scholarly journals Can Gafchromic EBT3 films effectively characterize small fields of 6 MV unflattened photon beams of Cyberknife system?

2018 ◽  
Vol 24 (4) ◽  
pp. 181-187
Author(s):  
Jerrin Amalraj ◽  
Ramasubramanian Velayudham ◽  
Pichandi Anchineyan

Abstract Shielded silicon diodes are commonly employed in commissioning of Cyberknife 6 MV photon beams. This study aims to measure output factors, off centered ratio (OCR), percentage depth dose (PDD) of 6 MV photons using shielded and unshielded diodes and to compare with Gafchromic EBT3 film measurements to investigate whether EBT3 could effectively characterize small 6 MV photon beams. Output factors, OCR and PDD were measured with shielded and unshielded silicon detectors in a radiation field analyzer system at reference condition. Water equivalent solid phantom were used while irradiating EBT3 films. From multiuser data, diodes underestimated output factor by 3% for collimator fields ≤ 10 mm, while EBT3 underestimated the output factor by 3.9% for 5 mm collimator. 1D Gamma analysis of OCR between diode and film, results in gamma ≤ 1 for all measured points with 1 mm distance to agreement (DTA) and 1% relative dose difference (DD). Dose at surface is overestimated with diodes compared to EBT3. PDD results were within 2% relative dose values between diode and EBT3 except for 5 mm collimator. Except for small collimator fields of up to 10 mm, results of output factor, OCR, PDD of all detectors used in this study exhibited similar results. Relative dose measurements with Gafchromic EBT3 in this work show that EBT3 films can be used effectively as an independent tool to verify commissioning beam data of small fields only after careful verification of methodology for any systematic errors with appropriate readout procedure.

2018 ◽  
Vol 18 (03) ◽  
pp. 280-284 ◽  
Author(s):  
Mohamed Bencheikh ◽  
Abdelmajid Maghnouj ◽  
Jaouad Tajmouati

AbstractAimThe purpose of this study is to introduce a new approach to assess the dosimetry quality of photon beam with energy and irradiation field size. This approach is based on percentage depth dose (PDD) fragmentation for investigating the dosimetry quality.Materials and methodsFor the investigation of the dosimetry quality of 6 and 18 MV photon beams, we have proceeded to fragment the PDD at different field sizes. This approach checks the overall PDD and is not restricted to the exponential decay regions, as per the International Atomic Energy Agency Technical Reports Series No 398 and the American Association of Physicist in Medicine Task Group 51 recommendations.Results and discussionThe 6 MV photon beam deposited more energy in the target volume than the 18 MV photon beam. The dose delivered by the 6 MV beam is greater by a factor of 1·5 than that delivered by the 18 MV beam in the build-up region and the dose delivered by the 6 MV beam is greater by a factor of 2·6 than that delivered by the 18 MV beam in the electronic equilibrium and the exponential decay regions.ConclusionThe dose measured at different points of the beam is higher for 6 MV than for 18 MV photon beam. Therefore, the 6 MV beam is more dosimetrically efficient than the 18 MV beam. Using the proposed approach, we can assess the dosimetry quality by taking into account overall PDD not only in the exponential decay region but also in the field.


2022 ◽  
Vol 12 (2) ◽  
pp. 600
Author(s):  
Serenella Russo ◽  
Silvia Bettarini ◽  
Barbara Grilli Leonulli ◽  
Marco Esposito ◽  
Paolo Alpi ◽  
...  

High-energy small electron beams, generated by linear accelerators, are used for radiotherapy of localized superficial tumours. The aim of the present study is to assess the dosimetric performance under small radiation therapy electron beams of the novel PTW microSilicon detector compared to other available dosimeters. Relative dose measurements of circular fields with 20, 30, 40, and 50 mm aperture diameters were performed for electron beams generated by an Elekta Synergy linac, with energy between 4 and 12 MeV. Percentage depth dose, transverse profiles, and output factors, normalized to the 10 × 10 cm2 reference field, were measured. All dosimetric data were collected in a PTW MP3 motorized water phantom, at SSD of 100 cm, by using the novel PTW microSilicon detector. The PTW diode E and the PTW microDiamond were also used in all beam apertures for benchmarking. Data for the biggest field size were also measured by the PTW Advanced Markus ionization chamber. Measurements performed by the microSilicon are in good agreement with the reference values for all the tubular applicators and beam energies within the stated uncertainties. This confirms the reliability of the microSilicon detector for relative dosimetry of small radiation therapy electron beams collimated by circular applicators.


