A NEW STRESS INDICATING RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR THE SIMULTANEOUS ESTIMATION OF ERTUGLIFLOZIN AND METFORMIN IN BULK AND ITS TABLET DOSAGE FORM

INDIAN DRUGS ◽  
2019 ◽  
Vol 56 (02) ◽  
pp. 39-46
Author(s):  
Babu D. China ◽  
◽  
C. Madhusudhana Chetty ◽  
Sk. Mastanamma

A simple, accurate, precise and robust reversed phase high performance liquid chromatographic (RP-HPLC) method was developed for the estimation of Ertugliflozin (ETZ) and metformin (MFN) in bulk and in tablet dosage form. The method was carried out by used Waters (5μm, C18 250 x 4.6 mm) column with mobile phase consists of 0.75 mM sodium dihydrogen orthophosphate buffer pH adjusted to 8.5, with NaOH, and acetonitrile in the ratio of 60:40 v/v, a flow rate of 1.5 mL/min and photodiode detection at 263 nm. The method was validated as per ICH guidelines with different parameters, the mean retention times of ertugliflozin and metformin were found to be 3.5& 2.0 min, respectively. The correlation coefficient values of calibration curves were found to be 0.999 for both ETZ and MFN, respectively. The LOD and LOQ for ertugliflozin and metformin were found to be 0.02-0.06 μg/mL and 17.5-58.3 μg/mL respectively.

INDIAN DRUGS ◽  
2015 ◽  
Vol 52 (08) ◽  
pp. 12-16
Author(s):  
S Vidyadhara ◽  
◽  
L. S Reddyvalam ◽  
T. Koduri ◽  
P. K. Borra ◽  
...  

A simple, accurate, precise high-performance liquid chromatographic (HPLC) method has been developed and validated for the simultaneous determination of abacavir sulphate (ABA) and lamivudine (LAM) in combined dosage form. Separation was performed on a C18 column [Agilent ODS UG 5 column, 250 mm x 4.5 mm], with methanol: water (50:50 V/V) isocratic elution using a flow rate of 1mL/min. Good sensitivity was observed with UV detection at 277 nm. After method development, the interference of other active compounds and excipients, repeatability and linearity, were investigated. Retention times of LAM and ABA were found to be 3.3 and 6.3 min, respectively. The method was validated over the range from 2.5-12.5 μg/mL for LAM and 5-25 μg/mL for ABA with correlation coefficients of 0.9997 and 0.9996, respectively. This method was shown to be accurate, robust, selective, linear, and repeatable and can be successfully employed in routine quality control for the simultaneous analysis of ABA and LAM in tablets.


Author(s):  
Sunkara Namratha ◽  
Vijayalakshmi A

Objective: Reversed-phase high-performance liquid chromatographic method (RP-HPLC) was developed for the assessment of lopinavir in the dosage form of tablet.Methods: Chromatogram was run through using Kromosil C18 4.5×150 mm using a mobile phase methanol: water of ratio 65:35% v/v with a rate of flow of 0.8 ml/min, measured by UV spectrometric detection at 265 nm. The method developed was validated in terms of precision, accuracy, linearity, and robustness parameters.Results: Retention time of lopinavir established at 2.482 min and percentage R.S.D of lopinavir found to be 1.0% and 0.5%, respectively. The method shows that good linearity range of 30–150 μg correlation coefficient of lopinavir was 0.997. The limit of detection was 2.97 and limit of quantification was 9.92, respectively. The percent purity of lopinavir was 99.87%.Conclusion: The suggested method (Rp-HPLC) for concurrent assay lopinavir was validated, which is appropriate method for the analysis oflopinavir quantitatively in tablet dosage forms and bulk.


