Bioabsorbable anterior lumbar plate fixation in conjunction with cage-assisted anterior interbody fusion

2002 ◽  
Vol 97 (4) ◽  
pp. 447-455 ◽  
Author(s):  
Denis J. DiAngelo ◽  
Jeffrey L. Scifert ◽  
Scott Kitchel ◽  
G. Bryan Cornwall ◽  
Bobby J. McVay

Object. An in vitro biomechanical study was conducted to determine the effects of anterior stabilization on cage-assisted lumbar interbody fusion biomechanics in a multilevel human cadaveric lumbar spine model. Methods. Three spine conditions were compared: harvested, bilateral multilevel cages (CAGES), and CAGES with bioabsorbable anterior plates (CBAP), tested under flexion—extension, lateral bending, and axial rotation. Measurements included vertebral motion, applied load, and bending/rotational moments. Application of anterior fixation decreased local motion and increased stiffness of the instrumented levels. Clinically, this spinal stability may serve to promote fusion. Conclusions. Coupled with the bioabsorbability of the plating material, the bioabsorbable anterior lumbar plating system is considered biomechanically advantageous.

Neurosurgery ◽  
2001 ◽  
Vol 49 (6) ◽  
pp. 1399-1408 ◽  
Author(s):  
Andrzej Maciejczak ◽  
Michał Ciach ◽  
Maciej Radek ◽  
Andrzej Radek ◽  
Jan Awrejcewicz

ABSTRACT OBJECTIVE To determine whether the Cloward technique of cervical discectomy and fusion increases immediate postoperative stiffness of single cervical motion segment after application of interbody dowel bone graft. METHODS We measured and compared the stiffness of single-motion segments in cadaveric cervical spines before and immediately after interbody fusion with the Cloward technique. Changes in range of motion and stiffness of the C5–C6 segment were measured in a bending flexibility test (flexion, extension, lateral bending and axial rotation) before and after a Cloward procedure in 11 fresh-frozen human cadaveric specimens from the 4th through the 7th vertebrae. RESULTS The Cloward procedure produced a statistically significant increase in stiffness of the operated segment in flexion and lateral bending when compared with the intact spine. The less stiff the segment before the operation, the greater the increase in its postoperative flexural stiffness (statistically significant). The Cloward procedure produced nonuniform changes in rotational and extensional stiffness that increased in some specimens and decreased in others. CONCLUSION Our data demonstrate that Cloward interbody fusion increases immediate postoperative stiffness of an operated segment only in flexion and lateral bending in cadaveric specimens in an in vitro environment. Thus, Cloward fusion seems a relatively ineffective method for increasing the stiffness of a construct. This may add to discussion on the use of spinal instrumentation and postoperative management of patients after cervical discectomy, which varies from bracing in hard collars through immobilization in soft collars to no external orthosis.


2019 ◽  
Vol 9 (8) ◽  
pp. 826-833
Author(s):  
Ripul Panchal ◽  
Anup Gandhi ◽  
Chris Ferry ◽  
Sam Farmer ◽  
Jeremy Hansmann ◽  
...  

Study Design: In vitro biomechanical study. Objectives: The objective of this in vitro biomechanical range-of-motion (ROM) study was to evaluate spinal segmental stability following fixation with a novel anterior cervical discectomy and fusion (ACDF) device (“novel device”) that possesses integrated and modular no-profile, half-plate, and full-plate fixation capabilities. Methods: Human cadaveric (n = 18, C3-T1) specimens were divided into 3 groups (n = 6/group). Each group would receive one novel device iteration. Specimen terminal ends were potted. Each specimen was first tested in an intact state, followed by anterior discectomy (C5/C6) and iterative instrumentation. Testing order: (1) novel device (group 1, no-profile; group 2, half-plate; group 3, full-plate); (2) novel device (all groups) with lateral mass screws (LMS); (3) traditional ACDF plate + cage; (4) traditional ACDF plate + cage + LMS. A 2 N·m moment was applied in flexion/extension (FE), lateral bending (LB), and axial rotation (AR) via a kinematic testing machine. Segmental ROM was tracked and normalized to intact conditions. Comparative statistical analyses were performed. Results: Key findings: (1) the novel half- and full-plate constructs provided comparable reduction in FE and LB ROM to that of traditional plated ACDF ( P ≥ .05); (2) the novel full-plate construct significantly exceeded all other anterior-only constructs ( P ≤ .05) in AR ROM reduction; and (3) the novel half-plate construct significantly exceeded the no-profile construct in FE ( P < .05). Conclusions: The novel ACDF device may be a versatile alternative to traditional no-profile and independent plating techniques, as it provides comparable ROM reduction in all principle motion directions, across all device iterations.


