg, Germany), Visante (Carl Zeiss, Oberkochen, Germany), and Zeiss Rescan 700 (Carl Zeiss, Oberkochen, Germany). Measurements from the machines were compared with the caliper distance measured during the strabismus surgery before disinsertion or after reattachment. The OCT machines had moderate (Bioptigen: 0.62) to good intraclass correlation coefficients (Rescan: 0.83, Spectralis: 0.85, Visante: 0.88) with intra-operative measurements. To our knowledge, this is the first study to use an operating microscope with integrated intra-operative OCT to image rectus muscle insertions. OCT is a useful tool in strabismus surgical patients in the pre-, intra-, and post-operative settings, particularly in patients who have had previous surgery, when the muscle insertion is unknown. The ability to accurately image rectus muscle insertions has significant implications for the management of strabismus patients.


2021 ◽  
Vol 10 (2) ◽  
pp. 174
Author(s):  
Katharina Breher ◽  
Alejandro Calabuig ◽  
Laura Kühlewein ◽  
Focke Ziemssen ◽  
Arne Ohlendorf ◽  
...  

Retinal shape presents a clinical parameter of interest for myopia, and has commonly been inferred indirectly from peripheral refraction (PRX) profiles. Distortion-corrected optical coherence tomography (OCT) scans offer a new and direct possibility for retinal shape estimation. The current study compared retinal curvatures derived from OCT scans vs. PRX measurements in three refractive profiles (0° and 90° meridians, plus spherical equivalent) for 25 participants via Bland–Altman analysis. The radial differences between both procedures were correlated to axial length using Pearson correlation. In general, PRX- and OCT-based retinal radii showed low correlation (all intraclass correlation coefficients < 0.21). PRX found flatter retinal curvatures compared to OCT, with the highest absolute agreement found with the 90° meridian (mean difference +0.08 mm) and lowest in the 0° meridian (mean difference +0.89 mm). Moreover, a negative relation between axial length and the agreement of both methods was detected especially in the 90° meridian (R = −0.38, p = 0.06). PRX measurements tend to underestimate the retinal radius with increasing myopia when compared to OCT measurements. Therefore, future conclusions from PRX on retinal shape should be made cautiously. Rather, faster and more clinically feasible OCT imaging should be performed for this purpose.


2011 ◽  
Vol 2011 ◽  
pp. 1-5 ◽  
Author(s):  
Ning Fan ◽  
Nina Huang ◽  
Dennis Shun Chiu Lam ◽  
Christopher Kai-shun Leung

Objective.To measure and compare photoreceptor layer thickness between normal and glaucomatous eyes using spectral-domain optical coherence tomography (OCT).Methods. Thirty-eight healthy normal volunteers and 47 glaucoma patients were included in the analysis. One eye from each participant was randomly selected for macula imaging by a spectral-domain OCT (3D OCT-1000, Topcon, Tokyo, Japan). The foveal and parafoveal (1.5 mm from the fovea) outer nuclear layer (ONL) and inner and outer segments (IS+OS) layer thicknesses were measured by a single masked observer. The measurements were repeated 3 times in a random sample of 30 normal eyes to determine the repeatability coefficient and intraclass correlation coefficient.Results.The measurement variabilities of photoreceptor thickness were low. The respective intraclass correlation coefficients of ONL and IS+OS thicknesses were 0.96 (95% confidence interval: 0.94–0.98) and 0.82 (95% confidence interval 0.70–0.90). While there were no differences in parafoveal ONL and IS+OS thicknesses between normal and glaucoma groups (), the foveal ONL thickness was greater in glaucomatous eyes () than in normal eyes.Conclusions.Glaucomatous damage may involve structural change in the photoreceptor layer.


2019 ◽  
Vol 2019 ◽  
pp. 1-7 ◽  
Author(s):  
Jingyuan Yang ◽  
Mingzhen Yuan ◽  
Erqian Wang ◽  
Youxin Chen

The aim of this study was to compare the repeatability of optical coherence tomography angiography (OCT-A) measurements of macular vessel density using four OCT-A systems, including Heidelberg Spectralis HRA, Optovue RTVue XR, Zeiss Cirrus HD-OCT 5000, and Topcon DRI OCT Triton. A cross-sectional design was used for this study. The vascular density and vascular length density of the superficial and deep retinal capillary plexuses were imaged with OCT-A using 3 mm and 6 mm scan patterns and were calculated using ImageJ. Comparisons of intraclass correlation coefficients (ICC) were conducted. We found that the OCT-A systems had various levels of repeatability. Zeiss had better repeatability for vessel density than the other systems (overall ICC = 0.936). Optovue had better repeatability for vessel length density when the 6 mm scan pattern was used (ICC = 0.680 and 0.700 for retinal superficial and deep capillary plexus, respectively). We concluded that repeatability varied when different scan patterns of various OCT-A systems were used for imaging the superficial retinal and deep capillary plexuses. Results should be seen as valid only for a given method. The repeatability of various OCT-A systems should be considered in clinical practice and in clinical trials that use OCT-A metrics as outcome measures.