Author(s):  
D. Chinababu

Aim: The objective of the study was simplest, accurate, precise and robust reversed phase high performance liquid chromatographic (RP-HPLC) method was developed for the estimation of Velpatasvir (VEL) and Sofosbuvir (SOF) in the bulk and its tablet dosage form. Study Design: The Quantitative and Qualitative estimation and designed forced degradation study of Velpatasvir & Sofosbuvir by RP-HPLC. Place and Duration of Study: The study was carried at Santhiram College of Pharmacy and time taken 4 months. Method: The method was attained by used Waters( 5µm, C18 250 x 4.6 mm) column with mobile phase consists of  0.5 mM disodium hydrogen phosphate buffer pH adjusted to 6.5, with Orthophosphoric acid and Methanol in the ratio of 78:22 v/v, a flow rate of 1.0 mL/min and ultraviolet detection at 285 nm. Results: The method was validated as per ICH guidelines with different parameters, the mean retention times of VEL and SOF were found to be 2.8 & 4.7 min respectively. The resolution between VEL and SOF was found to be 10.66. The Correlation coefficients for calibration curves within the detection range of 32.5 - 97.5 and 125 - 375 µg/mL were 0.999 for VEL and SOF respectively. The LOD and LOQ for VEL and SOF were found to be 0.0068-0.029 µg/mL and 0.104-0.342 µg/mL respectively. Conclusion: The results were indicated that the developed method was used for the routine analysis of VEL & SOF combined form in bulk and its commercial formulation. To the best of our knowledge, there was no method of RP-HPLC for the determination of VEL alone or in combination with SOF molecule.


2018 ◽  
Vol 8 (6-s) ◽  
pp. 78-82 ◽  
Author(s):  
Vishal Singh Solanki ◽  
RAM SINGH BISHNOI ◽  
Raviraj Baghel ◽  
Deepti Jain

A simple precise and economical reverse phase high performance liquid chromatographic method has been developed and validated for the simultaneous estimation of Cilnidipine (CDP), Atenolol (ATL) and Chlorthalidone (CTD).The chromatographic separation was achieved by using Hypersil- keystone C18 (4.6 x 250mm, 5μm) under isocratic conditions The mobile phase consisted of methanol and triple distilled water (80/20, v/v) having pH 7 with a flow rate of 1.0 mL/min. The eluents were monitored at 225 nm for simultaneous measurement.The selected chromatographic conditions were found to effectively separate CDP (Rt: 3.25 min), ATL (Rt: 5.366 min) and CTD (Rt: 9.025 min) having good resolution. The developed method was validated for linearity, accuracy, precision, LOD, LOQ, robustness and for system suitability parameters as per ICH guidelines. In this study, an excellent linearity was obtained with r2 = 0.999, r² = 0.999, r² = 1, for CDP, ATL and CTD respectively. The developed chromatographic method proved to be simple, precise, accurate, robust and reproducible Thus, this method would be employed for routine simultaneous quantification of CDP, ATL and CTD in bulk form or tablet dosage form.   Keywords: Cilnidipine, Atenolol and Chlorthalidone, RP-HPLC.


2014 ◽  
Vol 2014 ◽  
pp. 1-5 ◽  
Author(s):  
Patel Bhaumik ◽  
Gopani Mehul ◽  
Vikani Kartik ◽  
Patel Rashmin ◽  
Patel Mrunali

This paper describes validated reverse phase high-performance liquid chromatographic (RP-HPLC) method for simultaneous estimation of trihexyphenidyl hydrochloride (THP) and risperidone (RSP) in the pure powder form and in combined tablet dosage form. The HPLC separation was achieved on a core shell C18 (100 mm length × 4.6 mm, 2.6 μm particle size) using methanol : ammonium acetate buffer 1% (85 : 15 v/v; pH-6.5) as mobile phase and delivered at flow rate of 0.8 mL/min. The calibration plot showed good linear relationship with r2 = 0.997 ± 0.001 for THP and r2 = 0.998 ± 0.001 for RSP in concentration range of 50–175 μg/mL and 50–175 μg/mL, respectively. LOD and LOQ were found to be 0.40 and 1.29 μg/mL for THP and 1.24 and 3.92 μg/mL for RSP. Assay of THP and RSP was found to be 100.16 ± 0.03% and 99.83 ± 0.02%, respectively. THP and RSP were subjected to different stress conditions (acidic, basic, oxidative, thermal, and photolytic degradation). The degraded product peaks were well resolved from the pure drug peak. The method was successfully validated as per the ICH guidelines. The developed RP-HPLC method was successfully applied for the estimation of THP and RSP in tablet dosage form.


