Determination of Minimal Pressure Support Level During Weaning from Pressure Support Ventilation

1998 ◽  
Vol 45 (2) ◽  
pp. 380
Author(s):  
Bock Hyun Jung ◽  
Youn Suck Koh ◽  
Chae Man Lim ◽  
Sang Do Lee ◽  
Woo Sung Kim ◽  
...  
2017 ◽  
Vol 32 (4) ◽  
pp. 741-751 ◽  
Author(s):  
Tobias Becher ◽  
Dirk Schädler ◽  
Philipp Rostalski ◽  
Günther Zick ◽  
Inéz Frerichs ◽  
...  

1995 ◽  
Vol 23 (3) ◽  
pp. 302-306 ◽  
Author(s):  
A. Uchiyama ◽  
H. Imanaka ◽  
N. Taenaka ◽  
S. Nakano ◽  
Y. Fujino ◽  
...  

The triggering system in pressure support ventilation needs to respond rapidly, especially in neonates. The aim of this study was to compare the effects of flow-triggered and pressure-triggered pressure support ventilation on neonatal mechanical ventilation using an animal model. Respiratory flow, airway pressure, oesophageal pressure, and diaphragmatic electromyogram were measured during pressure support ventilation in five anaesthetized rabbits. The animals were connected to a VIPBIRD (Bird, U.S.A.) (CPAP mode, pressure support ventilation, 5 cm H2O) and PEEP 0 cm H2O). Flow-triggering sensitivity was set at 0.21/min, 0.51/min, 1.011/min, or 1.51/min. Pressure-triggering sensitivity was set at −1.0 cm H2O. Shorter trigger delay and longer pressure support time were observed in flow-triggering. There was also less diaphragmatic activity in flow-triggering as evidenced by the amplitude of integrated diaphragmatic electromyogram and negative deflection of oesophageal pressure. The findings suggest that flow-triggering will prove superior to pressure-triggering in pressure support ventilation for neonates.