2020 ◽  
pp. bjophthalmol-2020-316512
Author(s):  
Mingui Kong ◽  
Sungsoon Hwang ◽  
Hyeonyoung Ko ◽  
Ga-In Lee ◽  
Moonil Kang ◽  
...  

PurposeTo evaluate genetic influence on macular ganglion cell inner plexiform layer (GCIPL) thickness.MethodsMacular GCIPL thickness was measured with optical coherence tomography in nine macular subfields defined by the E TDRS. Intraclass correlation coefficients (ICC) of GCIPL thickness by different types of family relationships were estimated to assess intrafamilial resemblance. Then, heritability of GCIPL thickness was estimated.ResultsThree hundred and sixty-one Korean adults from 89 families with normal healthy eyes were included. GCIPL thickness was highest in inner subfields and lowest in fovea. Monozygotic twin pairs showed significantly higher ICCs of GCIPL thickness in all subfields compared to those in parent–offspring pairs and sibling pairs. GCIPL thickness was highly heritable in the centre (0.71) and outer subfields but moderate to highly heritable in inner subfields. Heritability of GCIPL thickness in outer subfields was 0.69, 0.67, 0.72 and 0.68 for superior, inferior, temporal and nasal fields, respectively. Heritability of GCIPL thickness in inner subfields was 0.55, 0.56, 0.75 and 050 for superior, inferior, temporal and nasal subfields, respectively.ConclusionMacular GCIPL thickness is significantly influenced by genetic factors. It varies according to subfields with moderate to high heritability in all subfields.


Author(s):  
Jeniffer Trenado ◽  
Sergio Rojas juárez

Introduction: The purpose was to assess the level of agreement and reproducibility between fluorescein angiography (FA) and optical coherence tomography angiography (OCTA), throughout the evaluation of the foveal avascular zone (FAZ) to diagnose diabetic macular ischemia (DMI). Methods: In this ambispective, observational, cross-sectional research, 41 patients underwent traditional FA and OCTA to obtain the level of agreement and reproducibility between them, using an intraclass correlation coefficient (ICC) in mixed models. Two raters independently graded the area of the FAZ using the ImageJ software and the Early Treatment Diabetic Retinopathy Study protocols. Spearman and rank-biserial correlational analysis were used to calculate the strength of linear relationship between the area of the retinal vessels, the thickness of subfoveal retinal and choroid layers, the presence of neovascularization, and the visual acuity. A classification, based on the radius of the FAZ, was proposed and used to categorize the severity of DMI into five different grades. Result: From 52 evaluated eyes, the level of agreement between the OCTA and FA among the raters had an ICC of 0.99 and 0.907 (p<.001), respectively. Furthermore, the reproducibility analysis had an ICC of 0.85 (p<.001). In the correlation analysis, the enlargement of the FAZ was associated with a decrease in the outer retinal layers (r=-0.458, p<0.001), the photoreceptor layer (r=-0.32, p=0.021), their outer segments (r=-0.32, p=0.021) and the subfoveal choroidal thickness (r=-0.483, p<0.001). A reduction of the vascular area was observed in higher grades of diabetic retinopathy (r= -0.395, p=0.38). Additionally, meager choroidal thickness was found in grater grades of DMI (r=-0.461, p=0.014). Conclusion: Compelling evidence of a high level of agreement and reproducibility between OCTA and FA was obtained to diagnose diabetic macular ischemia. Moreover, it is suggested that the increase in macular ischemia leads to the thinning of the photoreceptor, outer retinal and choroid layers.