2000 ◽  
Vol 92 (1) ◽  
pp. 87-92 ◽  
Author(s):  
Annette Kettler ◽  
Hans-Joachim Wilke ◽  
Rupert Dietl ◽  
Matthias Krammer ◽  
Christianto Lumenta ◽  
...  

Object. The function of interbody fusion cages is to stabilize spinal segments primarily by distracting them as well as by allowing bone ingrowth and fusion. An important condition for efficient formation of bone tissue is achieving adequate spinal stability. However, the initial stability may be reduced due to repeated movements of the spine during everyday activity. Therefore, in addition to immediate stability, stability after cyclic loading is of remarkable relevance; however, this has not yet been investigated. The object of this study was to investigate the immediate stabilizing effect of three different posterior lumbar interbody fusion cages and to clarify the effect of cyclic loading on the stabilization. Methods. Before and directly after implantation of a Zientek, Stryker, or Ray posterior lumbar interbody fusion cage, 24 lumbar spine segment specimens were each evaluated in a spine tester. Pure lateral bending, flexion—extension, and axial rotation moments (± 7.5 Nm) were applied continuously. The motion in each specimen was measured simultaneously. The specimens were then loaded cyclically (40,000 cycles, 5 Hz) with an axial compression force ranging from 200 to 1000 N. Finally, they were tested once again in the spine tester. Conclusions. In general, a decrease of movement in all loading directions was noted after insertion of the Zientek and Ray cages and an increase of movement after implantation of a Stryker cage. In all three cage groups greater stability was demonstrated in lateral bending and flexion than in extension and axial rotation. Reduced stability during cyclic loading was observed in all three cage groups; however, loss of stability was most pronounced when the Ray cage was used.


1999 ◽  
Vol 90 (1) ◽  
pp. 91-98 ◽  
Author(s):  
A. Giancarlo Vishteh ◽  
Neil R. Crawford ◽  
M. Stephen Melton ◽  
Robert F. Spetzler ◽  
Volker K. H. Sonntag ◽  
...  

Object. The authors sought to determine the biomechanics of the occipitoatlantal (occiput [Oc]—C1) and atlantoaxial (C1–2) motion segments after unilateral gradient condylectomy. Methods. Six human cadaveric specimens (skull with attached upper cervical spine) underwent nondestructive biomechanical testing (physiological loads) during flexion—extension, lateral bending, and axial rotation. Axial translation from tension to compression was also studied across Oc—C2. Each specimen served as its own control and underwent baseline testing in the intact state. The specimens were then tested after progressive unilateral condylectomy (25% resection until completion), which was performed using frameless stereotactic guidance. At Oc—C1 for all motions that were tested, mobility increased significantly compared to baseline after a 50% condylectomy. Flexion—extension, lateral bending, and axial rotation increased 15.3%, 40.8%, and 28.1%, respectively. At C1–2, hypermobility during flexion—extension occurred after a 25% condylectomy, during axial rotation after 75% condylectomy, and during lateral bending after a 100% condylectomy. Conclusions. Resection of 50% or more of the occipital condyle produces statistically significant hypermobility at Oc—C1. After a 75% resection, the biomechanics of the Oc—C1 and C1–2 motion segments change considerably. Performing fusion of the craniovertebral junction should therefore be considered if half or more of one occipital condyle is resected.


2000 ◽  
Vol 93 (2) ◽  
pp. 259-265 ◽  
Author(s):  
Masahiro Kanayama ◽  
Bryan W. Cunningham ◽  
Charles J. Haggerty ◽  
Kuniyoshi Abumi ◽  
Kiyoshi Kaneda ◽  
...  

Object. Interbody fusion devices are rapidly gaining acceptance as a method of ensuring lumbar interbody arthrodesis. Although different types of devices have been developed, the comparative reconstruction stability remains controversial. It also remains unclear how different stress-shielded environments are created within the devices. Using a calf spine model, this study was designed to compare the construct stiffness afforded by 11 differently designed lumbar interbody fusion devices and to quantify their stress-shielding effects by measuring pressure within the devices. Methods. Sixty-six lumbar specimens obtained from calves were subjected to anterior interbody reconstruction at L4–5 by using one of the following interbody fusion devices: four different threaded fusion cages (BAK device, BAK Proximity, Ray TFC, and Danek TIBFD), five different nonthreaded fusion devices (oval and circular Harms cages, Brantigan PLIF and ALIF cages, and InFix device); two different types of allograft (femoral ring and bone dowel) were used. Construct stiffness was evaluated in axial compression, torsion, flexion, and lateral bending. Prior to testing, a silicon elastomer was injected into the cages and intracage pressures were measured using pressure needle transducers. Conclusions. No statistical differences were observed in construct stiffness among the threaded cages and nonthreaded devices in most of the testing modalities. Threaded fusion cages demonstrated significantly lower intracage pressures compared with nonthreaded cages and structural allografts. Compared with nonthreaded cages and structural allografts, threaded fusion cages afforded equivalent reconstruction stiffness but provided more stress-shielded environment within the devices.


