transmission electron microscopy sample
Recently Published Documents


TOTAL DOCUMENTS

18
(FIVE YEARS 4)

H-INDEX

4
(FIVE YEARS 0)

Author(s):  
J. Demarest ◽  
B. Austin ◽  
J. Arjavac ◽  
M. Breton ◽  
M. Bergendahl ◽  
...  

Abstract Demarest et al. concluded in their previous report that a ten times improvement in placement accuracy was required to enable automated transmission electron microscopy (TEM) sample preparation, and wafer alignment by GDS coordinates demonstrated a factor of two improvement in comparison to optical or scanning electron microscope based processes. This paper provides an additional update on this project. The study is about a GDS based process developed to simplify the complicated workflow for examining discrete electrical failures. The results of this study indicated that the recipe prototype developed on a test structure had a unique feature that consisted of an approximately 45nm by 200nm Cu line segment. Executing the prototype recipe on a wafer at the same process point fabricated 6 months after the original wafer yielded four identical successful samples of about 30nm sample thickness. This technique can thus be extended to large 2D arrays of small structures.