2021 ◽  
Vol 27 (1) ◽  
pp. 25-29
Author(s):  
Labinot Kastrati ◽  
Gezim Hodolli ◽  
Sehad Kadiri ◽  
Elvin Demirel ◽  
Lutfi Istrefi ◽  
...  

Abstract Introduction: The aim of this study is to analyze the gradient of percentage depth dose for photon and electron beams of LINACs and to simplify the data set. Materials and Methods: Dosimetry measurements were performed in accordance with Technical Reports Series No. 398 IAEA. Results and discussion: The gradient of percentage depth dose was calculated and compared with the available published data. Conclusion: Instead of percentage depth dose for increasing and decreasing parts, the findings suggest using only two numbers for specific gradient of dose, separately. In this way, they can replace the whole set of the percentage depth dose (PDD).


2017 ◽  
Vol 13 (4) ◽  
pp. 4901-4907
Author(s):  
Ehab A. Hegazy

Many radiotherapy centers don’t pay attention to effect of ionization champers type on accuracy of quality control measurements. They use any available ionization champers in all quality control and data entry measurements (1,2) . Many studies were carried out in this field to compare different ionization champers in small fields but large fields were not completely compared before (3.4).  The aim of this work is to compare   output factor in large field using two different sizes ionization champers connected to electrometer. Final out put were obtained from the farmer and smidflex dosimeter irradiated with 6 MV photon beams. Important field side ranging from 20 cm to 70 cm side field is measured in whole body radiation (5, 6). For all examined large  field sizes a difference ranging from 1% to 5 % was found when added to other calibration errors it will exceeds the acceptable margin. The largest difference was found in field side 70 cm this may be due to large scattering radiation


2019 ◽  
Vol 25 (2) ◽  
pp. 101-110
Author(s):  
Itumeleng Setilo ◽  
Oluwaseyi Michael Oderinde ◽  
Freek Cp du Plessis

Abstract Introduction: Small fields photon dosimetry is associated with many problems. Using the right detector for measurement plays a fundamental role. This study investigated the measurement of relative output for small photon fields with different detectors. It was investigated for three-photon beam energies at SSDs of 90, 95, 100 and 110 cm. As a benchmark, the Monte Carlo simulation was done to calculate the relative output of these small photon beams for the dose in water. Materials and Methods: 6, 10 and 15 MV beams were delivered from a Synergy LINAC equipped with an Agility 160 multileaf collimator (MLC). A CC01 ion chamber, EFD-3G diode, PTW60019 microdiamond, EBT2 radiochromic film, and EDR2 radiographic film were used to measure the relative output of the linac. Measurements were taken in water for the CC01 ion chamber, EFD-3G diode, and the PTW60019. Films were measured in water equivalent RW3 phantom slabs. Measurements were made for 1 × 1, 2 × 2, 3 × 3, 4 × 4, 5 × 5 and a reference field of 10 × 10 cm2. Field sizes were defined at 100cm SSD. Relative output factors were also compared with Monte Carlo (MC) simulation of the LINAC and a water phantom model. The influence of voxel size was also investigated for relative output measurement. Results and Discussion: The relative output factor (ROF) increased with energy for all fields large enough to have lateral electronic equilibrium (LEE). This relation broke down as the field sizes decreased due to the onset of lateral electronic disequilibrium (LED). The high-density detector, PTW60019 gave the highest ROF for the different energies, with the less dense CC01 giving the lowest ROFs. Conclusion: These are results compared to MC simulation, higher density detectors give higher ROF values. Relative to water, the ROF measured with the air-chamber remained virtually unchanged. The ROFs, as measured in this study showed little variation due to increased SSDs. The effect of voxel size for the Monte Carlo calculations in water does not lead to significant ROF variation over the small fields studied.


2011 ◽  
Vol 38 (6Part22) ◽  
pp. 3667-3667
Author(s):  
R Krauss ◽  
M Morales ◽  
C Coffey ◽  
G Ding