2005 ◽  
Vol 2 (3) ◽  
pp. 339-343 ◽  
Author(s):  
Patrick W. Hitchon ◽  
Kurt Eichholz ◽  
Christopher Barry ◽  
Paige Rubenbauer ◽  
Aditya Ingalhalikar ◽  
...  

Object. The authors compared the biomechanical performance of the human cadaveric spine implanted with a metallic ball-and-cup artificial disc at L4–5 with the spine's intact state and after anterior discectomy. Methods. Seven human L2—S1 cadaveric spines were mounted on a biomechanical testing frame. Pure moments of 0, 1.5, 3.0, 4.5, and 6.0 Nm were applied to the spine at L-2 in six degrees of motion (flexion, extension, right and left lateral bending, and right and left axial rotation). The spines were tested in the intact state as well as after anterior L4–5 discectomy. The Maverick disc was implanted in the discectomy defect, and load testing was repeated. The artificial disc created greater rigidity for the spine than was present after discectomy, and the spine performed biomechanically in a manner comparable with the intact state. Conclusions. The results indicate that in an in vitro setting, this model of artificial disc stabilizes the spine after discectomy, restoring motion comparable with that of the intact state.


2001 ◽  
Vol 94 (1) ◽  
pp. 97-107 ◽  
Author(s):  
Annette Kettler ◽  
Hans-Joachim Wilke ◽  
Lutz Claes

Object. The aim of this in vitro study was to determine the influence of simulated postoperative neck movements on the stabilizing effect and subsidence of four different anterior cervical interbody fusion devices. Emphasis was placed on the relation between subsidence and spinal stability. Methods. The flexibility of 24 human cervical spine specimens was tested before and directly after being stabilized with a WING, BAK/C, AcroMed I/F cage, or with bone cement in standard flexibility tests under 50 N axial preload. Thereafter, 700 pure moment loading cycles (± 2 Nm) were applied in randomized directions to simulate physiological neck movements. Additional flexibility tests in combination with measurements of the subsidence depth were conducted after 50, 100, 200, 300, 500, and 700 loading cycles. In all four groups, simulated postoperative neck movements caused an increase of the range of motion (ROM) ranging from 0.4 to 3.1° and of the neutral zone from 0.1 to 4.2°. This increase in flexibility was most distinct in extension followed by flexion, lateral bending, and axial rotation. After cyclic loading, ROM tended to be lower in the group fitted with AcroMed cages (3.3° in right lateral bending, 3.5° in left axial rotation, 7.8° in flexion, 8.3° in extension) and in the group in which bone cement was applied (5.4°, 2.5°, 7.4°, and 8.8°, respectively) than in those fixed with the WING (6.3°, 5.4°, 9.7°, and 6.9°, respectively) and BAK cages (6.2°, 4.5°, 10.2°, and 11.6°, respectively). Conclusions. Simulated repeated neck movements not only caused an increase of the flexibility but also subsidence of the implants into the adjacent vertebrae. The relation between flexibility increase and subsidence seemed to depend on the implant design: subsiding BAK/C cages partially supported stability whereas subsiding WING cages and AcroMed cages did not.


2020 ◽  
pp. 219256822090561
Author(s):  
Ryan DenHaese ◽  
Anup Gandhi ◽  
Chris Ferry ◽  
Sam Farmer ◽  
Randall Porter

Study Design: In vitro cadaveric biomechanical study. Objective: Biomechanically characterize a novel lateral lumbar interbody fusion (LLIF) implant possessing integrated lateral modular plate fixation (MPF). Methods: A human lumbar cadaveric (n = 7, L1-L4) biomechanical study of segmental range-of-motion stiffness was performed. A ±7.5 Nċm moment was applied in flexion/extension, lateral bending, and axial rotation using a 6 degree-of-freedom kinematics system. Specimens were tested first in an intact state and then following iterative instrumentation (L2/3): (1) LLIF cage only, (2) LLIF + 2-screw MPF, (3) LLIF + 4-screw MPF, (4) LLIF + 4-screw MPF + interspinous process fixation, and (5) LLIF + bilateral pedicle screw fixation. Comparative analysis of range-of-motion outcomes was performed between iterations. Results: Key biomechanical findings: (1) Flexion/extension range-of-motion reduction with LLIF + 4-screw MPF was significantly greater than LLIF + 2-screw MPF ( P < .01). (2) LLIF with 2-screw and 4-screw MPF were comparable to LLIF with bilateral pedicle screw fixation in lateral bending and axial rotation range-of-motion reduction ( P = 1.0). (3) LLIF + 4-screw MPF and supplemental interspinous process fixation range-of-motion reduction was comparable to LLIF + bilateral pedicle screw fixation in all directions ( P ≥ .6). Conclusions: LLIF with 4-screw MPF may provide inherent advantages over traditional 2-screw plating modalities. Furthermore, when coupled with interspinous process fixation, LLIF with MPF is a stable circumferential construct that provides biomechanical utility in all principal motions.