2005 ◽  
Vol 15 (4) ◽  
pp. 486-492 ◽  
Author(s):  
M. Olmedo ◽  
C. Cadarso-Suarez ◽  
F. Gomez-Ulla ◽  
C. Val ◽  
I. Fernandez

Purpose To evaluate reproducibility of optic nerve head (ONH) morphometry measurements obtained by optical coherence tomography (the fast optical disk protocol of the Zeiss model 3000 OCT system) in normal and glaucomatous eyes. Methods Prospective instrument-evaluation study. ONH measurements for 20 eyes were obtained in eight scan sessions taken during two visits to an ophthalmology clinic (10 normal patients, 10 glaucoma patients, one eye per subject). At every one of the eight sessions for each eye, estimates of eight ONH morphometry variables (see Main outcome measures) were obtained. The first two sessions were performed by two operators, followed by a 30-minute break. The same operators then completed a third and fourth session. This sequence was duplicated on a second visit. Intrasession, intersession, intervisit, and interoperator reproducibility of the eight variables were calculated by the use of a components variance model. Intraclass correlation coefficients (ICC) were used to assess reliability. Results Vertical integrated rim area, horizontal integrated rim area, disk area, cup area, rim area, cup/disk area ratio, cup/disk horizontal ratio, cup/disk vertical ratio. With the exception of the horizontal integrated rim area and rim area in normal subjects, the factor subject was the most important source of variance for all variables. Reliability values as measured by ICC for normal eyes were above 81%, with the exception of measurements of the horizontal integrated rim area (23.1%), rim area (33.3%), and disk area (64.7%). For glaucomatous eyes all values were above 85%, with the exception of the disk area (68.1%). Conclusions ONH measurements obtained using the fast optical disk protocol of the Zeiss 3000 OCT system show good reproducibility, for both normal and glaucomatous eyes.


2016 ◽  
Vol 27 (3) ◽  
pp. 336-341 ◽  
Author(s):  
Rodolfo Mastropasqua ◽  
Lisa Toto ◽  
Peter A. Mattei ◽  
Marta Di Nicola ◽  
Isaia A.L. Zecca ◽  
...  

Purpose To assess the reproducibility and repeatability of foveal avascular zone (FAZ) area measurements using swept-source optical coherence tomography angiography (SS-OCTA) in healthy subjects. Methods Sixty-four eyes of 64 healthy volunteers were randomly subjected to FAZ area measurements using SS-OCTA by 2 examiners in 2 different sessions. Results The FAZ areas measured by the first and second observer were 0.269 ± 0.092 mm2 and 0.270 ± 0.090 mm2, respectively. Within subjects, the coefficients of variations were 2.44% (95% confidence interval [CI] 1.95% to 2.93%) and 2.66% (95% CI 2.00% to 3.31%) for the first and second observers, respectively. The coefficient of repeatability average measurements of FAZ area were 0.021 mm2 and 0.024 mm2. The intraclass correlation coefficient values were 0.993 (95% CI 0.989 to 0.996) and 0.991 (95% CI 0.986 to 0.995). Interobserver and intraobserver concordance correlation coefficients ranged from 0.998 (95% CI 0.997 to 0.999) to 0.999 (95% CI 0.998 to 0.999) and from 0.989 (95% CI 0.982 to 0.993) to 0.987 (95% CI 0.979 to 0.992), respectively. Conclusions The FAZ area measurements by means of SS-OCTA showed high reproducibility and repeatability in healthy eyes.


2021 ◽  
Author(s):  
Özgür Eroğul ◽  
Mehmet Murat Uzel ◽  
Leyla Eryiğit Eroğul ◽  
Ayşe Güzin Taşlıpınar Uzel ◽  
Hamidu Hamısı Gobeka ◽  
...  

Abstract PurposeTo investigate the effects of hemodialysis (HD) on the reproducibility of subfoveal choroidal thickness (SFCT) as measured by spectral domain-optical coherence tomography (SD-OCT)MethodsIn this study, 26 HD (26 eyes) patients had their pre- and post-HD SFCT measured, and the results were compared for reproducibility. Following a thorough ophthalmic examination, SD-OCT was performed three times in a row during a single session. The same physician measured SFCT after automatically identifying choroid with a software caliper. The reproducibility parameters, including intra-class correlation coefficients (ICCs), coefficients of variation (COV), and test-retest variability (TRTV) were then calculated.ResultsMales made up 53.85% of the 26 HD patients. There was a significant IOP difference between pre-HD (16.42±3.14 mmHg) and post-HD (14.21±2.78 mmHg) (P<0.001). SFCT decreased significantly from pre-HD 243.50±10.23 μm to post-HD 234.29±9.41 μm (P<0.001). ICC value increased significantly after HD, rising from 0.948 to 0.989 (P<0.001, for all). Pre- and post-HD COV values were 1.6% and 0.65%, respectively. Also, pre- and post-HD TRTV values were 7.864±1.996 μm and 3.074±1.536 μm, respectively.ConclusionThe reproducibility of SFCT as measured by OCT was lower during pre-HD compared to post-HD. Post-HD SD-OCT assessment appears to improve the reliability of clinical outcomes in the diagnosis and monitoring of HD patients.