INDIAN DRUGS ◽  
2021 ◽  
Vol 58 (01) ◽  
pp. 76-78
Author(s):  
Kailasa Mangasree ◽  
◽  
Biswabara Roy ◽  
Harunrasheed Shaik ◽  
Manjunath S. Yalagatti ◽  
...  

A reverse phase high performance liquid chromatographic method was developed for simultaneous estimation of gemcitabine and clarithromycin in bulk and tablet dosage form. The method was developed by using acetonitrile: methanol: 50 mM sodium acetate buffer (adjusted to pH-3 with orthophosphoric acid (40:40:20) as a mobile phase at the flow rate of 1mL/min. The retention time was found to be 2.365 and 5.999 minutes, respectively. The method showed linearity in the concentration range 10-50 µg/mL in respect of both drugs. The % RSD was less than 0.4% for gemcitabine and 0.2% for clarithromycin. Mean percentage recovery was found to be 99.63 and 99.69 for gemcitabine and clarithromycin. The tailing factor for gemcitabine and clarithromycin was not less than 0.9 and not more than 1.4 during the robustness study. The average percentage assay was calculated and found to be 99.83 for gemcitabine and 98.89 for charithromycin. The limit of detection and quantification for gemcitabine was found to be 0.41 and 1.79 µg/mL. For Clarithromycin, the value was 0.58 and 1.37 µg/mL respectively.


2020 ◽  
Vol 10 (5) ◽  
pp. 143-148
Author(s):  
T. Hanuman ◽  
T. Sivakkumar ◽  
S. Sridhar

A simple, specific and accurate reverse phase high performance liquid chromatographic method was developed for the simultaneous determination Sofosbuvir and Velpatasvirin pharmaceutical dosage form. The column used was Kromosil C18(150mm x 4.6 mm, 5mm)in isocratic mode, with mobile phase containing phosphate buffer andacetonitrile(70:30v/v). The buffer is prepared by adding 1.41gm of sodium dihyrogen ortho phosphate in a 1000ml of volumetric flask add about 900ml of milli-Q water added and degas to sonicate and finally make up the volume with water then pH adjusted to 3.5 with dil. orthophosphoric acid solution. The flow rate was 1.0ml/ min and effluents were monitored at 260 nm. The retention times of Sofosbuvir and Velpatasvirwere found to be 2.404min and 2.986 min, respectively. The linearity for Sofosbuvir and Velpatasvirwere in the range of 40-240µg/ml and 10-60 µg/ml respectively. The recoveries of Sofosbuvir and Velpatasvirwere found to be 99.64% and 99.25%, respectively. The proposed method was validated and successfully applied to the estimation of Sofosbuvir and Velpatasvirin combined tablet dosage forms. Keywords: Sofosbuvir, Velpatasvir, Validation, Buffer and ICH Guidelines.


1970 ◽  
Vol 3 (1) ◽  
pp. 49-53 ◽  
Author(s):  
Gaurav Patel ◽  
Sanjay Patel ◽  
Dhamesh Prajapiti ◽  
Rajendra Mehta