2002 ◽  
Vol 96 (3) ◽  
pp. 313-320 ◽  
Author(s):  
Frank Kandziora ◽  
Robert Pflugmacher ◽  
Katrin Ludwig ◽  
Georg Duda ◽  
Thomas Mittlmeier ◽  
...  

Object. The optimum fixation method to achieve atlantoaxial fusion after resection of the odontoid process remains a matter of discussion. Anterior atlantoaxial plate fixation has been described by Harms as a fixation procedure to be performed after transoral odontoid resection. In recent biomechanical and clinical studies investigators have shown that this procedure is a good alternative to established posterior atlantoaxial fixation techniques, but they have also indicated the biomechanical disadvantages of the Harms plate design. Therefore, three new anterior atlantoaxial plate designs were developed. The purpose of this study was to compare these three newly designed plate systems biomechanically with that used in Harms anterior atlantoaxial plate fixation. Methods. Twenty-four human craniocervical cadaveric specimens were tested in flexion, extension, axial rotation, and lateral bending in a nonconstrained testing apparatus by using a nondestructive stiffness method. Three-dimensional displacement of C1–2 was measured with an optical measurement system. Six different groups were examined: 1) control (24 specimens); 2) unstable (after odontoidectomy and dissection of the atlantoaxial ligaments; 24 specimens); 3) Harms (anterior atlantoaxial plate fixation according to Harms; six specimens); 4) subarticular atlantoaxial plate (SAAP; six specimens); 5) transpedicular atlantoaxial plate (TAAP; six specimens); and 6) subarticular atlantoaxial locking plate (SAALP; six specimens). Stiffness, range of motion, and neutral and elastic zones were determined. Compared with the Harms plate, stiffness was significantly higher when methods for placing the SAAP, TAAP, and SAALP devices were used (p < 0.05). Angular displacement of SAALPs was less than that demonstrated in any other group (p < 0.05). Stiffness values in any direction were significantly greater for the SAALP-fixed specimens than for the TAAP, SAAP, Harms, control, or unstable specimens (p < 0.05). Conclusions. Experimentally, the SAAP, TAAP, and Harms plate achieved less stable fixation than the SAALP. Therefore, if transoral odontoid resection is performed, SAALP-fixed spines will provide significantly improved stability compared with previous fixation devices and methods. This may be a necessary prerequisite for a fast and uneventful osseous fusion even without additional posterior stabilization.


2003 ◽  
Vol 99 (2) ◽  
pp. 221-228 ◽  
Author(s):  
Hideki Sudo ◽  
Itaru Oda ◽  
Kuniyoshi Abumi ◽  
Manabu Ito ◽  
Yoshihisa Kotani ◽  
...  

Object. Posterior lumbar interbody fusion (PLIF) was developed to overcome the limitations of posterolateral fusion in correcting spinal deformity and maintaining lumbar lordosis. In this study the authors compare the biomechanical effects of three different posterior reconstructions on the adjacent motion segment. Methods. Ten calf spinal (L2—S1) specimens underwent nondestructive flexion—extension testing (± 6 Nm). The specimens were destabilized at the L5—S1 levels after intact testing. This was followed by pedicle screw fixation with and without interbody cages as follows: 1) with straight rods (“aligned” posterolateral fusion); 2) with kyphotically prebent rods (“kyphotic” posterolateral fusion); and 3) with interbody cages combined with straight rods (“aligned” PLIF/posterolateral fusion). The range of motion (ROM) of the operative segments, the intradiscal pressure (IDP), and longitudinal lamina strain in the superior adjacent segment (L4–5) were analyzed. The ROM associated with aligned PLIF/posterolateral fusion-treated specimens was significantly less than both the aligned and kyphotic posterolateral fusion-treated procedures in both flexion and extension loading (p < 0.05). The aligned PLIF/posterolateral fusion was associated with greater IDP and the lamina strain compared with the aligned and kyphotic posterolateral fusion groups in flexion loading. Under extension loading, greater IDP and lamina strain were present in the kyphotic posterolateral fusion group than in the aligned posterolateral fusion group. The highest IDP and lamina strain were shown in the aligned PLIF/posterolateral fusion group. Conclusions. Compared with kyphotic posterolateral fusion, PLIF may lead to even higher load at the superior adjacent level because of the increased stiffness of the fixed segments even if local kyphosis is corrected by PLIF.