A reverse phase high performance liquid chromatographic (RP-HPLC) method has been developed for the simultaneous estimation of Amlodipine Besylate and Hydrochlorothiazide in combine dosage form. Amlodipine Besylate (AML) is a long acting calcium channel blocker and in the treatment of CVS disorder. Hydrochlorothiazide (HCT) is a diuretic and antihypertensive. The mobile phase used was a combination of Water: Methanol (70:30). The detection of the combined dosage form was carried out at 245nm and a flow rate employd was 0.5ml/min. The retention time for Amlodipine Besylate and Hydrochlorothiazide was found to be 6.95 and 2.65 min respectively. Linearity was obtained in the concentration range of 6 to 18μg/ml of Amlodipine Besylate and 6 to 18μg/ml of Hydrochlorothiazide with a correlation coefficient of 0.997 and 0.9974. Detector consists of photodiode array detector; the reversed phase column used was RP-C18 (5 μm size, 250mm, 4.6mm i.d.) at ambient temperature. The developed method was validated according to ICH guidelines and values of accuracy, precision and other statistical analysis were found to be in good accordance with the prescribed values. Thus the proposed method is precise, selective and rapid for simultaneous estimation of Amlodipine Besylate and Hydrochlorothiazide in routine analysis. Key Words: Simultaneous Estimation; Amlodipine Besylate; Hydrochlorothiazide; HPLC. DOI: 10.3329/sjps.v3i1.6798S. J. Pharm. Sci. 3(1): 49-53


2017 ◽  
Vol 4 (1) ◽  
pp. 5-10 ◽  
Author(s):  
Ramesh Jayaprakash ◽  
Senthil Kumar Natesan

A simple, economic, accurate, sensitive, specific and precise stability indicating reverse phase high performance liquid chromatographic [RP-HPLC] method for the determination of Regorafenib in pure and tablet dosage from was developed and validated. The chromatographic separation was carried out using Phenomenex Luna-C18 column (4.5x250 mm; 5 µm particle size) as a stationary phase and methanol: acetonitrile: water (55:25:20 v/v/v) as a mobile phase. The flow rate of 1 mL/min was used with PDA detection at 275 nm. The retention time of Regorafenib was 2.480 min. RP-HPLC method was developed with linearity range of 40-240 µg/mL of Regorafenib. The correlation coefficient [r2] was found to be 0.9999. The assay results obtained was in good agreement with the corresponding labeled amount by developed method within range of 98.83 ± 0.6937. Accuracy of the method was confirmed by recovery studies and the recoveries were found to be between 99.61 % and 100.22 %, the corresponding %RSD was found to be 0.2029. Precision, LOD, LOQ, specificity, robustness and ruggedness were performed as per ICH Q2(R1) guidelines and were within the acceptance criteria. This method can be conveniently used to detect the possible degradation product in the dosage form of Regorafenib during stability studies (acidic, alkaline, oxidative, thermal and photolytic). The method proved to be effective on the analysis of stressed marketed tablet formulation.


2012 ◽  
Vol 10 (2) ◽  
pp. 67-70
Author(s):  
Abdullah Al Masud ◽  
Mohammad Saydur Rahman ◽  
Towfika Islam ◽  
Saki Sultana ◽  
Moynul Hasan ◽  
...  

A simple, reproducible and efficient reversed phase high performance liquid chromatographic (RPHPLC) method has been developed for the estimation of a recently approved anti allergic drug, amlexanox in oral paste dosage form. The separations were carried out on a Zorbax Eclipse XBD, C18 column (150 x 4.6 mm; 5?m) at a flow rate of 1.50 ml/min. by using mobile phase comprising of mixed buffer (pH adjusted to 6.50) and methanol (50:50 v/v). The injection volume was 10 ?l and the peaks were detected at 244 nm. The linear dynamic range found to be in the concentration range of 15-35 ?g/ml and coefficient of correlation was found to be 0.999. The %RSD value was below 2.0 for intra-day and inter-day precision which indicated that the method was highly precise. The LOD (Limit of detection) and LOQ (Limit of quantitation) were found to be 3.8 ng/ml and 12.5 ng/ml, respectively which revealed that the method was highly sensitive. The percentage recovery of amlexanox ranged from 99.31 to 99.75%, indicating the accuracy of the method and absence of interference from the excipients present in the formulation. The proposed method was simple, fast, accurate and reproducible and hence can be applied for routine quality control operations of amlexanox in oral paste dosage form. Key words: Amlexanox, Anti allergic, RP-HPLC, LOD, LOQ. DOI: http://dx.doi.org/10.3329/dujps.v10i2.11782 Dhaka Univ. J. Pharm. Sci. 10(2): 67-70, 2011 (December)